Flash glucose monitoring system in special situations

ABSTRACT

The management of diabetes mellitus (DM) requires maintaining glycemic control, and patients must keep their blood glucose levels close to the normal range to reduce the risk of microvascular complications and cardiovascular events. While glycated hemoglobin (A1C) is currently the primary measure for glucose management and a key marker for long-term complications, it does not provide information on acute glycemic excursions and overall glycemic variability. These limitations may even be higher in some special situations, thereby compromising A1C accuracy, especially when wider glycemic variability is expected and/or when the glycemic goal is more stringent. To attain adequate glycemic control, continuous glucose monitoring (CGM) is more useful than self-monitoring of blood glucose (SMBG), as it is more convenient and provides a greater amount of data. Flash Glucose Monitoring (isCGM /FGM) is a widely accepted option of CGM for measuring interstitial glucose levels in individuals with DM. However, its application under special conditions, such as pregnancy, patients on hemodialysis, patients with cirrhosis, during hospitalization in the intensive care unit and during physical exercise has not yet been fully validated. This review addresses some of these specific situations in which hypoglycemia should be avoided, or in pregnancy, where strict glycemic control is essential, and the application of isCGM/FGM could alleviate the shortcomings associated with poor glucose control or high glycemic variability, thereby contributing to high-quality care.

Keywords
Diabetes mellitus; blood glucose self-monitoring; liver cirrhosis; renal dialysis; pregnancy

INTRODUCTION

The management of diabetes mellitus (DM) primarily requires effective glycemic control since reductions of glycemic variability can minimize the micro and macrovascular complications associated with the disease (11 Hirsch IB. Glycemic Variability and Diabetes Complications: Does It Matter? Of Course It Does! Diabetes Care. 2015;38(8):1610-4.). Glycated hemoglobin (A1c) remains the reference test to monitor the glucose level of patients with DM and estimate the average blood glucose levels of two to three months (22 Garber AJ, Handelsman Y, Grunberger G, Einhorn D, Abrahamson MJ, Barzilay JI, et al. Consensus Statement by the American Association of Clinical Endocrinologists and American College of Endocrinology on the Comprehensive Type 2 Diabetes Management Algorithm – 2020 Executive Summary. Endocr Pract. 2020;26(1):107-39.). However, this measure fails to consider fluctuations in blood glucose levels during the day and fails to detect acute events of hypoglycemia or postprandial hyperglycemia. Moreover, the method is unreliable in measuring blood glucose levels during anemia, in pregnant patients, and in many other situations (33 Beyond A1C Writing Group. Need for Regulatory Change to Incorporate Beyond A1C Glycemic Metrics. Diabetes Care. 2018;41(6):e92-4.,44 Bry L, Chen PC, Sacks DB. Effects of Hemoglobin Variants and Chemically Modified Derivatives on Assays for Glycohemoglobin. Clin Chem. 2001;47(2):153-63.).

Self-monitoring of blood glucose (SMBG), which is usually performed with finger-prick blood samples, though useful in the management of the disease, does not provide continuous data because the values are not obtained in a timely manner, depending upon the patient's decision to test (55 Shi T, Li D, Li G, Zhang Y, Xu K, Lu L. Modeling and Measurement of Correlation between Blood and Interstitial Glucose Changes. J Diabetes Res. 2016;2016:1-9.,66 Koschinsky T, Heinemann L. Sensors for glucose monitoring: technical and clinical aspects. Diabetes Metab Res Rev. 2001;17(2):113-23.). Various factors, such as pain and inconvenience, lead to lower evaluations frequency and difficulties in diagnosing nocturnal hypoglycemia (77 Dunn TC, Xu Y, Hayter G, Ajjan RA. Real-world flash glucose monitoring patterns and associations between self-monitoring frequency and glycaemic measures: A European analysis of over 60 million glucose tests. Diabetes Res Clin Pract. 2018;137:37-46.,88 Heinemann L. Finger Pricking and Pain: A Never Ending Story. J Diabetes Sci Technol. 2008;2(5):919-21.).

CONTINUOUS GLUCOSE MONITORING (CGM)

The FreeStyle Libre™ Flash Glucose Monitoring System (isCGM/FGM)(Abbott Diabetes Care, Alameda, CA) is an interstitial continuous glucose monitoring (CGM) system that uses a glucose sensor inserted into the skin over the triceps and a portable reader, which people can use to scan their blood glucose level. The sensor automatically measures glucose every minute, storing the values in 15-minutes intervals (99 Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787-94.). It is factory-calibrated and requires no fingerstick for calibration. However, a confirmation with capillary blood glucose (CBG) is recommended in the event of sensor-reported hypoglycemia, during rapidly changing glucose concentrations, in the first 24 hours after sensor insertion, and when no correspondence is identified between symptoms and interstitial glucose (1010 Leelarathna L, Wilmot EG. Flash forward: a review of flash glucose monitoring. Diabet Med. 2018;35(4):472-82.).

Thus, it is critical to understand that each physiological compartment follows different dynamics, and several factors cause concentration disparities between blood and interstitial fluid (ISF) glucose levels. These factors include the distribution of glucose between blood vessels and subcutaneous tissue, glucose permeability, blood flow, and the release of pancreatic hormones such as insulin and glucagon. Moreover, a physiological time lag between blood and ISF glucose levels is remarked. When such levels are changing rapidly, their measurements differ considerably (55 Shi T, Li D, Li G, Zhang Y, Xu K, Lu L. Modeling and Measurement of Correlation between Blood and Interstitial Glucose Changes. J Diabetes Res. 2016;2016:1-9.,66 Koschinsky T, Heinemann L. Sensors for glucose monitoring: technical and clinical aspects. Diabetes Metab Res Rev. 2001;17(2):113-23.).

Time in range (TIR) is the most relevant parameter to assess glucose control, which indicates the percentage of time that a person spends with their interstitial glucose levels in the target range, usually defined between 70 and 180 mg/dL (Figure 1). Moreover, it is possible to evaluate the average glucose, glucose management index, glucose variability, time in hypoglycemia (time below range, TBR) and hyperglycemia (time above range), and the number of reads that the user performed per day (1111 Danne T, Nimri R, Battelino T, Bergenstal RM, Close KL, DeVries JH, et al. International Consensus on Use of Continuous Glucose Monitoring. Diabetes Care. 2017;40(12):1631-40.).

Figure 1
Example of an ambulatory glucose profile report, a standardized single-page report developed by the International Diabetes Center and adopted by most of the CGM device manufacturers, including 14 days of data. The middle graph shows the median glucose over 24 hours and its variation (5th, 25th, 75th, and 95th percentile), while the bottom graph shows daily glucose profiles. The stacked bar graph at the top right corner displays the percentage of time spent within, below, and above the target range. The additional table at the top left corner describes glucose statistics and targets.

The primary goal for effective and safe glucose control for an individual with DM is to increase the TIR to over 70% while reducing the TBR to less than 4, thereby including the time in level 2 hypoglycemia (below 54 mg/dL) to less than 1%. Thus, CGM-based glycemic targets can be personalized to address specific needs of special diabetes populations (1212 Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593-603.).

The association between TIR and microvascular complications was reported in previous studies, where individuals with advanced diabetic retinopathy and chronic kidney disease spent less time on target, while the highest TIR was associated with lower frequency of these complications (1313 Beck RW, Bergenstal RM, Riddlesworth TD, Kollman C, Li Z, Brown AS, et al. Validation of Time in Range as an Outcome Measure for Diabetes Clinical Trials. Diabetes Care. 2019;42(3):400-5.,1414 Lu J, Ma X, Zhou J, Zhang L, Mo Y, Ying L, et al. Association of Time in Range, as Assessed by Continuous Glucose Monitoring, With Diabetic Retinopathy in Type 2 Diabetes. Diabetes Care. 2018;41(11):2370-6.). In randomized clinical trials, a TIR of 70% and 50% corresponded to an A1c of approximately 7% and 8%, respectively. Meanwhile, an 8% to 10% increase in TIR was associated with a 0.5% decrease in A1c (1515 Beck RW, Bergenstal RM, Cheng P, Kollman C, Carlson AL, Johnson ML, et al. The Relationships Between Time in Range, Hyperglycemia Metrics, and HbA1c. J Diabetes Sci Technol. 2019;13(4):614-26.,1616 Vigersky RA, McMahon C. The Relationship of Hemoglobin A1C to Time-in-Range in Patients with Diabetes. Diabetes Technol Ther. 2019;21(2):81-5.).

Recently, the new version of FreeStyle Libre 2TM has been launched in some countries; BluetoothTM technology has been integrated into the system, where an optional alarm is available to alert patients in case of high and low glucose levels (1717 Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 Diabetes. Diabetes Ther. 2017;8(3):573-86.). FreeStyle Libre 3TM is already approved in Europe, and it offers similar features to that of the previous version plus some additional features, such as continuous real-time glucose readings that automatically delivers glucose level to a smartphone every minute through a sensor. The FreeStyle Libre ProTM is a device that uses the same system, but it is blind to the patient. Healthcare professionals applied the sensor, scanned with a reader after 14 days, and downloaded the results stored in the sensor.

Bailey and cols. compared capillary and venous glucose with ISF glucose using the isCGM/FGM first-generation system in 2015. The overall mean absolute relative difference (MARD) was 11.4%. In zones A and B of the consensus error grid, the percentage of sensor results was 99.7%. Sensor accuracy was not affected by factors such as body mass index, age, type of diabetes, clinical site, insulin administration, or A1c (99 Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787-94.).

Another study evaluated FreeStyle Libre 2TM performance to compare plasma venous blood glucose. The updated system demonstrated improved analytical accuracy performance across the dynamic range during the 14-day sensor wear period, with a 9.2% overall MARD and 9.7% for the pediatric population (1818 Alva S, Bailey T, Brazg R, Budiman ES, Castorino K, Christiansen MP, et al. Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With Advanced Algorithm in Pediatric and Adult Population With Diabetes. J Diabetes Sci Technol. 2020;193229682095875.).

In some special situations, frequent glucose monitoring is crucial to avoid complications and SMBG is a barrier to optimal glucose control. Recent studies show a positive correlation between the frequency of blood glucose assessments and glycemic control (99 Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787-94.,1717 Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 Diabetes. Diabetes Ther. 2017;8(3):573-86.,1919 Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016;388(10057):2254-63.2121 Calliari LEP, Krakauer M, Vianna AGD, Ram Y, Barbieri DE, Xu Y, et al. Real-world flash glucose monitoring in Brazil: can sensors make a difference in diabetes management in developing countries? Diabetol Metab Syndr. 2020;12:3.).

