Neonatal hearing screening with transient evoked otoacoustic emissions using narrow band stimulus

Abstracts

PURPOSE:

to assess the sensitivity and specificity of an automated Transient Evoked Otoacoustic Emissions equipment using narrow band stimulus.

METHODS:

the hearing screening results of 300 newborns were analyzed. A portable automated equipment was used with narrow band stimulus to conduct the Transient Evoked Otoacoustic Emissions. All newborns were also submitted to the Brainstem Auditory Evoked Potentials using 100μs-duration click stimuli, which was considered the gold standard test.

RESULTS:

the proportion of "pass" results in the neonatal hearing screening was 90.7%. Considering the gold standard results, it was observed a false-positive rate of 9%. The Transient Evoked Otoacoustic Emissions with narrow band stimulus presented sensitivity of 100% and specificity of 92%.

CONCLUSION:

the narrow band stimulus was efficient enough to be used in neonatal hearing screening programs.

Infant, Newborn; Hearing Tests; Otoacoustic Emissions, Spontaneous; Evoked Potentials, Auditory; Hearing Loss


OBJETIVO:

avaliar a sensibilidade e a especificidade de um equipamento automático de Emissão Otoacustica Evocada Transiente que utiliza um estímulo de banda estreita, Narrow.

MÉTODOS:

foi analisado o resultado da Triagem Auditiva Neonatal de 300 neonatos. A Emissão Otoacústica Evocada Transiente foi realizada com um equipamento portátil automático, com o estímulo de banda estreita - Narrow. Todos os neonatos foram submetidos ao Potencial Evocado Auditivo de Tronco Encefálico com o estímulo clique, duração 100μs como padrão-ouro.

RESULTADOS:

o percentual de "passa" na Triagem Auditiva Neonatal foi de 90,7%. Considerando os resultados do padrão-ouro, foi observado uma taxa de falso-positivo de 9%. A sensibilidade e especificidade encontradas foram de 100% e 92%, respectivamente.

CONCLUSÃO:

o estímulo de banda estreita - Narrow mostrou-se eficaz para ser utilizado nos programa de Triagem Auditiva Neonatal Universal.

Recém-Nascido; Testes Auditivos; Emissões Otoacústicas Espontâneas; Potenciais Evocados Auditivos, Perda Auditiva


Introduction

Efforts are being made in order to establish a National Policy on attention to auditory health, considering that the impact caused by sensory deprivation during child development has turned hearing impairment a public health problem 11. Política nacional de atenção à saúde auditiva. [cited 2006 sep 21]. Disponível em: http://portalsaude.saude.gov.br/portalsaude/
http://portalsaude.saude.gov.br/portalsa...
. In consequence to the implementation of programs aiming towards early detection of hearing loss 22. Hilú MRPB, Zeigelboim BS. O conhecimento, a valorização da triagem auditiva neonatal e a intervenção precoce da perda auditiva. Rev CEFAC. 2007;9(4):563-70.

3. Soares CP, Marques LR, Flores NGC. Triagem auditiva neonatal: aplicabilidade clínica na rotina dos médicos pediatras neonatologistas. Rev CEFAC. 2008;10(1):110-6.
- 44. Dantas MBS, Anjos CAL, Camboim ED, Pimentel MCR. Resultados de um programa de triagem auditiva neonatal em Maceió. Rev. Bras. Otorrinolaringol. 2009;75(1):58-63., optimization of the procedures used to ensure the efficiency of the developed programs and universal access to Neonatal Hearing Screening (NHS) have been sought55. Côrtes-Andrade IF, Bento DV, Lewis DR. Emissões Otoacústicas Evocadas por Estímulo Transiente: protocolos de Triagem Auditiva Neonatal. Rev CEFAC [online]. 2013; 15(3):521-7. Available from: http://www.scielo.br/pdf/rcefac/2012nahead/48-11.pdf
http://www.scielo.br/pdf/rcefac/2012nahe...
. In this context, the time of test conduction, its sensitivity (efficacy) and specificity (efficiency) determined by the employed techniques are extremely important 66. Davis A, Bamford J, Stevens J. Performance of neonatal and infant hearing screens: sensitivity and specificity. Br J Audiol. 2001;35(1):3-15..

Literature has shown that the Transient Evoked Otoacoustic emissions Test (TEOAE) is the procedure deemed adequate for NHS, as it is a low-cost procedure that uses a weak intensity acoustic stimulus covering a wide frequency range, and also for being a quick and simple procedure 77. Joint committee on infant hearing (JCIH). Position statement. 2007 Avaliable from: < http://www.jcih.org/ExecSummFINAL.pdf >. Acesso em: 19/06/2009.
http://www.jcih.org/ExecSummFINAL.pdf ...

