Endothelial keratoplasty: Descemet stripping (DSEK) using TAN EndoGlideTM device: case series

PURPOSE: To report the results of Descemet stripping endothelial keratoplasty (DSEK) using the TAN EndoGlideTM device to facilitate the insertion of the endothelial membrane. METHODS: Prospective clinical study that included nine patients presenting corneal edema secondary to endothelial dysfunction. Best corrected visual acuity, refraction, central corneal thickness, endothelial cell density and complications were analyzed after a six-month follow-up. RESULTS: There was a significant improvement in the corneal edema and visual acuity in 7 patients (77.78%). The best corrected visual acuity ranged between 20/40 and 20/200. The average density of endothelial cells in six months varied between 1,305 cells/mm² and 2,346 cells/mm² with an average loss of 33.14% cells. Detachment of part of the graft was observed in one eye (11.11%) and primary failure of the endothelial transplantation occurred in 2 eyes (22.22%). CONCLUSION: The device TAN EndoGlideTM facilitates the introduction of the graft in Descemet stripping endothelial keratoplasty.

Corneal transplantation; Corneal edema; Corneal endothelial cell loss; Descemet membrane; Descemet stripping endothelial keratoplasty