The main advantages of the isCGM/FGM are its convenience for patients and the greater amount of data available from SMBG (2222 Al Hayek AA, Robert AA, Al Dawish MA. Differences of FreeStyle Libre Flash Glucose Monitoring System and Finger Pricks on Clinical Characteristics and Glucose Monitoring Satisfactions in Type 1 Diabetes Using Insulin Pump. Clin Med Insights Endocrinol Diabetes. 2019;12:117955141986110.). When comparing isCGM/FGM with SMBG, TBR was reduced by 38% in patients with type 1 diabetes mellitus (T1DM) in the IMPACT study and 43% in individuals with type 2 diabetes mellitus (T2DM) in the REPLACE study. In both studies, A1c was similar between groups, revealing a significant improvement in quality of life in groups using isCGM/FGM (1717 Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 Diabetes. Diabetes Ther. 2017;8(3):573-86.,1919 Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016;388(10057):2254-63.).

However, some shortcomings of isCGM/FGM include the lower accuracy for low glucose values and lack of a sound alarm to alert patients about hypoglycemia and hyperglycemia. However, this issue has been recently solved with the launching of a new version. Furthermore, the device offers no option to connect to an insulin pump, requiring an exclusive reader or an application in a compatible smartphone (1010 Leelarathna L, Wilmot EG. Flash forward: a review of flash glucose monitoring. Diabet Med. 2018;35(4):472-82.).

The present review discusses the benefits of isCGM/FGM in diagnosing poor glucose control or high glycemic variability, thereby enhancing effective decisions regarding the treatment of diabetes. In addition, the study discusses specific situations in which hypoglycemia is avoided or strict glycemic control is vital, in which the current metrics are inadequate (not mentioned). Many studies evaluated Libre 1 sensors.

In cases where Libre 2 or Libre Pro sensors were used, this information was included in the text.

PREGNANCY

The relationship between hyperglycemia during pregnancy and adverse outcomes for mother and fetus is well established in the literature (2323 Lepercq J, Le Ray C, Godefroy C, Pelage L, Dubois-Laforgue D, Timsit J. Determinants of a good perinatal outcome in 588 pregnancies in women with type 1 diabetes. Diabetes Metab. 2019;45(2):191-6.). These complications include spontaneous abortion, preterm births, large for gestational age (LGA), and neonatal hypoglycemia (2323 Lepercq J, Le Ray C, Godefroy C, Pelage L, Dubois-Laforgue D, Timsit J. Determinants of a good perinatal outcome in 588 pregnancies in women with type 1 diabetes. Diabetes Metab. 2019;45(2):191-6.). In women with T1DM and T2DM, tight glucose control is vital from the pre-pregnancy to the post-pregnancy period to reduce fetal and maternal risks (2424 Buhary B, Almohareb O, Aljohani N, Alzahrani S, Elkaissi S, Sherbeeni S, et al. Glycemic control and pregnancy outcomes in patients with diabetes in pregnancy: A retrospective study. Indian J Endocrinol Metab. 2016;20(4):481.,2525 Ringholm L, Mathiesen ER, Kelstrup L, Damm P. Managing type 1 diabetes mellitus in pregnancy – from planning to breastfeeding. Nat Rev Endocrinol. 2012;8(11):659-67.).

However, A1c is unreliable for assessing glucose control during pregnancy owing to physiological changes that compromise its measurement, which include an increase in production of red blood cells, increase of half-life of cells, decrease in the affinity of hemoglobin to glucose, and iron deficiency (2626 Nielsen LR, Ekbom P, Damm P, Glumer C, Frandsen MM, Jensen DM, et al. HbA1c Levels Are Significantly Lower in Early and Late Pregnancy. Diabetes Care. 2004;27(5):1200-1.,2727 Herranz L, Saez-de-Ibarra L, Grande C, Pallardo LF. Non-Glycemic-Dependent Reduction of Late Pregnancy A1C Levels in Women With Type 1 Diabetes. Diabetes Care. 2007;30(6):1579-80.).

The pregnancy affects glycemic control and the management of diabetes since insulin resistance decreases in the first trimester and increases in the second half of pregnancy (2828 García-Patterson A, Gich I, Amini SB, Catalano PM, de Leiva A, Corcoy R. Insulin requirements throughout pregnancy in women with type 1 diabetes mellitus: three changes of direction. Diabetologia. 2010;53(3):446-51.). The risk of severe hypoglycemia is a barrier to maintaining strict glycemic control. Hypoglycemia can be five times more frequent at the beginning of pregnancy than in the pre-gestational period in women with T1DM (2929 Ringholm L, Pedersen-Bjergaard U, Thorsteinsson B, Damm P, Mathiesen ER. Hypoglycaemia during pregnancy in women with Type 1 diabetes. Diabet Med. 2012;29(5):558-66.).

Women with gestational diabetes mellitus (GDM) face a particular challenge. Thus, they must understand disease management, which includes techniques to monitor and control blood glucose, the effects of diet, the effects of physical exercise, and medication on their glucose. They must optimize their glucose control within the pregnancy time frame to minimize complications (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.).

SMBG should be performed at least four to eight times a day to reach the rigorous glycemic targets in pregnancy. Moreover, frequent tests are critical for the patient on intensive insulin treatment; however, conducted less frequently when on diet therapy or a basal insulin regimen (3131 National Institute for Health and Care Excellence (UK). Diabetes in pregnancy: management from preconception to the postnatal period [Internet]. London; 2015 [cited 2020 Apr 1]. (National Institute for Health and Care Excellence: Clinical Guidelines). Available from: http://www.ncbi.nlm.nih.gov/books/NBK555331/). Nonetheless, SMBG may fail to capture transient glucose excursions, as the values are assessed intermittently, hindering proper control (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.).

Continuous glucose monitoring provides additional data without discomfort, which has been attributed to SMBG. The CONCEPTT (Continuous Glucose Monitoring in Women With Type 1 Diabetes in Pregnancy Trial) was a multicenter randomized controlled trial involving 215 pregnant women with T1DM. The study reported improved neonatal outcomes with a 5% increase in TIR among the real-time CGM (rtCGM) group (3232 Feig DS, Donovan LE, Corcoy R, Murphy KE, Amiel SA, Hunt KF, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial. Lancet. 2017;390(10110):2347-59.). Scott and cols. analyzed the data from the CONCEPTT study and revealed that the rt-CGM reduced fetal glucose exposure, especially during the day, suggesting that it was easier to monitor the effect of feeding with rtCGM. The strategy assists in preventing and managing fluctuations in glucose levels (3333 Scott EM, Feig DS, Murphy HR, Law GR. Continuous Glucose Monitoring in Pregnancy: Importance of Analyzing Temporal Profiles to Understand Clinical Outcomes. Diabetes Care. 2020;43(6):1178-84.).

Kristensen and cols. observed that a 5% to 7% increase in TIR during the second and third trimesters of pregnancy in women with T1DM was associated with a lower risk of LGA newborns and with a decrease in neonatal complications (3434 Kristensen K, Ögge LE, Sengpiel V, Kjölhede K, Dotevall A, Elfvin A, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes: an observational cohort study of 186 pregnancies. Diabetologia. 2019;62(7):114353.). Murphy and cols. reported a positive association between glycemic control and the outcomes of pregnant women with T1DM and T2DM (3535 Murphy HR, Rayman G, Lewis K, Kelly S, Johal B, Duffield K, et al. Effectiveness of continuous glucose monitoring in pregnant women with diabetes: randomised clinical trial. BMJ. 2008;337:a1680.). The same benefits were observed in a recent systematic review that focused on the effect of CGM on GDM (3636 Yu F, Lv L, Liang Z, Wang Y, Wen J, Lin X, et al. Continuous Glucose Monitoring Effects on Maternal Glycemic Control and Pregnancy Outcomes in Patients With Gestational Diabetes Mellitus: A Prospective Cohort Study. J Clin Endocrinol Metab. 2014;99(12):4674-82.).

Despite the positive outcomes presented by the recent CONCEPTT study, over 80% of the women experienced frustrations with CGM, such as connectivity issues, alarms, and calibration errors. Moreover, 48% of participants experienced skin reactions, such as bleeding, erythema, and discomfort following sensor use (3232 Feig DS, Donovan LE, Corcoy R, Murphy KE, Amiel SA, Hunt KF, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial. Lancet. 2017;390(10110):2347-59.). Other reported barriers to the use of CGM include technical challenges, calibration, skin irritation, frequent alarms (especially during sleep), accuracy, and cost, which discourage its use (3737 Alfadhli E, Osman E, Basri T. Use of a real time continuous glucose monitoring system as an educational tool for patients with gestational diabetes. Diabetol Metab Syndr. 2016;8(1):48.,3838 Secher AL, Madsen AB, Ringholm L, Barfred C, Stage E, Andersen HU, et al. Patient satisfaction and barriers to initiating real-time continuous glucose monitoring in early pregnancy in women with diabetes: CGM in early pregnancy. Diabet Med. 2012;29(2):272-7.).

Thus, isCGM/FGM is more acceptable by pregnant women than other CGM devices owing to its simplicity of handling the device, the absence of alarms, or the possibility of adjusting personalized glucose limits (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.). Moreover, isCGM/FGM was recently approved for pregnant women, as it targets 70% TIR between 63-140 mg/dL and < 4% TBR (1212 Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593-603.).

Scott and cols. conducted a multicenter study to evaluate the accuracy and safety of FreeStyle LibreTM, used to compare the control of 74 hyperglycemic pregnant women (T1DM, n = 24, T2DM, n = 11, and GDM, n = 39) with SMBG values. The results demonstrated good safety and acceptable reliability, compared to that reported in the non-pregnant population (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.).

In another study, Sola-Gazagnes and cols. evaluated a discrepancy between isCGM/FGM results and SMBG based on the outcome of the study of 33 pregnant women. Although close, the results of the sensor underestimated those of the SMBG, in which 25% to 35% of management choices would have diverged if based on isCGM/FGM than SMBG (3939 Sola-Gazagnes A, Faucher P, Jacqueminet S, Ciangura C, Dubois-Laforgue D, Mosnier-Pudar H, et al. Disagreement between capillary blood glucose and flash glucose monitoring sensor can lead to inadequate treatment adjustments during pregnancy. Diabetes Metab. 2020;46(2):158-63.).