8. Kemp DT. Otoacustic emissions, their oringinin cochlear function, and use. Br. Med. Bull. 2002;63(1):223-41.
- 99. Walsh KP, Pasanen EG, McFadden D. Evidence for dynamic cochlear processing in otoacoustic emissions and behavior (A). J. Acoust. Soc. Am. 2009;125:2720. .

However, one of the main difficulties in registering the TEOAE during NHS is typically the effect of noise 88. Kemp DT. Otoacustic emissions, their oringinin cochlear function, and use. Br. Med. Bull. 2002;63(1):223-41., is it is not always possible to conduct NHS in acoustically treated rooms. Thus, the screening generally requires the newborn to be calm during the test and in adequate conditions examination tends to be quicker, detecting a robust emission sign level 1010. Vohr BR, White KR, Brancia Maxon A, Johnson MJ. Factors affecting the interpretation of transient-evoked otoacousticm emission results in neonatal hearing screening. Semin Hear. 1993;14:57-72. .

Therefore, different states of activity may affect the TEOAE, such as, for example, sneezes, mandible and muscular movements. Both the physiological processes as well as the state of activity may result in weak response level of the TEOAE, where the physiological noises are reported as impacting the range between 500-1500 Hz 1111. Rhoades K, McPherson B, Smyth V, Kei J, Baglioni A. Effects of background noise on click-evoked otoacoustic emissions. Ear Hear. 1998;19(6):450-62.. Thus, the authors agree that robust responses are common in medium and high frequency bands, though usually with lower general reproducibility, due to the influence of noise in the low frequencies 88. Kemp DT. Otoacustic emissions, their oringinin cochlear function, and use. Br. Med. Bull. 2002;63(1):223-41. , 1212. Norton SJ, Gorga MP, Widen JE, Folsom RC, Sininger Y, Cone-Wesson B et al. Identification of neonatal hearing impairment: evaluation of transient evoked otoacoustic emission, distortion product otoacoustic emission, and auditory brain stem response test performance. Ear Hear. 2000;21(5):508-28. .

Since then, equipment manufacturers have tried to maintain undesired responses at a low level, by using filters1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75. . The high-pass and low-pass filters are electronic and allow the analysis of a single specific signal frequency band, thus increasing the signal-to-noise ratio in TEOAE registration1414. Gong Q, Ye D. A study of the reduction of the stimulus artifact using a predictor-subtractor-restorer (PSR) filter. Conf Proc IEEE Eng Med Biol Soc. 2005;2:1185-8.. The high-pass filter allows only signals with frequencies above the established values to be analyzed; reducing the contribution of low frequencies on the TEOAE trace, and the low-pass filter does the same enabling the analysis of frequencies below the established value1515. Glattke TJ, Robinette MS. Transient evoked otoacoustic emissions in population with normal hearing sensitivity. In: Robinette MS, Glattke TJ . Otoacoustic emissions: clinical applications. 3nd. ed. New York: Thieme; 2007. P.87-107..

In the year of 1994, authors have described and implemented the use of a parameter that uses the narrow-band filter, named Narrow. The research studied the sensitivity and the specificity in 162 ears and the results showed that 100% of newborns passed in the TEOAE with the Narrow parameter, and that the sensitivity and specificity of screening with the narrow-band criteria were 100% and 98.1%, respectively. The group classified as noisy had a higher mean for the variable test time and the authors concluded that the low frequencies are filtered before affecting the signal-to-noise ratio result 1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75. .

In 1999, researchers in Milan, aiming to reduce the initial artifact associated to the TEOAE test, studied the efficacy of the effect of the "high-pass" and "low-pass" band filters, using the results from test signal and reproducibility, calculated for the general response as well as for each frequency band. The TEOAE were tested bilaterally in 629 newborns, and the cut-off frequency occurred outside the 1.6-4.2 kHz interval. The results using the band filter were compared to the TEOAE test that did not use the filter. Results showed that 46.0% of newborns "passed" the TEOAE before using the "high-pass" and "low-pass" filters, increasing the "pass" results percentage to 47.4% after using the filter. The study also concluded that the reproducibility with the "high-pass" and "low-pass" filter was always greater than when the test was conducted without the filter1616. Ravazzani P, Tognola G, Grandori F. Optimal band pass filtering of transient- evoked otoacoustic emissions in neonates. Audiology. 1999;38(2):69-74. . Another study has also observed that the reproducibility and the signal-to-noise ratio may be improved when "high-pass" and "low-pass" filters are used when registering TEOAE1717. Driscoll C, Kei J, McPherson B . Outcomes of transient evoked otoacoustic emission testing in 6-year-old school children: a comparison with pure tone screening and tympanometry. Int J Pediatr Otorhinolaryngol. 2001;57(1):67-76. and that they are both especially useful to reduce the interference cause by noise below 1.0 kHz 1818. Vicky Z, McPherson B . A review of otoacoustic emission hearing screening technology. Audiological Medicine. 2008;6(2):100-14..

