Surface modification of recording electrodes

Abstract

Waterborne Polyurethanes (PUs) are a family of polymers that contains urethane linkages synthesized in an aqueous environment and are thus free of organic solvents. Recently, waterborne PUs have been extensively studied for biomedical applications because of their biocompatibility. The present work investigates the following: (1) the impact on electrical performance of electrode materials (platinum and silicon) modified chemically by a layer of waterborne PU, and (2) the behavior of rat cardiac fibroblasts and rat cardiomyocytes when in contact with an electrode surface. Diisocyanate and poly(caprolactone diol) were the main reagents for producing PUs. The electrochemical impedance of the electrode/electrolyte interface was accessed by electrochemical impedance spectroscopy. The cellular viability, proliferation, and morphology changes were investigated using an MTT assay. Cardiomyocyte adherence was observed by scanning electron microscopy. The obtained surface was uniform, flat, and transparent. The film showed good adhesion, and no peeling was detected. The electrochemical impedance decreased over time and was influenced by the ionic permeability of the PU layer. The five samples did not show cytotoxicity when in contact with neonatal rat cells.

Implantable probes; bio-tolerability; surface modification; electrochemical impedance; waterborne polyurethanes


ARTIGO TÉCNICO CIENTÍFICO

Surface modification of recording electrodes

Iaci Miranda PereiraI; Sandhra Maria de CarvalhoII; Rodrigo Lambert OréficeII; Marcelo Bariatto Andrade FontesIII; Lilian Anee Muniz ArantesIV; Núbia Figueiró, Maria de Fátima LeiteIV; Hercules Pereira NevesV

IBrazilian Army Technology Center, Materials Laboratory, LM/CTEx, Brazil

IIDepartment of Metallurgical and Materials Engineering, Federal University of Minas Gerais, Brazil

IIIFaculty of Technology of Sao Paulo, Brazil

IVDepartment of Physiology and Biophysics, Federal University of Minas Gerais, Brazil

VInteruniversity Microelectronics Centre, IMEC, Belgium

Corresponding author Corresponding author: Iaci Miranda Pereira Brazilian Army Technology Center, Materials Laboratory Av. das Américas, 28705, Guaratiba CEP 23020-470, Rio de Janeiro, RJ, Brazil e-mail: iaci@ctex.eb.br

ABSTRACT

Waterborne Polyurethanes (PUs) are a family of polymers that contains urethane linkages synthesized in an aqueous environment and are thus free of organic solvents. Recently, waterborne PUs have been extensively studied for biomedical applications because of their biocompatibility. The present work investigates the following: (1) the impact on electrical performance of electrode materials (platinum and silicon) modified chemically by a layer of waterborne PU, and (2) the behavior of rat cardiac fibroblasts and rat cardiomyocytes when in contact with an electrode surface. Diisocyanate and poly(caprolactone diol) were the main reagents for producing PUs. The electrochemical impedance of the electrode/electrolyte interface was accessed by electrochemical impedance spectroscopy. The cellular viability, proliferation, and morphology changes were investigated using an MTT assay. Cardiomyocyte adherence was observed by scanning electron microscopy. The obtained surface was uniform, flat, and transparent. The film showed good adhesion, and no peeling was detected. The electrochemical impedance decreased over time and was influenced by the ionic permeability of the PU layer. The five samples did not show cytotoxicity when in contact with neonatal rat cells.

Keywords: Implantable probes, bio-tolerability, surface modification, electrochemical impedance, waterborne polyurethanes.

Introduction

Implantable probes for stimulation or recording have received increased attention. Miniaturized electrodes have been implanted to stimulate excitable tissue, detect bioelectric events, and measure chemical substances[1]. In electrical stimulation and recording, electrodes must be sufficiently close to cells and need sufficiently low impedance to maximize the signal to noise ratio[2]. Furthermore, the inflammatory response elicited by the implant should be minimized; therefore, the bio-tolerability of the implant or substrate material is one of the key factors to consider when designing the electrode. For instance, implantable pacemakers and cardioverter-defibrillators are used to treat tachycardia and severe heart failure. These devices use electrodes to deliver electrical impulses to regulate heart rhythm. However, past implantation studies suggest that implanted intravascular electrodes can cause histological changes and death[3].

Waterborne polyurethanes (PUs) are possible candidates for the chemical modification of an electrode surface. In addition to their biocompatibility, PUs have ionized groups attached to their macromolecules that lead to significant improvement in adhesive properties and allow ionic conduction. Figure 1 shows a neural probe coated with a PU layer.