Clinical studies have compared mostly pre-prandial capillary glucose measurements, which is a limitation. Despite recommendations for further studies, the use of isCGM/FGM can play an important role in improving glycemic control and the quality of life of pregnant women with hyperglycemia owing to its greater acceptability and simplicity compared to other CGM devices. However, the concomitant use of capillary glycemia should be recommended in association with isCGM/FGM for pregnant women with DM, mainly to confirm glucose values close to the lower limit (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.).

Further studies should investigate whether improved glycemic control and pregnancy outcomes for the mother and baby can be achieved with prolonged use of isCGM/FGM during pregnancy (3030 Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.).

DIALYSIS

The prevalence of DM in recent decades is the main factor responsible for the substantial global increase in end-stage renal disease (ESRD). Currently, more than 3 million people worldwide are estimated to be receiving treatment for kidney failure, and the number is predicted to reach more than 5 million by 2035 (4040 Liyanage T, Ninomiya T, Jha V, Neal B, Patrice HM, Okpechi I, et al. Worldwide access to treatment for end-stage kidney disease: a systematic review. Lancet. 2015;385(9981):1975-82.). Chronic kidney disease, especially ESRD, causes disturbances of glucose homeostasis in subjects with and without diabetes (4141 Schmitz O, Ørskov L, Lund S, Møller N, Christiansen JS, Ørskov H. Glucose Metabolism in Chronic Renal Failure with Reference to GH Treatment of Uremic Children. J Pediatr Endocrinol Metab. 1993;6(1):53-9.). Since the kidney is a relevant site for glucose and insulin metabolism, glycemic control is particularly challenging in ESRD patients with diabetes, owing to a higher risk of hypo and hyperglycemia (4242 Abe M, Kaizu K, Matsumoto K. Plasma Insulin is Removed by Hemodialysis: Evaluation of the Relation Between Plasma Insulin and Glucose by Using a Dialysate With or Without Glucose. Ther Apher Dial. 2007;11(4):280-7.).

Although no evidence-based guideline is available for glycemic targets for hemodialyzed patients with T2DM, adequate glycemic control in those populations seems to be a predictor of survival (4343 Morioka T, Emoto M, Tabata T, Shoji T, Tahara H, Kishimoto H, et al. Glycemic Control Is a Predictor of Survival for Diabetic Patients on Hemodialysis. Diabetes Care. 2001;24(5):909-13.). The Dialysis Outcomes and Practice Patterns Study included 9201 hemodialysis (HD) patients with T1DM or T2DM. Their results showed a U-shaped association between A1c and mortality, with the lowest mortality at A1c levels of 7% to 7.9%, thereby increasing progressively for either lower or higher A1c levels (4444 Ramirez SPB, McCullough KP, Thumma JR, Nelson RG, Morgenstern H, Gillespie BW, et al. Hemoglobin A1c Levels and Mortality in the Diabetic Hemodialysis Population: Findings from the Dialysis Outcomes and Practice Patterns Study (DOPPS). Diabetes Care. 2012;35(12):2527-32.). Despite being challenging, adequate glycemic control to reduce glycemic variations should be targeted in this population (4545 Yoo DE, Park JT, Oh HJ, Kim SJ, Lee MJ, Shin DH, et al. Good Glycemic Control Is Associated with Better Survival in Diabetic Patients on Peritoneal Dialysis: A Prospective Observational Study. PLoS One. 2012;7(1):e30072.).

In patients undergoing renal replacement therapy (RRT), the accuracy of A1c as a marker of glycemic control may be impaired by the altered half-life of red blood cells, erythropoietin therapy, anemia, uremic environment, and frequent blood transfusions, resulting in lowering A1c values (4646 Gai M, Merlo I, Dellepiane S, Cantaluppi V, Leonardi G, Fop F, et al. Glycemic Pattern in Diabetic Patients on Hemodialysis: Continuous Glucose Monitoring (CGM) Analysis. Blood Purif. 2014;38(1):68-73.). The latest Kidney Improving Global Outcomes guidelines highlight the inaccuracy of A1c, suggesting using continuous monitoring devices as an alternative (4747 de Boer IH, Caramori ML, Chan JCN, Heerspink HJL, Hurst C, Khunti K, et al. KDIGO 2020 Clinical Practice Guideline for Diabetes Management in Chronic Kidney Disease. Kidney Int. 2020;98(4):S1-115.).

CGM has emerged as a promising tool in glycemic control for patients with DM and undergoing the RRT. Although the available evidence about the management is currently scarce (4848 Divani M, Georgianos PI, Didangelos T, Iliadis F, Makedou A, Hatzitolios A, et al. Comparison of Glycemic Markers in Chronic Hemodialysis Using Continuous Glucose Monitoring. Am J Nephrol. 2018;47(1):21-9.). In recent studies, CGM has proved useful in detecting asymptomatic hypoglycemia and glycemic variations, as well as in evaluating the effect of some medications in patients undergoing HD (4949 Jin Y, Su X, Yin G, Xu X, Lou J, Chen J, et al. Blood glucose fluctuations in hemodialysis patients with end stage diabetic nephropathy. J Diabetes Complications. 2015;29(3):395-9.,5050 Jung HS, Kim HI, Kim MJ, Yoon JW, Ahn HY, Cho YM, et al. Analysis of Hemodialysis-Associated Hypoglycemia in Patients with Type 2 Diabetes Using a Continuous Glucose Monitoring System. Diabetes Technol Ther. 2010;12(10):801-7.).

Since isCGM/FGM does not require calibration and has a longer useful life, it could be an adequate option for improving the quality of life and facilitating the treatment of DM individuals with ESRD. Flash glucose monitoring with FreeStyle Libre ProTM was shown to be acceptable by Yajima and cols. Parkes’ error grid analysis against SMBG showed that 49.0% and 51.0% of ISF glucose levels fell into zones A and B, respectively using isCGM/FGM. However, for isCGM/FGM, MARD against SMBG was significantly higher than that of CGM (19.5% vs. 8.1%, P <.0001) (5151 Yajima T, Takahashi H, Yasuda K. Comparison of Interstitial Fluid Glucose Levels Obtained by Continuous Glucose Monitoring and Flash Glucose Monitoring in Patients With Type 2 Diabetes Mellitus Undergoing Hemodialysis. J Diabetes Sci Technol. 2020;14(6):1088-94.). Some studies reported similar results, revealing that isCGM/FGM average levels of 5 to 9.2 mg/dL were lower than the average of SMBG in patients with DM not undergoing HD (5252 Kumagai R, Muramatsu A, Fujii M, Katakura Y, Ito K, Fujie K, et al. Comparison of glucose monitoring between Freestyle Libre Pro and iPro2 in patients with diabetes mellitus. J Diabetes Investig. 2019;10(3):851-6.,5353 Ólafsdóttir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, Skrtic S, et al. A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2017;19(3):164-72.). Possible explanations for the results were the effect of the edema in the arm and the abdomen, where sensors were placed to record isCGM/FGM and CGM, respectively. In addition, hematocrit may affect blood glucose meter performance in these patients, as low hematocrit values result in high readings.

Hissa and cols. conducted a 3-week prospective study to compare capillary and ISF glucose in patients with T2DM undergoing dialysis, which was measured by FreeStyle LibreTM. Results showed that MARD values were between 16.5 and 19.0% in the first week. In the second week, the MARD values ranged between 25.3 and 28,8%. Regarding the Clarke and Parkes error grid, 90.3% of patients were in Zone AB of Parkes and 89.6% of Clarke. The lower frequency in the AB zones could be partly because they used only measurements taken during dialysis, when there is manipulation of fluid volume and greater probability of finding discrepancies between capillary and ISF measurements (5454 Hissa MRN, Hissa PNG, Guimarães SB, Hissa MN. Use of continuous glucose monitoring system in patients with type 2 mellitus diabetic during hemodialysis treatment. Diabetol Metab Syndr. 2021;13(1):104.).

Another study focusing on HD patients with T2DM revealed higher estimated A1c (eA1c) from using glycated albumin, BMI, and hemoglobin than eA1c using isCGM/FGM, particularly in patients with decreased BMI (5555 Ushigome E, Matsusaki S, Watanabe N, Hashimoto T, Nakamura N, Fukui M. Critical discrepancy in blood glucose control levels evaluated by glycated albumin and estimated hemoglobin A1c levels determined from a flash continuous glucose monitoring system in patients with type 2 diabetes on hemodialysis. J Diabetes Investig. 2020;11(6):1570-4.).

A pilot study of 10 patients undergoing HD and using FreeStyle Libre ProTM to record the presence of hypoglycemia revealed glucose levels of less than 70 mg/dL in 90% of patients. However, the glucose levels of 4 out of 10 were less than 55 mg/dL during the dialysis period. All episodes were asymptomatic, which is more dangerous and difficult to detect (5656 Javherani R, Purandare V, Bhatt A, Kumaran S, Sayyad M, Unnikrishnan A. Flash glucose monitoring in subjects with diabetes on hemodialysis: A pilot study. Indian J Endocrinol Metab. 2018;22(6):848.).

Mild or severe hypoglycemic episodes are known to be associated with an increased risk of cardiovascular events, hospitalization, and mortality. ESRD is often associated with neuropathy and impairing the perception of hypoglycemia in patients with diabetes. Although their risks can be reduced using fluid enriched with glucose, glycemic patterns are still difficult to predict. However, the application of isCGM/FGM could be helpful in this situation (5656 Javherani R, Purandare V, Bhatt A, Kumaran S, Sayyad M, Unnikrishnan A. Flash glucose monitoring in subjects with diabetes on hemodialysis: A pilot study. Indian J Endocrinol Metab. 2018;22(6):848.,5757 Sobngwi E, Ashuntantang G, Ndounia E, Dehayem M, Azabji-Kenfack M, Kaze F, et al. Continuous interstitial glucose monitoring in non-diabetic subjects with end-stage renal disease undergoing maintenance haemodialysis. Diabetes Res Clin Pract. 2010;90(1):22-5.).