Different pass/fail parameters and criteria may be used in automatic equipment that may change the sensitivity and specificity of the test used in a population of newborns. However, safe criteria should be chosen in order to ensure that false-positives and false-negatives do not occur in excessive or allowed numbers, maintaining this test's reliability. Thus, the purpose of the present study is to assess the sensitivity and specificity of a TEOAE automatic equipment that uses a narrow-band stimulus (narrow) with a "high-pass" and "low-pass" filter.

Methods

The Newborn Hearing Screening was conducted at the Amparo Maternal Hospital, located in the city of São Paulo. The facility has a partnership with the Pontifical Catholic University of São Paulo (PUCSP) for the conduction of scientific investigations.

The study was approved by the Research Ethics Committee of the Pontifical Catholic University of São Paulo - PUC-SP, protocol number 063/2010, and was funded by the Coordination for Improvement of Higher Level Personnel (CAPES). In compliance with the bioethics determinations of the Research Ethics Committee at PUCSP, only the newborns whose parents signed the Free and Informed Consent Term were included in the study.

The newborns included in the study were required to be over 24 hours old and over 37 weeks gestational age. Those with congenital syndromes or neurological disorders evidenced in the records were not included.

The subjects of this study were 300 newborns (600 ears), evaluated in the period between November 2010 and November 2011, of which 154 (51.3%) were males and 146 (48.7%) were females. Only 42 (14%) newborns had Risk Indicators for Hearing Loss (RIHL).

Literature shows that the "fail" index in the TEOAE may be attributed to several factors, such as: probe adjustment and monitoring, facilitating maneuver and newborn state of consciousness. In order to reduce false-positives, these factores were controlled during data collection. For test administration, all newborns were placed in the cribs provided by the hospital, laying on their sides, or in their mothers' arms.

TEOAE research was conducted using a portable Otoport Lite automatic equipment by Otodynamics. The stimulus was a non-linear click, lasting 80 microseconds and intensity of 84dB peSPL. 260 stimuli were used for response collection, and the maximum test time was 300 seconds.

The TEOAE automatic device conducts tests with different screening parameters that are already part of the device's standard protocol and may be chosen and adjusted by the evaluator. In this study, the inserted standard was the one named narrow band, or Narrow parameter, with the purpose of reducing noise, and conducts an algorithmic and vector analysis in a frequency range between 841 Hz - 4757 Hz, using a filter between the frequencies of 1600 Hz - 3200 Hz. The response is automatically provided by the device through the signals "Pass" (pass) or "Refer" (fail).

Then, all newborns underwent the Brainstem Evoked Auditory Potential (BEAP) using an Eclipse Black Box - software EP25 equipment by Interacoustics MedPC. The parameters were: 100μs click stimulus; Repetition Rate in 27.7Hz; Alternate Polarity; Filter 100-3000Hz and a 25ms Window. This test was considered to gold-standard to ensure and verify the sensitivity and specificity of the responses obtained in the TEOAE.

For this study, the Minimum Response Level (MRL) considered normal was 20 dBnHL. The BEAP was initially tested in 40 dBnHL intensity and, then in 20 dBnHL intensity. When the presence of wave V in 20 dBnHL was not observed, the electrophysiological threshold was tested, beginning in 80 dBnHL for assessment of auditory integrity. Then, intensity was decreased in 20dB steps until wave V could no longer be observed, and then increased in 10 dBnHL steps, until this wave could be registered and reproduced once again.

The electrophysiological threshold was considered the lowest intensity in which wave V could be observed and reproduced, and wave V was considered as the greatest negative deflection occurring between 5 and 20 milliseconds after stimulus presentation.

It should be noted that the EP25 software has a detection method that considers residual noise (40 nV) and wave reproducibility (50%) as criteria te determine presence or absence of response. Therefore, when the register reached these levels, the software would automatically stop response registration after a minimum of, at least, 800 presented stimuli. However, in the absence of response, the researcher was able to withdraw this criterion, allowing the answer to be evoked with up to 2000 presented stimuli.