The PU coating may promote cell adhesion and typically has high electrical impedance without direct pathways for electrical stimulation and recording. In this study, we are interested in determining the impact on electrical performance of electrode materials chemically modified by a layer of waterborne PU. The objectives of this study are (1) to assess electrochemical impedance of the electrode/electrolyte interface and (2) to study the behavior of cardiac fibroblasts when in contact with the electrode surface, as cardiac fibroblasts and cardiomyocytes contribute to cardiac development, myocardial structure, cell signaling, and electro-mechanical function in healthy and diseased myocardium[4].

Materials and Methods

Polymer synthesis

Poly(caprolactone diol) (PCL1250 - Mn = 1250 g mol - 1 and PCL2000 - Mn = 2000 g mol - 1), isophorone diisocyanate (IPDI), 2,2-bis(hydroxymethyl) propionic acid (DMPA) and dibutyl tin dilaurate (DBDLT) were obtained from Aldrich (St. Louis, MO, USA). Triethylamine (TEA, 98%) and hydrazine (HZ, 25%) were purchased from Vetec (Rio de Janeiro, RJ, Brazil). These chemicals were employed throughout this work without any previous treatment.

The PU dispersion was synthesized in an aqueous environment. Synthesis procedures for PU are described in detail elsewhere[5]. Using the conventional prepolymer method, PU was synthesized with a 250 mL three-neck glass flask equipped with a heating mantel, a mechanical stirrer, and a thermometer with nitrogen atmosphere. In the first step, prepolymer was obtained by reacting PCL1250 (23.7 wt %), PCL2000 (37.9 wt %), DMPA (3.8 wt %) and IPDI (29.6 wt %) at a 2 NCO/OH ratio for about 3.5 hours. During this time, DBDLT was added twice. The reaction was carried out at 70-75 °C. After cooling to 40 °C, the carboxylic acid groups were neutralized by adding TEA (2.9 wt %). The mixture was then gently stirred for 40 minutes. The mixture was dispersed in water, and PU chains were extended by adding HZ (2.1 wt %) and deionized water to the neutralized prepolymer during high-speed stirring. The mixture was stirred for 60 minutes, ensuring that the reaction was completed. This chemical procedure was very successful in producing PU dispersions in water with about 30 % solid content.

Substrate preparation

Two different sets of substrates, silicon and platinum, were studied. The 25 × 25 mm2 samples were cut from bare standard silicon wafers and standard silicon wafers coated with platinum. Depositing a 200 nm layer of platinum on the silicon wafers using physical vapor deposition produced the platinum-coated wafers.

The substrate surface was modified by spin coating a PU layer on the surface. First, samples were rinsed in deionized water, immersed in alcohol for 10 minutes and dried. An excess amount of PU solution was dispensed on the sample, which was subsequently spun at different speeds for 10 minutes. The sample was then heated to 40 °C on a hotplate for 2 minutes for curing. The scotch tape test was used for the adhesion test; no peeling was found. The platinum samples were named "Pt", and the silicon samples were named "Si". After surface modification, the Pt and Si samples were listed as Pt/PU and Si/PU, respectively.

Thickness measurements

After spin coating, the thickness profile of Si/PU samples was investigated. Measurements were performed at three different sample positions: the center and the two opposite corners. Thickness measurements were executed using a Tencor Instruments Alpha Step 200 surface profilometer.

Laser machine

To stimulate the electrodes on the samples, an array of holes was patterned on the PU layer using laser ablation. This array consisted of a grid of 24 × 24 holes. A custom-built laser set-up integrating a CO2 laser (GSIL Impact SSM 2150) was used to cut the PU film. The sample was mounted on a fixed working table, and the machining process was completed by moving the projection lens at 1 mm/s. Velocity and laser displacement were computer controlled. After surface patterning, the Pt/PU and Si/PU samples were labeled Pt/PU/Grid and Si/PU/Grid, respectively. The holes sizes and distribution were investigated by optical microscopy (Nikon Optiphot 200). Scanning electron microscopy (SME - JEOL 5600LV) was employed to observe the laser cutting effect on the polymer surface.

Electrochemical impedance spectroscopy

Electrochemical impedance spectroscopy was carried out using an electrochemical analyzer (CompactStat, IviumTechnologies). All experiments were performed in standard physiological saline solution (0.9% NaCl). Investigations were performed on substrates spun at 4000 rpm.