Additional studies are needed to determine the accuracy and safety of this form of blood glucose monitoring in patients with ESRD. Due to a lack of evidence at present, it is not recommended for people undergoing peritoneal dialysis. For those on HD, isCGM/FGM should be applied with caution. It is crucial to assess patterns rather than focusing on specific glucose values (5858 Galindo RJ, Beck RW, Scioscia MF, Umpierrez GE, Tuttle KR. Glycemic Monitoring and Management in Advanced Chronic Kidney Disease. Endocr Rev. 2020;41(5):756-74.).

INTENSIVE CARE UNIT

Dysglycemia, which includes stress-induced hyperglycemia, hypoglycemia, and excessive glycemic variability, is common amongst critically ill patients (5959 Aramendi I, Burghi G, Manzanares W. Dysglycemia in the critically ill patient: current evidence and future perspectives. Rev Bras Ter Intensiva. 2017;29(3):364-72.). Previous studies associate poor glycemic control with increased morbidity and mortality in critically ill patients (6060 Krinsley JS. Glycemic variability: A strong independent predictor of mortality in critically ill patients. Crit Care Med. 2008;36(11):3008-13.). A recent meta-analysis demonstrated that intensive glucose control among critically ill patients reduces all-cause mortality. However, severe hypoglycemia is more frequent in this context, justifying continuous assessment of glycemic status (6161 Yao R, Ren C, Wu G, Zhu Y, Xia Z, Yao Y. Is intensive glucose control bad for critically ill patients? A systematic review and meta-analysis. Int J Biol Sci. 2020;16(9):1658-75.). Currently, glucose control in the intensive care unit (ICU) is based on intermittent measurements using handheld meters for point-of-care testing. Handheld glucose meters are not designed for ICU settings, and their accuracy is questionable and markedly inferior to central laboratory or blood gas analysis, especially in patients with anemia or hypoxia, or those exposed to certain drugs (6262 Garingarao CJP, Buenaluz-Sedurante M, Jimeno CA. Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock. J Diabetes Sci Technol. 2014;8(5):937-44.). In critically ill patients, between 4% and 15% of hypoglycemic events are undetected, more frequently when there is a long time interval between glucose measurements (6363 van Steen S, Rijkenberg S, Limpens J, van der Voort P, Hermanides J, DeVries J. The Clinical Benefits and Accuracy of Continuous Glucose Monitoring Systems in Critically Ill Patients – A Systematic Scoping Review. Sensors. 2017;17(12):146.).

Meanwhile, the CGM systems are gradually gaining space, as they can overcome these limitations and detect acute changes in glucose levels without overloading the nursing staff (6464 Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. 2014;18(4):453.). However, some limitations of current CGM systems are remarked, originating from physiological and technical aspects. CGM systems’ technical performance and accuracy must be reliable to be used in daily practice. Data reliability about the patients with diabetes cannot be automatically transferred to a different situation like the ICU, where many variables (edema, hypotension, and vasoactive drugs, among others) can affect CGM performance (6565 Block C, Manuel-y-Keenoy B, Rogiers P, Jorens P, Gaal L. Glucose Control and Use of Continuous Glucose Monitoring in the Intensive Care Unit: A Critical Review. Curr Diabetes Rev. 2008;4(3):234-44.).

A systematic review by van Steen and cols. identified 32 studies that addressed the accuracy of CGM. However, only five randomized controlled trials studies, focusing on the ICU population, explored the efficacy of CGM (6363 van Steen S, Rijkenberg S, Limpens J, van der Voort P, Hermanides J, DeVries J. The Clinical Benefits and Accuracy of Continuous Glucose Monitoring Systems in Critically Ill Patients – A Systematic Scoping Review. Sensors. 2017;17(12):146.). Despite the conclusive evidence about the CGM failure to improve glycemic control, no formal meta-analysis could be conducted owing to a low number of studies, small sample size, group heterogeneity, and difference in glucose target values.

In addition, the literature is inconsistent regarding the detection of hypoglycemia by CGM. While Holzinger and cols. showed that CGM reduces episodes of hypoglycemia (6666 Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-Time Continuous Glucose Monitoring in Critically Ill Patients: A prospective randomized trial. Diabetes Care. 2010;33(3):467-72.), Boom and cols. revealed no increase in the detection of hypoglycemic events with the system (6464 Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. 2014;18(4):453.).

Ancona and cols. reported high reliability of the FreeStyle LibreTM system in eight adult patients admitted to the ICU; all had been diagnosed with DM. The authors observed acceptable values and clinical accuracy of arterial blood glucose, even higher than the capillary blood glucose levels (6767 Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-74.).

Another study, evaluating the performance of isCGM/FGM in the pediatric ICU setting, did not show a satisfactory result, with a tendency of underestimating glucose levels. The differences in arterial blood gas, capillary blood, and biochemical serum are greater in the hypoglycemic and normoglycemic range than in the hyperglycemic range (6868 Kotzapanagiotou E, Tsotridou E, Volakli E, Dimitriadou M, Chochliourou E, Kalamitsou S, et al. Evaluation of continuous flash glucose monitoring in a pediatric ICU setting. J Clin Monit Comput. 2020;34(4):843-52.). The observation could explain the satisfactory results of the study by Ancona and cols., in which all patients had diabetes before hospitalization and, therefore, usually higher glucose values.

Zhang and cols. evaluated FreeStyle LibreTM feasibility and accuracy in 17 patients with COVID-19 and hyperglycemia admitted to the ICU. The error grid analyses against venous blood glucose showed acceptable clinical accuracy, with 97.1% of glucose readings falling into zones A and B in Clarke error grid analysis and 97.7% in zones A and B in consensus error grid analysis. However, MARD was 22.4%, which was higher than reported in outpatients with diabetes (6969 Zhang Y, Liu X, Zhang J, Fu J, Li S, Chen S, et al. Evaluation for the feasibility and accuracy of Freestyle Libre Flash Glucose Monitoring System used by COVID-19 patients in intensive care unit. J Diabetes. 2021;13(7):603-5.).

The lower values found with the isCGM/FGM may not reflect inaccuracy. However, it may reflect the difference in the glucose level measurement in plasma before glucose consumption by cells, as compared with interstitial after glucose utilization. Glucose diffusion is determined by blood supply, which may be impaired in patients with a critical illness, vasopressor therapy, or both. Simultaneously, exogenous insulin administration may augment glucose uptake through subcutaneous cells. Importantly, interstitial edema, common in critically ill patients, may further dilute the subcutaneous glucose and contribute to an increased glucose gradient between compartments (6767 Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-74.).

Assessing the accuracy of subcutaneous and intravascular CGM devices is still a challenge, despite randomized clinical trials pointing to a lower accuracy of the subcutaneous device. However, the intravascular device requires an invasive procedure for placement, with an increased risk of peripheral venous thrombosis and infection. Schierenbeck and cols. showed a MARD of 30.5% versus 6.5% between FreeStyle LibreTM and Eirus (intravascular) after cardiac surgery (7070 Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 Different Continuous Glucose Monitoring Systems in Patients Undergoing Cardiac Surgery. J Diabetes Sci Technol. 2017 Jan;11(1):108-16.). One of the hypotheses, although not reported by the authors, could be the possible reason for poor subcutaneous tissue perfusion, which was provoked by perioperative hypothermia, limiting its effectiveness (6767 Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-74.).

Though a definitive conclusion has not been reached, the expert consensus is that CGM could offer improved glucose control with less risk of hypoglycemic events in ICU patients, justifying future research through a randomized controlled clinical trial (7171 Salinas PD, Mendez CE. Glucose Management Technologies for the Critically Ill. J Diabetes Sci Technol. 2019;13(4):682-90.).

CIRRHOSIS

DM and glucose intolerance are observed in patients with liver cirrhosis since the liver plays a central role in glucose metabolism (7272 Garcia-Compean D, Jaquez-Quintana JO, Gonzalez-Gonzalez JA, Maldonado-Garza H. Liver cirrhosis and diabetes: Risk factors, pathophysiology, clinical implications and management. World J Gastroenterol. 2009;15(3):280-8.). A large cohort study showed that DM is an independent risk factor for developing chronic nonalcoholic liver disease, along with hepatocellular carcinoma (7373 El-Serag HB, Tran T, Everhart JE. Diabetes increases the risk of chronic liver disease and hepatocellular carcinoma. Gastroenterology. 2004;126(2):460-8.). Furthermore, cirrhosis is a major cause of death in patients with DM (7474 de Marco R, Locatelli F, Zoppini G, Verlato G, Bonora E, Muggeo M. Cause-specific mortality in type 2 diabetes. The Verona Diabetes Study. Diabetes Care. 1999;22(5):756-61.).

In patients with cirrhosis, A1c generally underestimates glycemic status. Possible explanations include the shortened erythrocyte life span and anemia, frequently observed in patients with advanced liver disease, whether due to bleeding or hemolysis related to hypersplenism (7575 Silva TE, Ronsoni MF, Schiavon LL. Challenges in diagnosing and monitoring diabetes in patients with chronic liver diseases. Diabetes Metab Syndr Clin Res Rev. 2018;12(3):431-40.). Individuals with chronic liver disease have insulin resistance and hyperinsulinemia. Moreover, postprandial hyperglycemia is frequently observed in these patients. After an overnight fast, patients with cirrhosis will develop a metabolic profile similar to that found in normal individuals after two to three days of fasting, due to their low hepatic glycogen supply (7676 Owen OE, Trapp VE, Reichard GA, Mozzoli MA, Moctezuma J, Paul P, et al. Nature and quantity of fuels consumed in patients with alcoholic cirrhosis. J Clin Invest. 1983;72(5):1821-32.). SMBG provides information according to the recommended frequency, which usually occurs before and after meals. However, nocturnal hypoglycemia is not normally captured, and it is difficult to obtain information on fluctuations in glucose levels (7777 Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial. Diabetes Ther. 2017;8(1):55-73.). Honda and cols. reported a direct relationship between a deterioration of the hepatic functional reserve and higher glycemic variability with CGM in individuals with DM. Their study identified hidden abnormalities of glucose fluctuations in this population, along with the presence of asymptomatic nocturnal hypoglycemia (7878 Honda F, Hiramatsu A, Hyogo H, Aikata H, Daijo K, Teraoka Y, et al. Evaluation of glycemic variability in chronic liver disease patients with type 2 diabetes mellitus using continuous glucose monitoring. PLoS One. 2018;13(4):e0195028.).

