Acessibilidade / Reportar erro

Reply to: Epistaxis as a complication of high-flow nasal cannula therapy in adults

TO THE EDITOR

We appreciate the comments by Drs Nair and Esquinas to our article,(11 Veiga VC, Silva LM, Sady ER, Maia IS, Cavalcanti AB. Epistaxis as a complication of high-flow nasal cannula therapy in adults. Rev Bras Ter Intensiva. 2022;33(4):640-3.) to which we provide clarifications.

First, according to the institutional protocol, the use of high-flow nasal cannula (HFNC) therapy was contraindicated in patients with predominant mouth breathing, considering its ineffectiveness.(22 Spicuzza L, Schisano M. High-flow nasal cannula oxygen therapy as an emerging option for respiratory failure: the present and the future. Ther Adv Chronic Dis. 2020;11:2040622320920106.)

Second, all seven patients with epistaxis reported in our case series were evaluated and managed by the institution´s otorhinolaryngology team, including an exploration of the nasal cavity utilizing anterior rhinoscopy. Epistaxis was unilateral in five patients and bilateral in two patients. The Epistaxis Severity Score could not be calculated because it is not applicable to isolated acute epistaxis. Among the 7 patients with epistaxis associated with HFNC use, one case was more severe and needed to be treated with an epistaxis-specific device (Rapid Rhyno®). Two other cases were controlled with topical epinephrine. No patients needed chemical or electrical cauterization.

Third, the Vapotherm® technical manual indicates that the temperature of the delivered gas mixture is accurate within plus or minus 2ºC of the programmed temperature.(33 Vapotherm®. Precision Flow® Plus. Technical Service Manual [cited 2022 June 29]. Available from: https://vapotherm.com/docs/english/3101151%20Rev%20C%20PF%20Plus%20Technical%20Service%20Manual.pdf
https://vapotherm.com/docs/english/31011...
) However, the manufacturer states that the temperature to which patients are exposed depends not only on the temperature titration in the device but also on the titrated flow rate and the ambient temperature. In our institution, the ambient temperature is controlled between 20ºC and 22ºC. We did not find a study that evaluated the accuracy of temperature and humidification in different scenarios of flow or ambient temperatures for the HNFC device used in our patients. Additionally, as the use of HFNCs in the patients whose cases were reported was part of regular clinical practice, there were no measurements of the accuracy of the temperature and humidification of the gas mixture delivered to patients.

REFERÊNCIAS

  • 1
    Veiga VC, Silva LM, Sady ER, Maia IS, Cavalcanti AB. Epistaxis as a complication of high-flow nasal cannula therapy in adults. Rev Bras Ter Intensiva. 2022;33(4):640-3.
  • 2
    Spicuzza L, Schisano M. High-flow nasal cannula oxygen therapy as an emerging option for respiratory failure: the present and the future. Ther Adv Chronic Dis. 2020;11:2040622320920106.
  • 3
    Vapotherm® Precision Flow® Plus. Technical Service Manual [cited 2022 June 29]. Available from: https://vapotherm.com/docs/english/3101151%20Rev%20C%20PF%20Plus%20Technical%20Service%20Manual.pdf
    » https://vapotherm.com/docs/english/3101151%20Rev%20C%20PF%20Plus%20Technical%20Service%20Manual.pdf

Publication Dates

  • Publication in this collection
    04 Nov 2022
  • Date of issue
    2022

History

  • Received
    28 June 2022
  • Accepted
    29 June 2022
Associação de Medicina Intensiva Brasileira - AMIB Rua Arminda, 93 - Vila Olímpia, CEP 04545-100 - São Paulo - SP - Brasil, Tel.: (11) 5089-2642 - São Paulo - SP - Brazil
E-mail: rbti.artigos@amib.com.br