In the cases where absence of wave V in 20 dBnNA, in air conduction, the mothers were referred to a complete audiological investigation in a Child Hearing Center, in an interval of approximately 30 days.

During the research at the hospital, four result combinations could occur (Table 1):

Table 1:
Result possibilities in the newborn hearing screening with Transient Evoked Otoacoustic Emissions and gold-standard using Brainstem Evoked Auditory Potential

Statistical measurements were conducted and these reflect the validity of the "pass-fail" criteria of the TEOAE through narrow-band - Narrow considering BEAP as the gold-standard. The following aspects were analyzed: sensitivity, specificity, positive predictive value, negative predictive value, accuracy, likelihood ratio and Youden index, corresponding to the Narrow "pass-fail" criteria (Altman, 1999). The statistical analysis was made using the Minitab 16 e SPSS 18 applications.

Results

Table 2 shows the percentages of TEOAE results for the "pass-fail" criteria for narrow-band (Narrow), where it may be seen that the "pass" percentage was 90.7%.

Table 2:
Distribution of results of the Transient Evoked Otoacoustic Emission for the "pass-fail" criterion considering the 600 tested ears

When analyzing the BEAP, gold-standard, it was observed that five (0.8%) ears had absent responses and 595 (99.2%) present responses.

The distributions of joint and marginal frequencies of the BEAP diagnosis and in the TEOAE for the "pass-fail" criteria investigated in this study are found in Table 3.

Table 3:
Distribution of the results in the Brainstem Evoked Auditory Potential (present and absent) and in the Transient Evoked Otoacoustic Emission (pass and fail)

It may be seen that, in the Narrow "pass-fail" criteria, the five ears with absent responses in the BEAP were classified as "fail". Thus, there were no false-negatives. The number of newborns with absent responses in the TEOAE and present responses in BEAP was zero.

Table 4 shows the results regarding time of examination.

Table 4:
Descriptive Statistics Values observed for examination time, in seconds

The false-positive proportion estimate shows that, of the 595 with present responses in the BEAP, 51 were classified as "fail" by the Narrow "pass-fail" criteria. Thus, the false-positive proportion was 0.09.

Sensitivity, considering the gold-standard, was 100% and specificity was 92% with a confidence interval (95%) of [0.89; 0.94]. The positive predictive value was calculated and a 0.09 value was obtained with a confidence interval (95%) of [0.03; 0.20] showing that the pass response accurately indicates a present response. Since there was a low prevalence of deaf subjects in the sample, the negative predictive value is not reliable and was therefore excluded from the study.

Even though predictive values are useful in practice, they depend on the prevalence of the interesting result (fail). In this case, the prevalence of "fail" was low, and therefore there is a strong indication that the "pass" result in the TEOAE accurately indicates present response, and little indication that a "fail" result will really indicate a patient with an absent response.

The value of "pass-fail" accuracy of the TEOAE for the Narrow criteria is 92% in a confidence interval [0.89; 0.94] and the observed value of the likelihood ration (LR) was 11.7 in the Narrow "pass-fail" criteria.

The Youden index was used and showed that the sensitivity of the Narrow "pass-fail" criteria was equal to 1, which coincided with its specificity.

Discussion

Different parameters may be applied to the software of newborn screening automatic equipment, and among these are those using broad band and narrow band (Narrow) stimuli1414. Gong Q, Ye D. A study of the reduction of the stimulus artifact using a predictor-subtractor-restorer (PSR) filter. Conf Proc IEEE Eng Med Biol Soc. 2005;2:1185-8. , 1515. Glattke TJ, Robinette MS. Transient evoked otoacoustic emissions in population with normal hearing sensitivity. In: Robinette MS, Glattke TJ . Otoacoustic emissions: clinical applications. 3nd. ed. New York: Thieme; 2007. P.87-107.. However, differences among the parameters may lead to an increase in the number of false-positives and false-negatives, decreasing the quality of the program 1919. National center for hearing assessment and management - NCHAM. Issues & evidence: prevalence of congenital hearing loss. [homepage na internet]. Utah, US; 2002. [atualizado em 11abr 2008; [acesso em 23 02 13]. Disponível em <http://www.infanthearing.org/status/unhsstate.html>
http://www.infanthearing.org/status/unhs...
.

The present study has analyzed the application of an automatic TEOAE equipment using a narrow band stimulus - Narrow, highlighting aspects related to sensitivity, specificity and examination time, extremely important features in a NHS program.