A recent study evaluated isCGM/FGM performance in patients with diabetes and liver cirrhosis (LC). Thirty-one patients in the study group and 30 controls with diabetes, but without liver disease, were analyzed. The results showed a strong agreement between isCGM/FGM readings and capillary glycemia (7979 Costa D, Lourenço J, Monteiro AM, Castro B, Oliveira P, Tinoco MC, et al. Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus. Sci Rep. 2020;10(1):7460.). A MARD of 12.68% was found in the LC group, versus 10.55% in the control group, similar to the results of previous studies using different target populations (99 Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787-94.,6767 Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-74.,7979 Costa D, Lourenço J, Monteiro AM, Castro B, Oliveira P, Tinoco MC, et al. Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus. Sci Rep. 2020;10(1):7460.). Nevertheless, MARD was persistently higher in those with LC as compared to the control, possibly due to the fluid overload characteristic of cirrhotic permanent hyperdynamic circulation and fluid retention state, significantly impacting isCGM/FGM's analytical accuracy. However, the clinical usability of isCGM/FGM in patients with LC was confirmed through Consensus Error Grid analysis, which resulted in 80.36% of values in zone A and 99.83% of values in zones A + B (7979 Costa D, Lourenço J, Monteiro AM, Castro B, Oliveira P, Tinoco MC, et al. Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus. Sci Rep. 2020;10(1):7460.).

The systemic changes caused by liver dysfunction generate imprecision in markers used to monitor glycemic control. The reason is that in the initial stages, fasting blood glucose levels are normal in 23% of those with evident diabetes, revealing a proving difficult diagnosis (8080 Nishida T, Tsuji S, Tsujii M, Arimitsu S, Haruna Y, Imano E, et al. Oral Glucose Tolerance Test Predicts Prognosis of Patients with Liver Cirrhosis. Am J Gastroenterol. 2006;101(1):70-5.).

Although the impact of early diagnosis and treatment of glycemic changes in patients with cirrhosis is unknown, it is tempting to speculate this as beneficial. While monitoring glycemic control in all DM stages, the isCGM/FGM could facilitate the monitoring and identification of glycemic variability and provide the appropriate therapy for identified changes. However, future studies are required to confirm the accuracy of the tool and its application for clinical practice (7979 Costa D, Lourenço J, Monteiro AM, Castro B, Oliveira P, Tinoco MC, et al. Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus. Sci Rep. 2020;10(1):7460.).

PHYSICAL EXERCISE

Physical activity (PA) plays important role in blood glucose management and the overall vitality of individuals with diabetes and prediabetes (8181 Riddell MC, Gallen IW, Smart CE, Taplin CE, Adolfsson P, Lumb AN, et al. Exercise management in type 1 diabetes: a consensus statement. Lancet Diabetes Endocrinol. 2017;5(5):377-90.). Regular exercise can prevent or delay the development of T2DM. At the same time, it is part of an effective non-pharmacological intervention. Previous studies demonstrated that aerobic or combined (aerobic and resistance) exercise can reduce glucose variability in patients with diabetes (8282 Corrêa APS, Figueira FR, Umpierre D, Casali KR, Schaan BD. Inspiratory muscle loading: a new approach for lowering glucose levels and glucose variability in patients with Type 2 diabetes. Diabet Med. 2015;32(9):1255-7.,8383 Figueira FR, Umpierre D, Casali KR, Tetelbom PS, Henn NT, Ribeiro JP, et al. Aerobic and Combined Exercise Sessions Reduce Glucose Variability in Type 2 Diabetes: Crossover Randomized Trial. PLoS One. 2013;8(3):e57733.). In addition, circulating markers of oxidative stress and inflammation, influenced by different exercise protocols, have been related to change in glucose variability in T2DM (8484 Figueira FR, Umpierre D, Bock PM, Waclawovsky G, Guerra AP, Donelli A, et al. Effect of exercise on glucose variability in healthy subjects: randomized crossover trial. Biol Sport. 2019;36(2):141-8.).

Moreover, physical exercises play vital roles in managing T1DM. In a large cross-sectional study of 18,028 adults with T1DM, patients who were most active in PA had better A1c levels, more favorable body mass indexes (BMIs), less dyslipidemia and hypertension, and fewer diabetes-related complications (retinopathy, microalbuminuria) than those who were less active (8585 Bohn B, Herbst A, Pfeifer M, Krakow D, Zimny S, Kopp F, et al. Impact of Physical Activity on Glycemic Control and Prevalence of Cardiovascular Risk Factors in Adults With Type 1 Diabetes: A Cross-sectional Multicenter Study of 18,028 Patients. Diabetes Care. 2015;38(8):1536-43.).

Insulin action in muscle and liver can be modified by intensive exercise sessions and regular PA. Acute, aerobic exercise increases muscle glucose uptake up to fivefold through insulin-independent mechanisms. After exercise, glucose uptake remains elevated by insulin-independent (∼2 hours) and insulin-dependent (up to 48 hours) mechanisms if exercise is prolonged (8686 Magkos F, Tsekouras Y, Kavouras SA, Mittendorfer B, Sidossis LS. Improved insulin sensitivity after a single bout of exercise is curvilinearly related to exercise energy expenditure. Clin Sci. 2008;114(1):59-64.). Improvements in insulin action may last for 24 hours following shorter-duration activities (∼20 minutes) and if the intensity is elevated to near maximal effort (8787 Gillen JB, Little JP, Punthakee Z, Tarnopolsky MA, Riddell MC, Gibala MJ. Acute high-intensity interval exercise reduces the postprandial glucose response and prevalence of hyperglycaemia in patients with type 2 diabetes. Diabetes Obes Metab. 2012;14(6):575-7.). Even low-intensity aerobic exercise lasting 60 minutes or longer enhances insulin action in adults with obesity and insulin resistance for at least 24 hours (8888 Newsom SA, Everett AC, Hinko A, Horowitz JF. A Single Session of Low-Intensity Exercise Is Sufficient to Enhance Insulin Sensitivity Into the Next Day in Obese Adults. Diabetes Care. 2013;36(9):2516-22.).

In T1DM, blood glucose responses to PA are highly variable, considering the duration and intensity of the exercise, initial blood glucose levels, the individual anaerobic threshold, and the amount of insulin in circulation (8181 Riddell MC, Gallen IW, Smart CE, Taplin CE, Adolfsson P, Lumb AN, et al. Exercise management in type 1 diabetes: a consensus statement. Lancet Diabetes Endocrinol. 2017;5(5):377-90.). In general, aerobic exercise decreases blood glucose levels if performed during postprandial periods along with the insulin dose administered at the meal before exercise, and prolonged activity may cause hypoglycemia (8989 Tansey MJ, Tsalikian E, Beck RW, Mauras N, Buckingham BA, Weinzimer SA, et al. The Diabetes Research in Children Network (DirecNet) Study Group. The Effects of Aerobic Exercise on Glucose and Counterregulatory Hormone Concentrations in Children With Type 1 Diabetes. Diabetes Care. 2006;29(1):20-5.,9090 Mallad A, Hinshaw L, Schiavon M, Dalla Man C, Dadlani V, Basu R, et al. Exercise effects on postprandial glucose metabolism in type 1 diabetes: a triple-tracer approach. Am J Physiol-Endocrinol Metab. 2015;308(12):E1106-15.). Exercise while fasting may produce a lesser decrease or a small increase in blood glucose (9191 Turner D, Luzio S, Gray BJ, Bain SC, Hanley S, Richards A, et al. Algorithm that delivers an individualized rapid-acting insulin dose after morning resistance exercise counters post-exercise hyperglycaemia in people with Type 1 diabetes. Diabet Med. 2016;33(4):506-10.). Anaerobic exercises and high-intensity interval training (HIIT) may provide better glucose stability, decrease blood glucose level, or slightly raise the glucose level (9292 Yardley JE, Kenny GP, Perkins BA, Riddell MC, Balaa N, Malcolm J, et al. Resistance Versus Aerobic Exercise: Acute effects on glycemia in type 1 diabetes. Diabetes Care. 2013;36(3):537-42.,9494 Brazeau AS, Rabasa-Lhoret R, Strychar I, Mircescu H. Barriers to Physical Activity Among Patients With Type 1 Diabetes. Diabetes Care. 2008;31(11):2108-9.). Due to greater glycemic instability that often requires therapy adjustments, a higher frequency of glucose monitoring is required in those who are physically active. A previous study identified hypoglycemia as the stronger barrier to regular PA in adults with T1DM (9494 Brazeau AS, Rabasa-Lhoret R, Strychar I, Mircescu H. Barriers to Physical Activity Among Patients With Type 1 Diabetes. Diabetes Care. 2008;31(11):2108-9.).

Thus, the CGM appears to be a possible ally for this population, allowing frequent assessments of glucose levels more conveniently and painlessly. Information, obtained in real-time during and after PA, could help in insulin and carbohydrate intake adjustments during and after exercise, thereby reducing extreme fluctuations in blood glucose. However, recently, CGM devices have been extensively evaluated in the context of physical exercise due to the apparent delay between interstitial glucose readings and blood glucose in situations with rapid changes in concentrations (9595 Zaharieva DP, Turksoy K, McGaugh SM, Pooni R, Vienneau T, Ly T, et al. Lag Time Remains with Newer Real-Time Continuous Glucose Monitoring Technology During Aerobic Exercise in Adults Living with Type 1 Diabetes. Diabetes Technol Ther. 2019;21(6):313-21.).

A study evaluating the accuracy of CGM during prolonged aerobic exercise in patients with T1DM observed a delay of 12 ± 11 minutes between CGM and SMBG, in addition to an increase in MARD of 13% during the exercise (9595 Zaharieva DP, Turksoy K, McGaugh SM, Pooni R, Vienneau T, Ly T, et al. Lag Time Remains with Newer Real-Time Continuous Glucose Monitoring Technology During Aerobic Exercise in Adults Living with Type 1 Diabetes. Diabetes Technol Ther. 2019;21(6):313-21.). In line with this, Biagi and cols. reported an increase in MARD from 9.5% to 16.5% during aerobic exercise; whereas, the anaerobic exercise showed no significant difference (9696 Biagi L, Bertachi A, Quirós C, Giménez M, Conget I, Bondia J, et al. Accuracy of Continuous Glucose Monitoring before, during, and after Aerobic and Anaerobic Exercise in Patients with Type 1 Diabetes Mellitus. Biosensors. 2018;8(1):22.).