In vitro testing

Culture of cardiac fibroblasts

Neonatal cardiac fibroblasts were isolated from the hearts of 3-day-old Wistar rats according to instructions from the biochemical supplier (Worthington Biochemical Corporation, Lakewood, USA). The cells were resuspended in Dulbecco's modified Eagle's medium (DMEM) supplemented with 10 % fetal bovine serum (FBS), 100 units/ml penicillin, 100 units/ml streptomycin and 0.25 g/mL anfotericin-b, all obtained from Sigma (St. Louis, MO, USA). Cardiac fibroblasts were plated into culture dishes and incubated at 37 °C in a 5 % CO2 incubator. The cultured cells were used in experiments on the fourth day of culture[6].

Culture of cardiomyocytes

Neonatal cardiomyocytes were isolated from hearts of 3-day-old Wistar rats according to instructions from the biochemical supplier (Worthington Biochemical Corporation, Lakewood, USA). The cells were cultured in Dulbecco's modified Eagle's medium (DMEM) supplemented with 10 % fetal bovine serum (FBS) from Gibco (Grand Island, NY, USA), 100 units/ml penicillin, 100 mg/mL streptomycin and 0.25 g/mL anfotericin-b, all obtained from Sigma (St. Louis, MO, USA). The cells were plated in 6 well culture plates and incubated at 37 °C in a 5% CO2 incubator. The cultured cells were used in experiments on the second day of culture[6].

Toxicity assay resazurin

Toxicity assays were implemented separately for cultures of cardiac fibroblasts (on the fourth day of culture) and cardiomyocytes (on the second day of culture). The cell cultures were plated (1 X 106 cells/cm3) on the material surface in 6 well culture plates. After 48 hours, the entire medium was aspirated and, in each well, 2 mL of culture medium was replaced with serum. Thereafter, 200 µL of resazurin (0.1 mg/mL; Sigma Aldrich, USA) was added in each well and incubated for 18 hours in an incubator at 37 °C and 5 % CO2. Afterwards, 100 µl of each well were removed and transferred to a 96 well plate. Quantification was performed in a spectrophotometer (ADAP 1.6, Anthos Labtec Instruments) with dual filters (570 nm and 590 nm). The positive control was cells in a PBS (5X)/ trypsin solution (GibcoW, Invitrogen Brasil Ltd., Sao Paulo, SP, Brazil). The negative control was cells on cover glass slides (VWR Scientific, USA). Tests were performed in triplicate (n = 3). The sterilization method was ultraviolet radiation. Tests were performed according to ISO 10993.

Scanning electron microscopy

Si, Si/PU and Pt/PU samples were prepared for SEM. Initially, cardiomyocytes were fixed on wafers with 2.5 % glutaraldehyde in 0.1 M sodium phosphate buffer (pH 7.3) for 2 hours at 8 °C. Afterwards, samples were rinsed three times with PBS and post-fixed twice in osmium tetroxide. Subsequently, samples were dehydrated with ascending grades of ethanol, critical point dried, coated with 15 nm gold and examined using electron microscopy (Quanta 200v FEG FEI).

Results and Discussion

The chemically modified surface obtained by spin coating was uniform, flat, and transparent. Figure 2a presents the film thickness profile obtained at 500 rpm for Si/PU samples. Figure 2b shows the film thickness profile obtained at different spinning speeds for Si/PU samples.


Figure 3a presents a typical sample covered with a PU layer. The picture shows 4 of the holes on the grid. The average diameter of a hole was 251.4 µm ± 14 µm, and they were arranged 717 µm ± 23 µm apart. Figure 3b shows a SEM image of a typical hole and the exposed platinum layer. As observed, laser cutting ablated the PU layer. The zone affected by heating was restrained to a few microns.


Figure 4a presents the electrochemical impedance results (|Z|) for surfaces without modification. As expected, the PU layer acts as an insulator coating whose capacitance alters the impedance spectra, as shown in Figure 4b. As observed in Figure 4c, the array of holes produced impedance values that are consistent with an array of Pt microelectrodes.