Newborn hearing screening was conducted in 300 newborns (600 ears) using TEOAE. The results show that 90.7% of the ears "passed" hearing screening with narrow band stimulus "Narrow". Studies using the broad band stimulus have reported worse results, 66.7%2020. Clarke P, Iqbal M, Mitchell S. A comparison of transientevoked otoacoustic emissions and automated auditory brainstem responses for pre-discharge neonatal hearing screening. Int J Audiol. 2003;42(8):443-7., 86.1% 2121. Pereira PKS, Martins AS, Vieira MR, Azevedo MF. Programa de triagem auditiva neonatal: associação entre perda auditiva e fatores de risco. Pró-Fono R. Atual. Cient. 2007;19(3):267-78. and 64% 2222. Freitas VS, Alvarenga KF, Bevilacqua MC, Martinez MAN, Costa OA. Análise crítica de três protocolos de triagem auditiva neonatal. Pró-Fono R. Atual. Cient. 2009;21(3):201-6., considering the first moment of screening. It is thus observed that when the broad band stimulus is used, the amount of "fail" results is higher and considering that there were no cases of false-negatives for the "Narrow" stimulus in the present study, the results show that using this stimulus may result in better efficiency of the screening program with a reduction of new appointments for retesting.

In this context, considering the results for the gold-standard, the present study had a false-positive rate of 9%. However, a study mentioned above2222. Freitas VS, Alvarenga KF, Bevilacqua MC, Martinez MAN, Costa OA. Análise crítica de três protocolos de triagem auditiva neonatal. Pró-Fono R. Atual. Cient. 2009;21(3):201-6. that used the broad band stimulus estimated a 36% rate of false-positives in NHS using TEOAE. Another study, conducted in England, that also used the broad band stimulus, observed a rate of 11.2% of false-positives 2323. Thornton AR, Marotta N, Kennedy CR. The order of testing effect in otoacoustic emissions and its consequences for sex and ear differences in neonates. Hear Res. 2003;184(1-2):123-30.. A reference from 2005 has also sown a greater percentage (12.2%) than found in the present study using narrow band2424. Von Specht H, Beyer U. Hearing screening with transient evoked otoacoustic emissions in preterm neonates [Internet]. 2007. Available from: http://www.auditio.com/ierasg/abstract/126.htm
http://www.auditio.com/ierasg/abstract/1...
.

The differences found between the studies mentioned above may be related to the fact that the narrow band stimulus - Narrow restricts (with the use of high-pass and low-pass filters) the assessment of lower frequencies, mainly emphasizing frequency regions between 1600Hz-3200Hz 1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75.. Previous studies reinforce that satisfactory signal-to-noise relationships in frequency bands of 1.0kHz and 1.5kHz, as part of the "pass-fail" criteria in a NHS program with TEOAE would not be necessary 55. Côrtes-Andrade IF, Bento DV, Lewis DR. Emissões Otoacústicas Evocadas por Estímulo Transiente: protocolos de Triagem Auditiva Neonatal. Rev CEFAC [online]. 2013; 15(3):521-7. Available from: http://www.scielo.br/pdf/rcefac/2012nahead/48-11.pdf
http://www.scielo.br/pdf/rcefac/2012nahe...
, 1010. Vohr BR, White KR, Brancia Maxon A, Johnson MJ. Factors affecting the interpretation of transient-evoked otoacousticm emission results in neonatal hearing screening. Semin Hear. 1993;14:57-72.. However, several factors beyond the type of stimulus may be related to the differences found among the several studies, including hours of life and the newborn's state of consciousness during screening, using or not using a facilitation maneuver and environmental noise. In the present study, these factors were controlled and the newborn was over 24 hours old and was naturally sleeping during screening, similarly to the 2009 study2222. Freitas VS, Alvarenga KF, Bevilacqua MC, Martinez MAN, Costa OA. Análise crítica de três protocolos de triagem auditiva neonatal. Pró-Fono R. Atual. Cient. 2009;21(3):201-6. , differing only by the fact that, in the latter, the TEOAE equipment was not portable or automatic. In the 2003 study2020. Clarke P, Iqbal M, Mitchell S. A comparison of transientevoked otoacoustic emissions and automated auditory brainstem responses for pre-discharge neonatal hearing screening. Int J Audiol. 2003;42(8):443-7. screening was conducted by a researcher with no formal training in hearing screenings, which may also have influenced test results; moreover, the mean age during screening was 20 hours, while in the present study all newborns were over 24 hours old, and, therefore, the presence of vernix in the ear may have influenced the results.