In a previous study, the MARD for HIIT changed from 10.4% before exercise to 17.8% during training, in addition to a delay of 35 minutes in reaching half of the maximum glucose value compared to SMBG (9797 Li A, Riddell MC, Potashner D, Brown RE, Aronson R. Time Lag and Accuracy of Continuous Glucose Monitoring During High Intensity Interval Training in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2019;21(5):286-94.).

Aberer and cols. compared three different CGM devices – FreeStyle LibreTM (Abbott), Dexcom G4 PlatinumTM (Dexcom), and Medtronic MiniMed 640GTM (Medtronic) – during moderate aerobic activity, performed by individuals with T1DM, both before and after a meal. The outcome of the comparison reveals a high level of accuracy in all three devices during the exercise. The Abbott system was reported to have the best accuracy, with the lowest MARD (13.2 ± 10.9%) (9898 Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, Elsayed H, et al. Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes. Diabetes Obes Metab. 2017;19(7):1051-5.).

The fact that isCGM/FGM does not require calibration by the user and has longer sensor durability makes the system more convenient in different conditions of everyday life, including exercises (9999 Giani E, Macedoni M, Barilli A, Petitti A, Mameli C, Bosetti A, et al. Performance of the Flash Glucose Monitoring System during exercise in youth with Type 1 diabetes. Diabetes Res Clin Pract. 2018;146:321-9.). Giani and cols. assessed the performance of isCGM/FGM during an interval training exercise in young people with T1DM. The isCGM/FGM MARD during exercise was higher compared to the glucometer (12.5% vs. 5.7%). During the workout session, an increase in the MARD value was recorded from 5.5% at the beginning of the session to 15.8% at 30 minutes, suggesting a possible degradation of MARD during the training session, even if the variation was not statistically significant, probably due to the small sample size (9999 Giani E, Macedoni M, Barilli A, Petitti A, Mameli C, Bosetti A, et al. Performance of the Flash Glucose Monitoring System during exercise in youth with Type 1 diabetes. Diabetes Res Clin Pract. 2018;146:321-9.).

Another recent study evaluated the performance of FreeStyle LibreTM in T1DM patients during moderate aerobic exercise. Immediately before the exercise testing, isCGM/FGM performance in resting conditions showed an overall MARD of 13.7%; whereas overall MARD during exercise increased to 22% and reached 36.3% during exercise-induced hypoglycemia (100100 Moser O, Eckstein ML, Mueller A, Birnbaumer P, Aberer F, Koehler G, et al. Impact of physical exercise on sensor performance of the FreeStyle Libre intermittently viewed continuous glucose monitoring system in people with Type 1 diabetes: a randomized crossover trial. Diabet Med. 2019;36(5):606-11.).

Zaharieva and cols. evaluated simultaneously two rt-CGM devices during aerobic exercise and one isCGM/FGM in a male patient with T1DM. They observed a significant delay of these devices concerning SMBG, with differences in measurements greater in the first 30 minutes of the exercise (101101 Zaharieva DP, Riddell MC, Henske J. The Accuracy of Continuous Glucose Monitoring and Flash Glucose Monitoring During Aerobic Exercise in Type 1 Diabetes. J Diabetes Sci Technol. 2019;13(1):140-1.).

The CORRIDA study showed the superiority of rt-CGM to isCGM/FGM in reducing hypoglycemia and improving TIR in adults with T1DM and normal hypoglycemia awareness, demonstrating the value of rt-CGM alarms during exercise for daily diabetes self-management (102102 Hásková A, Radovnická L, Petruželková L, Parkin CG, Grunberger G, Horová E, et al. Real-time CGM Is Superior to Flash Glucose Monitoring for Glucose Control in Type 1 Diabetes: The CORRIDA Randomized Controlled Trial. Diabetes Care. 2020;43(11):2744-50.).

The performance of isCGM/FGM was assessed during the daily exercise and the challenges faced by individuals with T1DM. The overall MARD during inpatient phases was 14.3%. However, the overall MARD during acute exercise was 29.8%, not sufficiently accurate and required confirmatory blood glucose measurements (103103 Moser O, Eckstein ML, McCarthy O, Deere R, Pitt J, Williams DM, et al. Performance of the Freestyle Libre flash glucose monitoring (flash GM) system in individuals with type 1 diabetes: A secondary outcome analysis of a randomized crossover trial. Diabetes Obes Metab. 2019;21(11):2505-12.).

Although isCGM/FGM is a potentially useful tool during and after PA, the lower performance of the system observed in a study requires additional confirmation of the results, especially in light of the change in the therapeutic protocol (100100 Moser O, Eckstein ML, Mueller A, Birnbaumer P, Aberer F, Koehler G, et al. Impact of physical exercise on sensor performance of the FreeStyle Libre intermittently viewed continuous glucose monitoring system in people with Type 1 diabetes: a randomized crossover trial. Diabet Med. 2019;36(5):606-11.).

In conclusion, the development of an isCGM/FGM has transformed the management and treatment of patients with diabetes in recent years. Several clinical situations, previously presented difficulties for glucose monitoring, were contemplated in light of the benefits of the isCGM/FGM. Though current scientific evidence supports the use of the flash system in individuals with DM, future studies are required to reinforce its performance and safety.

  • Funding statement: the authors received no specific funding for this work.