Figure 5a shows |Z| measured after different time intervals. In the present study, the ionic conductivity is influenced by saline solution absorption through the polymer layer. Due to PU´s low ionic conductivity, investigations were performed on the thinnest coating, the 1.1±0.05 µm thick substrate. Lower ionic conductivity is estimated for thicker layer. Although, the saline solution depth of penetration could not be accurately determined it is possible to notice that, as time elapses, the penetration of ions toward the PU layer increases. As a consequence, |Z| values drop. After 24 hours, at the low frequency range (<103 Hz), |Z| values drop 20-fold, as shown in Figure 5a. Figure 5b shows |Z| of the Pt/PU/Grid, where no appreciable impedance change was observed.


The mitochondrial activity of cells exposed to Si, Pt, Si/PU, Pt/PU or Pt/PU/Grid was assessed by resazurin analysis. This assay is specifically used to evaluate mitochondrial function and cell viability. Samples did not exhibit cytotoxicity when in contact with cardiac fibroblasts or cardiomyocytes, as indicated in Figure 6. After 48 hours, the cardiac fibroblasts exposed to Si, Pt, Si/PU and Pt/PU/Grid samples showed no significant differences compared to the control group. However, compared to the control group, a 15 % increase in cell viability for cardiac fibroblasts was observed for cells in contact with the Pt/PU sample, Figure 6a. Furthermore, an increase in cardiomyocyte cell viability of 113 % compared to the control group was observed when in contact with the Pt/PU samples (Figure 6b). We also observed a 50 % increase in cell viability for cardiomyocytes in contact with the Si/PU sample, Figure 6b. The samples Si, Pt and Pt/PU/Grid showed no significant effect on cell viability when compared to the control group. Experiments were performed with n = 3 (p < 0.05 statistical analysis: One way/ANOVA/Bonferroni/GraphPad Prism).


Scanning electron microscopy of neonatal rat cardiomyocytes in contact with the samples of Si, Si/Pu and Pt/PU showed cell adhesion on the material surface (Figure 7).


Conclusion

PU films caused an expected increase in impedance. However, over a relatively short period of time, there was a marked decrease in impedance, which is evidence of the ionic permeability of the PU layer. The five samples did not show cytotoxicity when in contact with neonatal rat cells. We observed an increase in cell viability when in contact with Si or Pt samples after surface modification. The viability of neonatal rat cardiomyocytes and cardiac fibroblasts was tested by the resazurin assay based on the mitochondria activity of the cell, according to ISO 10993. We also observed good adhesion of cardiomyocytes on Si or Pt samples after surface modification. These results show that treating the sample with PU to increase biocompatibility only has an appreciable detrimental effect on electrical performance within the first day or so of implantation. This makes PU a suitable candidate for the surface modification of implantable electrodes.

Acknowledgements

The authors acknowledge financial support from the following institutions: National Council for Scientific and Technological Development (CNPq) and the State of Minas Gerais Research Foundation (FAPEMIG).

Received: 05/08/13

Revised: 07/05/13

Accepted: 09/05/13

  • 1. Geddes, L. A. & Roeder, R. - Ann Biomed Eng., 31, p.879 (2003). PMid:12971619. http://dx.doi.org/10.1114/1.1581292
  • 2. Thanawala, S.; Palyvoda, O.; Georgiev, D. G.; Khan, S. P.; Al-Homoudi, I. A. & Auner G. - J. Mater. Sci.: Mater. Med., 18, p.1745 (2007). PMid:17483885. http://dx.doi.org/10.1007/s10856-007-3054-1
  • 3. Dvorak, P.; Novak, M.; Kamaryt, P.; Slana, B.; Lipoldova, J. & Dvorak, P. - Europace, 14, p.117 (2012). PMid:21893512. http://dx.doi.org/10.1093/europace/eur274
  • 4. Camelliti, P.; Borg, T. K. & Kohl, P. - Cardiovasc. Res., 65, p.40 (2005). PMid:15621032. http://dx.doi.org/10.1016/j.cardiores.2004.08.020
  • 5. Pereira, I. M. & Oréfice, R. L. - Polymer, 51, p.1744 (2010). http://dx.doi.org/10.1016/j.polymer.2010.02.037
  • 6
    Guide for the Care and Use of Laboratory Animals. DHHS [NIH], US Dept. of Health and Human Services, Bethesda, p. 85 (1985).
  • Corresponding author:

    Iaci Miranda Pereira
    Brazilian Army Technology Center, Materials Laboratory
    Av. das Américas, 28705, Guaratiba
    CEP 23020-470, Rio de Janeiro, RJ, Brazil