Furthermore, the present study was concerned with controlling procedures that may influence the "pass-fail" results in NHS testing, such as probe adjustment, using a "facilitation maneuver", probe monitoring and stability 2222. Freitas VS, Alvarenga KF, Bevilacqua MC, Martinez MAN, Costa OA. Análise crítica de três protocolos de triagem auditiva neonatal. Pró-Fono R. Atual. Cient. 2009;21(3):201-6. , 2525. Guedes MC, Passos SN, Gomez MVSG, Bento RF. Estudo da reprodutibilidade das emissões otoacusticas em indivíduos normais. Rev. Bras. Otorrinolaringol. 2002;68(1):34-8. , 2626. Simonek C, Azevedo MF . Respostas falso-positivas na triagem auditiva neonatal universal: possíveis causas. Rev CEFAC. 2011;13(2):292-8.. Hours of life (≥24 hours), the newborn's state of consciousness and noise were also controlled1212. Norton SJ, Gorga MP, Widen JE, Folsom RC, Sininger Y, Cone-Wesson B et al. Identification of neonatal hearing impairment: evaluation of transient evoked otoacoustic emission, distortion product otoacoustic emission, and auditory brain stem response test performance. Ear Hear. 2000;21(5):508-28. , 2727. Brazelton TB. The manual. In: Brazelton TB . Neonatal Behavioral Assesment Scale. 2. ed. Philadelphia: Spastics International medical Publications, 1984. P. 17-77. , 2828. Korres SG, Balatsouras DG, Nikolopoulos T, Korres GS, Ferekidis E. Making universal newborn hearing screening a success. Int J Pediatr Otorhinolaryngol. 2006;70(2):241-6..

It should be noted that the implications caused by the false-positive rate imply in the increase in program cost due to unnecessary referral for audiological diagnosis, as well as in emotional stress of parents and/or caregivers 2929. Keren R, Helfand M, Homer C, McPhillips H, Lieu TA. Projected cost-effectiveness of statewide universal newborn hearing screening. Pediatrics. 2002;110(5):855-64..

The quality of NHS programs may be measured based on the efficacy and efficiency of the procedure employed 3030. Fletcher RH, Fletcher SW. Prevenção. In: Fletcher R, Fletcher S. Epidemiologia clínica: elementos essenciais. Porto Alegre: Artes Médicas, 2006, 4.ed. P. 179-201.. In both hearing and hearing impaired individuals, screening procedures may provide true or false outcomes, confirmed by a reference test, considered a gold-standard. The main purpose of NHS programs is using a screening test with precise "pass-fail" criteria.

A program is considered effective when it uses a test where the sensitivity is equal or close to 100%3030. Fletcher RH, Fletcher SW. Prevenção. In: Fletcher R, Fletcher S. Epidemiologia clínica: elementos essenciais. Porto Alegre: Artes Médicas, 2006, 4.ed. P. 179-201.. When analyzing this study's results it has been observed that the Narrow criterion had 100% sensitivity. This means that no hearing impairment went undetected; that is, the three (100%) individuals (5 ears) with hearing loss were accurately identified. Researcheres, seeking to validate the Narrow parameter have also found 100% sensitivity and 92% specificity in NHS using TEOAE1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75..

In addition to high sensitivity and specificity, examination time is an important aspect to be considered, since NHS should be universal 99. Walsh KP, Pasanen EG, McFadden D. Evidence for dynamic cochlear processing in otoacoustic emissions and behavior (A). J. Acoust. Soc. Am. 2009;125:2720. , 3131. Lewis DR, Marone SAM, Mendes BCA, Cruz OLM, Nóbrega M. Comitê multiprofissional em saúde auditiva: COMUSA. Braz. j. otorhinolaryngol. (Impr.) 2010;6(1):121-8.. The present study showed that the mean time for conduction of the TEOAE, using the narrow band parameter - Narrow was 17.5 seconds. A 1994 study compared narrow and broad band parameters and concluded that the mean test time for the Narrow "pass-fail" criterion was smaller. The time since probe adjustment and result appearance was timed and recorded. The authors state that the time for broad band is 70% longer than that for the Narrow - narrow band parameter. Test time for the broad band parameter criterion was 155 seconds and for the Narrow parameter for narrow band criterion test time was 92 seconds. The authors concluded that test time in the "frequency band criterion" is probably longer due to noise in low frequencies. In the Narrow parameter for narrow band stimulus, low frequencies are filtered before they can affect the signal-to-noise ratio results1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75..

However, the time found in this study (17.5 seconds) was also shorter when compared to the 60 to 80 seconds reported by a study conducted in 2002, that also assessed time after probe placement, in 52 newborns without risk factors for hearing impairment, aged up to 48 hours3232. Fuzetti C, Lewis DR . Emissões otoacústicas espontâneas e evocadas por estímulo transiente em recém nascidos. Pró-Fono R Atual Científica. 2003;15(2):189-98..