REFERENCES

  • 1
    Hirsch IB. Glycemic Variability and Diabetes Complications: Does It Matter? Of Course It Does! Diabetes Care. 2015;38(8):1610-4.
  • 2
    Garber AJ, Handelsman Y, Grunberger G, Einhorn D, Abrahamson MJ, Barzilay JI, et al. Consensus Statement by the American Association of Clinical Endocrinologists and American College of Endocrinology on the Comprehensive Type 2 Diabetes Management Algorithm – 2020 Executive Summary. Endocr Pract. 2020;26(1):107-39.
  • 3
    Beyond A1C Writing Group. Need for Regulatory Change to Incorporate Beyond A1C Glycemic Metrics. Diabetes Care. 2018;41(6):e92-4.
  • 4
    Bry L, Chen PC, Sacks DB. Effects of Hemoglobin Variants and Chemically Modified Derivatives on Assays for Glycohemoglobin. Clin Chem. 2001;47(2):153-63.
  • 5
    Shi T, Li D, Li G, Zhang Y, Xu K, Lu L. Modeling and Measurement of Correlation between Blood and Interstitial Glucose Changes. J Diabetes Res. 2016;2016:1-9.
  • 6
    Koschinsky T, Heinemann L. Sensors for glucose monitoring: technical and clinical aspects. Diabetes Metab Res Rev. 2001;17(2):113-23.
  • 7
    Dunn TC, Xu Y, Hayter G, Ajjan RA. Real-world flash glucose monitoring patterns and associations between self-monitoring frequency and glycaemic measures: A European analysis of over 60 million glucose tests. Diabetes Res Clin Pract. 2018;137:37-46.
  • 8
    Heinemann L. Finger Pricking and Pain: A Never Ending Story. J Diabetes Sci Technol. 2008;2(5):919-21.
  • 9
    Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther. 2015;17(11):787-94.
  • 10
    Leelarathna L, Wilmot EG. Flash forward: a review of flash glucose monitoring. Diabet Med. 2018;35(4):472-82.
  • 11
    Danne T, Nimri R, Battelino T, Bergenstal RM, Close KL, DeVries JH, et al. International Consensus on Use of Continuous Glucose Monitoring. Diabetes Care. 2017;40(12):1631-40.
  • 12
    Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593-603.
  • 13
    Beck RW, Bergenstal RM, Riddlesworth TD, Kollman C, Li Z, Brown AS, et al. Validation of Time in Range as an Outcome Measure for Diabetes Clinical Trials. Diabetes Care. 2019;42(3):400-5.
  • 14
    Lu J, Ma X, Zhou J, Zhang L, Mo Y, Ying L, et al. Association of Time in Range, as Assessed by Continuous Glucose Monitoring, With Diabetic Retinopathy in Type 2 Diabetes. Diabetes Care. 2018;41(11):2370-6.
  • 15
    Beck RW, Bergenstal RM, Cheng P, Kollman C, Carlson AL, Johnson ML, et al. The Relationships Between Time in Range, Hyperglycemia Metrics, and HbA1c. J Diabetes Sci Technol. 2019;13(4):614-26.
  • 16
    Vigersky RA, McMahon C. The Relationship of Hemoglobin A1C to Time-in-Range in Patients with Diabetes. Diabetes Technol Ther. 2019;21(2):81-5.
  • 17
    Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Use of Flash Glucose-Sensing Technology for 12 months as a Replacement for Blood Glucose Monitoring in Insulin-treated Type 2 Diabetes. Diabetes Ther. 2017;8(3):573-86.
  • 18
    Alva S, Bailey T, Brazg R, Budiman ES, Castorino K, Christiansen MP, et al. Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With Advanced Algorithm in Pediatric and Adult Population With Diabetes. J Diabetes Sci Technol. 2020;193229682095875.
  • 19
    Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016;388(10057):2254-63.
  • 20
    Pickup JC, Freeman SC, Sutton AJ. Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: meta-analysis of randomised controlled trials using individual patient data. BMJ. 2011;343:d3805.
  • 21
    Calliari LEP, Krakauer M, Vianna AGD, Ram Y, Barbieri DE, Xu Y, et al. Real-world flash glucose monitoring in Brazil: can sensors make a difference in diabetes management in developing countries? Diabetol Metab Syndr. 2020;12:3.
  • 22
    Al Hayek AA, Robert AA, Al Dawish MA. Differences of FreeStyle Libre Flash Glucose Monitoring System and Finger Pricks on Clinical Characteristics and Glucose Monitoring Satisfactions in Type 1 Diabetes Using Insulin Pump. Clin Med Insights Endocrinol Diabetes. 2019;12:117955141986110.
  • 23
    Lepercq J, Le Ray C, Godefroy C, Pelage L, Dubois-Laforgue D, Timsit J. Determinants of a good perinatal outcome in 588 pregnancies in women with type 1 diabetes. Diabetes Metab. 2019;45(2):191-6.
  • 24
    Buhary B, Almohareb O, Aljohani N, Alzahrani S, Elkaissi S, Sherbeeni S, et al. Glycemic control and pregnancy outcomes in patients with diabetes in pregnancy: A retrospective study. Indian J Endocrinol Metab. 2016;20(4):481.
  • 25
    Ringholm L, Mathiesen ER, Kelstrup L, Damm P. Managing type 1 diabetes mellitus in pregnancy – from planning to breastfeeding. Nat Rev Endocrinol. 2012;8(11):659-67.
  • 26
    Nielsen LR, Ekbom P, Damm P, Glumer C, Frandsen MM, Jensen DM, et al. HbA1c Levels Are Significantly Lower in Early and Late Pregnancy. Diabetes Care. 2004;27(5):1200-1.
  • 27
    Herranz L, Saez-de-Ibarra L, Grande C, Pallardo LF. Non-Glycemic-Dependent Reduction of Late Pregnancy A1C Levels in Women With Type 1 Diabetes. Diabetes Care. 2007;30(6):1579-80.
  • 28
    García-Patterson A, Gich I, Amini SB, Catalano PM, de Leiva A, Corcoy R. Insulin requirements throughout pregnancy in women with type 1 diabetes mellitus: three changes of direction. Diabetologia. 2010;53(3):446-51.
  • 29
    Ringholm L, Pedersen-Bjergaard U, Thorsteinsson B, Damm P, Mathiesen ER. Hypoglycaemia during pregnancy in women with Type 1 diabetes. Diabet Med. 2012;29(5):558-66.
  • 30
    Scott EM, Bilous RW, Kautzky-Willer A. Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes. Diabetes Technol Ther. 2018;20(3):180-8.
  • 31
    National Institute for Health and Care Excellence (UK). Diabetes in pregnancy: management from preconception to the postnatal period [Internet]. London; 2015 [cited 2020 Apr 1]. (National Institute for Health and Care Excellence: Clinical Guidelines). Available from: http://www.ncbi.nlm.nih.gov/books/NBK555331/
  • 32
    Feig DS, Donovan LE, Corcoy R, Murphy KE, Amiel SA, Hunt KF, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial. Lancet. 2017;390(10110):2347-59.
  • 33
    Scott EM, Feig DS, Murphy HR, Law GR. Continuous Glucose Monitoring in Pregnancy: Importance of Analyzing Temporal Profiles to Understand Clinical Outcomes. Diabetes Care. 2020;43(6):1178-84.
  • 34
    Kristensen K, Ögge LE, Sengpiel V, Kjölhede K, Dotevall A, Elfvin A, et al. Continuous glucose monitoring in pregnant women with type 1 diabetes: an observational cohort study of 186 pregnancies. Diabetologia. 2019;62(7):114353.
  • 35
    Murphy HR, Rayman G, Lewis K, Kelly S, Johal B, Duffield K, et al. Effectiveness of continuous glucose monitoring in pregnant women with diabetes: randomised clinical trial. BMJ. 2008;337:a1680.
  • 36
    Yu F, Lv L, Liang Z, Wang Y, Wen J, Lin X, et al. Continuous Glucose Monitoring Effects on Maternal Glycemic Control and Pregnancy Outcomes in Patients With Gestational Diabetes Mellitus: A Prospective Cohort Study. J Clin Endocrinol Metab. 2014;99(12):4674-82.
  • 37
    Alfadhli E, Osman E, Basri T. Use of a real time continuous glucose monitoring system as an educational tool for patients with gestational diabetes. Diabetol Metab Syndr. 2016;8(1):48.
  • 38
    Secher AL, Madsen AB, Ringholm L, Barfred C, Stage E, Andersen HU, et al. Patient satisfaction and barriers to initiating real-time continuous glucose monitoring in early pregnancy in women with diabetes: CGM in early pregnancy. Diabet Med. 2012;29(2):272-7.
  • 39
    Sola-Gazagnes A, Faucher P, Jacqueminet S, Ciangura C, Dubois-Laforgue D, Mosnier-Pudar H, et al. Disagreement between capillary blood glucose and flash glucose monitoring sensor can lead to inadequate treatment adjustments during pregnancy. Diabetes Metab. 2020;46(2):158-63.
  • 40
    Liyanage T, Ninomiya T, Jha V, Neal B, Patrice HM, Okpechi I, et al. Worldwide access to treatment for end-stage kidney disease: a systematic review. Lancet. 2015;385(9981):1975-82.
  • 41
    Schmitz O, Ørskov L, Lund S, Møller N, Christiansen JS, Ørskov H. Glucose Metabolism in Chronic Renal Failure with Reference to GH Treatment of Uremic Children. J Pediatr Endocrinol Metab. 1993;6(1):53-9.
  • 42
    Abe M, Kaizu K, Matsumoto K. Plasma Insulin is Removed by Hemodialysis: Evaluation of the Relation Between Plasma Insulin and Glucose by Using a Dialysate With or Without Glucose. Ther Apher Dial. 2007;11(4):280-7.
  • 43
    Morioka T, Emoto M, Tabata T, Shoji T, Tahara H, Kishimoto H, et al. Glycemic Control Is a Predictor of Survival for Diabetic Patients on Hemodialysis. Diabetes Care. 2001;24(5):909-13.
  • 44
    Ramirez SPB, McCullough KP, Thumma JR, Nelson RG, Morgenstern H, Gillespie BW, et al. Hemoglobin A1c Levels and Mortality in the Diabetic Hemodialysis Population: Findings from the Dialysis Outcomes and Practice Patterns Study (DOPPS). Diabetes Care. 2012;35(12):2527-32.
  • 45
    Yoo DE, Park JT, Oh HJ, Kim SJ, Lee MJ, Shin DH, et al. Good Glycemic Control Is Associated with Better Survival in Diabetic Patients on Peritoneal Dialysis: A Prospective Observational Study. PLoS One. 2012;7(1):e30072.
  • 46
    Gai M, Merlo I, Dellepiane S, Cantaluppi V, Leonardi G, Fop F, et al. Glycemic Pattern in Diabetic Patients on Hemodialysis: Continuous Glucose Monitoring (CGM) Analysis. Blood Purif. 2014;38(1):68-73.
  • 47
    de Boer IH, Caramori ML, Chan JCN, Heerspink HJL, Hurst C, Khunti K, et al. KDIGO 2020 Clinical Practice Guideline for Diabetes Management in Chronic Kidney Disease. Kidney Int. 2020;98(4):S1-115.
  • 48
    Divani M, Georgianos PI, Didangelos T, Iliadis F, Makedou A, Hatzitolios A, et al. Comparison of Glycemic Markers in Chronic Hemodialysis Using Continuous Glucose Monitoring. Am J Nephrol. 2018;47(1):21-9.
  • 49
    Jin Y, Su X, Yin G, Xu X, Lou J, Chen J, et al. Blood glucose fluctuations in hemodialysis patients with end stage diabetic nephropathy. J Diabetes Complications. 2015;29(3):395-9.
  • 50
    Jung HS, Kim HI, Kim MJ, Yoon JW, Ahn HY, Cho YM, et al. Analysis of Hemodialysis-Associated Hypoglycemia in Patients with Type 2 Diabetes Using a Continuous Glucose Monitoring System. Diabetes Technol Ther. 2010;12(10):801-7.
  • 51
    Yajima T, Takahashi H, Yasuda K. Comparison of Interstitial Fluid Glucose Levels Obtained by Continuous Glucose Monitoring and Flash Glucose Monitoring in Patients With Type 2 Diabetes Mellitus Undergoing Hemodialysis. J Diabetes Sci Technol. 2020;14(6):1088-94.
  • 52
    Kumagai R, Muramatsu A, Fujii M, Katakura Y, Ito K, Fujie K, et al. Comparison of glucose monitoring between Freestyle Libre Pro and iPro2 in patients with diabetes mellitus. J Diabetes Investig. 2019;10(3):851-6.
  • 53
    Ólafsdóttir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, Skrtic S, et al. A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2017;19(3):164-72.
  • 54
    Hissa MRN, Hissa PNG, Guimarães SB, Hissa MN. Use of continuous glucose monitoring system in patients with type 2 mellitus diabetic during hemodialysis treatment. Diabetol Metab Syndr. 2021;13(1):104.
  • 55
    Ushigome E, Matsusaki S, Watanabe N, Hashimoto T, Nakamura N, Fukui M. Critical discrepancy in blood glucose control levels evaluated by glycated albumin and estimated hemoglobin A1c levels determined from a flash continuous glucose monitoring system in patients with type 2 diabetes on hemodialysis. J Diabetes Investig. 2020;11(6):1570-4.
  • 56
    Javherani R, Purandare V, Bhatt A, Kumaran S, Sayyad M, Unnikrishnan A. Flash glucose monitoring in subjects with diabetes on hemodialysis: A pilot study. Indian J Endocrinol Metab. 2018;22(6):848.
  • 57
    Sobngwi E, Ashuntantang G, Ndounia E, Dehayem M, Azabji-Kenfack M, Kaze F, et al. Continuous interstitial glucose monitoring in non-diabetic subjects with end-stage renal disease undergoing maintenance haemodialysis. Diabetes Res Clin Pract. 2010;90(1):22-5.
  • 58
    Galindo RJ, Beck RW, Scioscia MF, Umpierrez GE, Tuttle KR. Glycemic Monitoring and Management in Advanced Chronic Kidney Disease. Endocr Rev. 2020;41(5):756-74.
  • 59
    Aramendi I, Burghi G, Manzanares W. Dysglycemia in the critically ill patient: current evidence and future perspectives. Rev Bras Ter Intensiva. 2017;29(3):364-72.
  • 60
    Krinsley JS. Glycemic variability: A strong independent predictor of mortality in critically ill patients. Crit Care Med. 2008;36(11):3008-13.
  • 61
    Yao R, Ren C, Wu G, Zhu Y, Xia Z, Yao Y. Is intensive glucose control bad for critically ill patients? A systematic review and meta-analysis. Int J Biol Sci. 2020;16(9):1658-75.
  • 62
    Garingarao CJP, Buenaluz-Sedurante M, Jimeno CA. Accuracy of Point-of-Care Blood Glucose Measurements in Critically Ill Patients in Shock. J Diabetes Sci Technol. 2014;8(5):937-44.
  • 63
    van Steen S, Rijkenberg S, Limpens J, van der Voort P, Hermanides J, DeVries J. The Clinical Benefits and Accuracy of Continuous Glucose Monitoring Systems in Critically Ill Patients – A Systematic Scoping Review. Sensors. 2017;17(12):146.
  • 64
    Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. 2014;18(4):453.
  • 65
    Block C, Manuel-y-Keenoy B, Rogiers P, Jorens P, Gaal L. Glucose Control and Use of Continuous Glucose Monitoring in the Intensive Care Unit: A Critical Review. Curr Diabetes Rev. 2008;4(3):234-44.
  • 66
    Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-Time Continuous Glucose Monitoring in Critically Ill Patients: A prospective randomized trial. Diabetes Care. 2010;33(3):467-72.
  • 67
    Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-74.
  • 68
    Kotzapanagiotou E, Tsotridou E, Volakli E, Dimitriadou M, Chochliourou E, Kalamitsou S, et al. Evaluation of continuous flash glucose monitoring in a pediatric ICU setting. J Clin Monit Comput. 2020;34(4):843-52.
  • 69
    Zhang Y, Liu X, Zhang J, Fu J, Li S, Chen S, et al. Evaluation for the feasibility and accuracy of Freestyle Libre Flash Glucose Monitoring System used by COVID-19 patients in intensive care unit. J Diabetes. 2021;13(7):603-5.
  • 70
    Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 Different Continuous Glucose Monitoring Systems in Patients Undergoing Cardiac Surgery. J Diabetes Sci Technol. 2017 Jan;11(1):108-16.
  • 71
    Salinas PD, Mendez CE. Glucose Management Technologies for the Critically Ill. J Diabetes Sci Technol. 2019;13(4):682-90.
  • 72
    Garcia-Compean D, Jaquez-Quintana JO, Gonzalez-Gonzalez JA, Maldonado-Garza H. Liver cirrhosis and diabetes: Risk factors, pathophysiology, clinical implications and management. World J Gastroenterol. 2009;15(3):280-8.
  • 73
    El-Serag HB, Tran T, Everhart JE. Diabetes increases the risk of chronic liver disease and hepatocellular carcinoma. Gastroenterology. 2004;126(2):460-8.
  • 74
    de Marco R, Locatelli F, Zoppini G, Verlato G, Bonora E, Muggeo M. Cause-specific mortality in type 2 diabetes. The Verona Diabetes Study. Diabetes Care. 1999;22(5):756-61.
  • 75
    Silva TE, Ronsoni MF, Schiavon LL. Challenges in diagnosing and monitoring diabetes in patients with chronic liver diseases. Diabetes Metab Syndr Clin Res Rev. 2018;12(3):431-40.
  • 76
    Owen OE, Trapp VE, Reichard GA, Mozzoli MA, Moctezuma J, Paul P, et al. Nature and quantity of fuels consumed in patients with alcoholic cirrhosis. J Clin Invest. 1983;72(5):1821-32.
  • 77
    Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, Rayman G. Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial. Diabetes Ther. 2017;8(1):55-73.
  • 78
    Honda F, Hiramatsu A, Hyogo H, Aikata H, Daijo K, Teraoka Y, et al. Evaluation of glycemic variability in chronic liver disease patients with type 2 diabetes mellitus using continuous glucose monitoring. PLoS One. 2018;13(4):e0195028.
  • 79
    Costa D, Lourenço J, Monteiro AM, Castro B, Oliveira P, Tinoco MC, et al. Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus. Sci Rep. 2020;10(1):7460.
  • 80
    Nishida T, Tsuji S, Tsujii M, Arimitsu S, Haruna Y, Imano E, et al. Oral Glucose Tolerance Test Predicts Prognosis of Patients with Liver Cirrhosis. Am J Gastroenterol. 2006;101(1):70-5.
  • 81
    Riddell MC, Gallen IW, Smart CE, Taplin CE, Adolfsson P, Lumb AN, et al. Exercise management in type 1 diabetes: a consensus statement. Lancet Diabetes Endocrinol. 2017;5(5):377-90.
  • 82
    Corrêa APS, Figueira FR, Umpierre D, Casali KR, Schaan BD. Inspiratory muscle loading: a new approach for lowering glucose levels and glucose variability in patients with Type 2 diabetes. Diabet Med. 2015;32(9):1255-7.
  • 83
    Figueira FR, Umpierre D, Casali KR, Tetelbom PS, Henn NT, Ribeiro JP, et al. Aerobic and Combined Exercise Sessions Reduce Glucose Variability in Type 2 Diabetes: Crossover Randomized Trial. PLoS One. 2013;8(3):e57733.
  • 84
    Figueira FR, Umpierre D, Bock PM, Waclawovsky G, Guerra AP, Donelli A, et al. Effect of exercise on glucose variability in healthy subjects: randomized crossover trial. Biol Sport. 2019;36(2):141-8.
  • 85
    Bohn B, Herbst A, Pfeifer M, Krakow D, Zimny S, Kopp F, et al. Impact of Physical Activity on Glycemic Control and Prevalence of Cardiovascular Risk Factors in Adults With Type 1 Diabetes: A Cross-sectional Multicenter Study of 18,028 Patients. Diabetes Care. 2015;38(8):1536-43.
  • 86
    Magkos F, Tsekouras Y, Kavouras SA, Mittendorfer B, Sidossis LS. Improved insulin sensitivity after a single bout of exercise is curvilinearly related to exercise energy expenditure. Clin Sci. 2008;114(1):59-64.
  • 87
    Gillen JB, Little JP, Punthakee Z, Tarnopolsky MA, Riddell MC, Gibala MJ. Acute high-intensity interval exercise reduces the postprandial glucose response and prevalence of hyperglycaemia in patients with type 2 diabetes. Diabetes Obes Metab. 2012;14(6):575-7.
  • 88
    Newsom SA, Everett AC, Hinko A, Horowitz JF. A Single Session of Low-Intensity Exercise Is Sufficient to Enhance Insulin Sensitivity Into the Next Day in Obese Adults. Diabetes Care. 2013;36(9):2516-22.
  • 89
    Tansey MJ, Tsalikian E, Beck RW, Mauras N, Buckingham BA, Weinzimer SA, et al. The Diabetes Research in Children Network (DirecNet) Study Group. The Effects of Aerobic Exercise on Glucose and Counterregulatory Hormone Concentrations in Children With Type 1 Diabetes. Diabetes Care. 2006;29(1):20-5.
  • 90
    Mallad A, Hinshaw L, Schiavon M, Dalla Man C, Dadlani V, Basu R, et al. Exercise effects on postprandial glucose metabolism in type 1 diabetes: a triple-tracer approach. Am J Physiol-Endocrinol Metab. 2015;308(12):E1106-15.
  • 91
    Turner D, Luzio S, Gray BJ, Bain SC, Hanley S, Richards A, et al. Algorithm that delivers an individualized rapid-acting insulin dose after morning resistance exercise counters post-exercise hyperglycaemia in people with Type 1 diabetes. Diabet Med. 2016;33(4):506-10.
  • 92
    Yardley JE, Kenny GP, Perkins BA, Riddell MC, Balaa N, Malcolm J, et al. Resistance Versus Aerobic Exercise: Acute effects on glycemia in type 1 diabetes. Diabetes Care. 2013;36(3):537-42.
  • 93
    Mitchell TH, Abraham G, Schiffrin A, Leiter LA, Marliss EB. Hyperglycemia After Intense Exercise in IDDM Subjects During Continuous Subcutaneous Insulin Infusion. Diabetes Care. 1988;11(4):311-7.
  • 94
    Brazeau AS, Rabasa-Lhoret R, Strychar I, Mircescu H. Barriers to Physical Activity Among Patients With Type 1 Diabetes. Diabetes Care. 2008;31(11):2108-9.
  • 95
    Zaharieva DP, Turksoy K, McGaugh SM, Pooni R, Vienneau T, Ly T, et al. Lag Time Remains with Newer Real-Time Continuous Glucose Monitoring Technology During Aerobic Exercise in Adults Living with Type 1 Diabetes. Diabetes Technol Ther. 2019;21(6):313-21.
  • 96
    Biagi L, Bertachi A, Quirós C, Giménez M, Conget I, Bondia J, et al. Accuracy of Continuous Glucose Monitoring before, during, and after Aerobic and Anaerobic Exercise in Patients with Type 1 Diabetes Mellitus. Biosensors. 2018;8(1):22.
  • 97
    Li A, Riddell MC, Potashner D, Brown RE, Aronson R. Time Lag and Accuracy of Continuous Glucose Monitoring During High Intensity Interval Training in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2019;21(5):286-94.
  • 98
    Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, Elsayed H, et al. Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes. Diabetes Obes Metab. 2017;19(7):1051-5.
  • 99
    Giani E, Macedoni M, Barilli A, Petitti A, Mameli C, Bosetti A, et al. Performance of the Flash Glucose Monitoring System during exercise in youth with Type 1 diabetes. Diabetes Res Clin Pract. 2018;146:321-9.
  • 100
    Moser O, Eckstein ML, Mueller A, Birnbaumer P, Aberer F, Koehler G, et al. Impact of physical exercise on sensor performance of the FreeStyle Libre intermittently viewed continuous glucose monitoring system in people with Type 1 diabetes: a randomized crossover trial. Diabet Med. 2019;36(5):606-11.
  • 101
    Zaharieva DP, Riddell MC, Henske J. The Accuracy of Continuous Glucose Monitoring and Flash Glucose Monitoring During Aerobic Exercise in Type 1 Diabetes. J Diabetes Sci Technol. 2019;13(1):140-1.
  • 102
    Hásková A, Radovnická L, Petruželková L, Parkin CG, Grunberger G, Horová E, et al. Real-time CGM Is Superior to Flash Glucose Monitoring for Glucose Control in Type 1 Diabetes: The CORRIDA Randomized Controlled Trial. Diabetes Care. 2020;43(11):2744-50.
  • 103
    Moser O, Eckstein ML, McCarthy O, Deere R, Pitt J, Williams DM, et al. Performance of the Freestyle Libre flash glucose monitoring (flash GM) system in individuals with type 1 diabetes: A secondary outcome analysis of a randomized crossover trial. Diabetes Obes Metab. 2019;21(11):2505-12.

Publication Dates

  • Publication in this collection
    10 June 2022
  • Date of issue
    Nov-Dec 2022

History

  • Received
    25 Apr 2021
  • Accepted
    04 Feb 2022
Sociedade Brasileira de Endocrinologia e Metabologia Rua Botucatu, 572 - Conjuntos 81/83, 04023-061 São Paulo SP Brasil, Tel: (55 11) 5575-0311 - São Paulo - SP - Brazil
E-mail: aem.editorial.office@endocrino.org.br