The high-pass and low-pass filters are tools that are commonly used by screening equipment to decrease examination time, since they increase the signal-to-noise ratio and response reproducibility in TEOAE 1616. Ravazzani P, Tognola G, Grandori F. Optimal band pass filtering of transient- evoked otoacoustic emissions in neonates. Audiology. 1999;38(2):69-74.. Test time using a broad band stimulus is probably longer due to the noise of low frequency bands, since, for the narrow band stimulus - Narrow these frequencies are filtered before they affect the signal-to-noise ratio result1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75..

Techniques such as decreasing the time window should be used in automatic TEOAE equipment in order to favor the response level recording quality of the TEOAE 1010. Vohr BR, White KR, Brancia Maxon A, Johnson MJ. Factors affecting the interpretation of transient-evoked otoacousticm emission results in neonatal hearing screening. Semin Hear. 1993;14:57-72. , 1313. Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75.. Thus, this study's results that showed a considerably short examination time with high sensitivity (100%) and specificity (92%) indicate this criterion's efficacy, making UNHS easier in noisy environments.

New technologies that seek to maximize results and guarantee effectiveness of the procedures in UNHS are being developed. However, there are still few studies regarding the efficiency and efficacy of the narrow band stimulus in NHS procedures using TEOAE, showing the need for new investigations concerning these aspects.

Conclusion

The "pass-fail" criterion using the narrow band stimulus offered acceptable indexes of sensitivity and specificity, proving efficient for use in UNHS programs.

  • 1
    Política nacional de atenção à saúde auditiva. [cited 2006 sep 21]. Disponível em: http://portalsaude.saude.gov.br/portalsaude/
    » http://portalsaude.saude.gov.br/portalsaude/
  • 2
    Hilú MRPB, Zeigelboim BS. O conhecimento, a valorização da triagem auditiva neonatal e a intervenção precoce da perda auditiva. Rev CEFAC. 2007;9(4):563-70.
  • 3
    Soares CP, Marques LR, Flores NGC. Triagem auditiva neonatal: aplicabilidade clínica na rotina dos médicos pediatras neonatologistas. Rev CEFAC. 2008;10(1):110-6.
  • 4
    Dantas MBS, Anjos CAL, Camboim ED, Pimentel MCR. Resultados de um programa de triagem auditiva neonatal em Maceió. Rev. Bras. Otorrinolaringol. 2009;75(1):58-63.
  • 5
    Côrtes-Andrade IF, Bento DV, Lewis DR. Emissões Otoacústicas Evocadas por Estímulo Transiente: protocolos de Triagem Auditiva Neonatal. Rev CEFAC [online]. 2013; 15(3):521-7. Available from: http://www.scielo.br/pdf/rcefac/2012nahead/48-11.pdf
    » http://www.scielo.br/pdf/rcefac/2012nahead/48-11.pdf
  • 6
    Davis A, Bamford J, Stevens J. Performance of neonatal and infant hearing screens: sensitivity and specificity. Br J Audiol. 2001;35(1):3-15.
  • 7
    Joint committee on infant hearing (JCIH). Position statement. 2007 Avaliable from: < http://www.jcih.org/ExecSummFINAL.pdf >. Acesso em: 19/06/2009.
    » http://www.jcih.org/ExecSummFINAL.pdf
  • 8
    Kemp DT. Otoacustic emissions, their oringinin cochlear function, and use. Br. Med. Bull. 2002;63(1):223-41.
  • 9
    Walsh KP, Pasanen EG, McFadden D. Evidence for dynamic cochlear processing in otoacoustic emissions and behavior (A). J. Acoust. Soc. Am. 2009;125:2720.
  • 10
    Vohr BR, White KR, Brancia Maxon A, Johnson MJ. Factors affecting the interpretation of transient-evoked otoacousticm emission results in neonatal hearing screening. Semin Hear. 1993;14:57-72.
  • 11
    Rhoades K, McPherson B, Smyth V, Kei J, Baglioni A. Effects of background noise on click-evoked otoacoustic emissions. Ear Hear. 1998;19(6):450-62.
  • 12
    Norton SJ, Gorga MP, Widen JE, Folsom RC, Sininger Y, Cone-Wesson B et al. Identification of neonatal hearing impairment: evaluation of transient evoked otoacoustic emission, distortion product otoacoustic emission, and auditory brain stem response test performance. Ear Hear. 2000;21(5):508-28.
  • 13
    Brass D, Watkins P, Kemp DT . Assessment of an implementation of a narrow band, neonatal otoacoustic emission screening method. Ear Hear. 1994;15(6):467-75.
  • 14
    Gong Q, Ye D. A study of the reduction of the stimulus artifact using a predictor-subtractor-restorer (PSR) filter. Conf Proc IEEE Eng Med Biol Soc. 2005;2:1185-8.
  • 15
    Glattke TJ, Robinette MS. Transient evoked otoacoustic emissions in population with normal hearing sensitivity. In: Robinette MS, Glattke TJ . Otoacoustic emissions: clinical applications. 3nd. ed. New York: Thieme; 2007. P.87-107.
  • 16
    Ravazzani P, Tognola G, Grandori F. Optimal band pass filtering of transient- evoked otoacoustic emissions in neonates. Audiology. 1999;38(2):69-74.
  • 17
    Driscoll C, Kei J, McPherson B . Outcomes of transient evoked otoacoustic emission testing in 6-year-old school children: a comparison with pure tone screening and tympanometry. Int J Pediatr Otorhinolaryngol. 2001;57(1):67-76.
  • 18
    Vicky Z, McPherson B . A review of otoacoustic emission hearing screening technology. Audiological Medicine. 2008;6(2):100-14.
  • 19
    National center for hearing assessment and management - NCHAM. Issues & evidence: prevalence of congenital hearing loss. [homepage na internet]. Utah, US; 2002. [atualizado em 11abr 2008; [acesso em 23 02 13]. Disponível em <http://www.infanthearing.org/status/unhsstate.html>
    » http://www.infanthearing.org/status/unhsstate.html
  • 20
    Clarke P, Iqbal M, Mitchell S. A comparison of transientevoked otoacoustic emissions and automated auditory brainstem responses for pre-discharge neonatal hearing screening. Int J Audiol. 2003;42(8):443-7.
  • 21
    Pereira PKS, Martins AS, Vieira MR, Azevedo MF. Programa de triagem auditiva neonatal: associação entre perda auditiva e fatores de risco. Pró-Fono R. Atual. Cient. 2007;19(3):267-78.
  • 22
    Freitas VS, Alvarenga KF, Bevilacqua MC, Martinez MAN, Costa OA. Análise crítica de três protocolos de triagem auditiva neonatal. Pró-Fono R. Atual. Cient. 2009;21(3):201-6.
  • 23
    Thornton AR, Marotta N, Kennedy CR. The order of testing effect in otoacoustic emissions and its consequences for sex and ear differences in neonates. Hear Res. 2003;184(1-2):123-30.
  • 24
    Von Specht H, Beyer U. Hearing screening with transient evoked otoacoustic emissions in preterm neonates [Internet]. 2007. Available from: http://www.auditio.com/ierasg/abstract/126.htm
    » http://www.auditio.com/ierasg/abstract/126.htm
  • 25
    Guedes MC, Passos SN, Gomez MVSG, Bento RF. Estudo da reprodutibilidade das emissões otoacusticas em indivíduos normais. Rev. Bras. Otorrinolaringol. 2002;68(1):34-8.
  • 26
    Simonek C, Azevedo MF . Respostas falso-positivas na triagem auditiva neonatal universal: possíveis causas. Rev CEFAC. 2011;13(2):292-8.
  • 27
    Brazelton TB. The manual. In: Brazelton TB . Neonatal Behavioral Assesment Scale. 2. ed. Philadelphia: Spastics International medical Publications, 1984. P. 17-77.
  • 28
    Korres SG, Balatsouras DG, Nikolopoulos T, Korres GS, Ferekidis E. Making universal newborn hearing screening a success. Int J Pediatr Otorhinolaryngol. 2006;70(2):241-6.
  • 29
    Keren R, Helfand M, Homer C, McPhillips H, Lieu TA. Projected cost-effectiveness of statewide universal newborn hearing screening. Pediatrics. 2002;110(5):855-64.
  • 30
    Fletcher RH, Fletcher SW. Prevenção. In: Fletcher R, Fletcher S. Epidemiologia clínica: elementos essenciais. Porto Alegre: Artes Médicas, 2006, 4.ed. P. 179-201.
  • 31
    Lewis DR, Marone SAM, Mendes BCA, Cruz OLM, Nóbrega M. Comitê multiprofissional em saúde auditiva: COMUSA. Braz. j. otorhinolaryngol. (Impr.) 2010;6(1):121-8.
  • 32
    Fuzetti C, Lewis DR . Emissões otoacústicas espontâneas e evocadas por estímulo transiente em recém nascidos. Pró-Fono R Atual Científica. 2003;15(2):189-98.