Acessibilidade / Reportar erro

Evaluation of temperature increase during in-office bleaching

ABSTRACT

The use of light sources in the bleaching process reduces the time required and promotes satisfactory results. However, these light sources can cause an increase in the pulp temperature.

Objective

The purpose of the present study was to measure the increase in intrapulpal temperature induced by different light-activated bleaching procedures with and without the use of a bleaching gel.

Material and Methods

A human maxillary central incisor was sectioned 2 mm below the cementoenamel junction. A K-type thermocouple probe was introduced into the pulp chamber. A 35% hydrogen peroxide bleaching gel was applied to the vestibular tooth surface. The light units used were a conventional halogen, a hybrid light (only LED and LED/Laser), a high intensity LED, and a green LED light. Temperature increase values were compared by two-way ANOVA and Tukey´s tests (p<0.05).

Results

There were statistically significant differences in temperature increases between the different light sources used and between the same light sources with and without the use of a bleaching gel. The presence of a bleaching gel generated an increase in intra-pulpal temperature in groups activated with halogen light, hybrid light, and high intensity LED. Compared to the other light sources, the conventional halogen lamp applied over the bleaching gel induced a significant increase in temperature (3.83±0.41°C). The green LED unit with and without gel application did not produce any significant intrapulpal temperature variations.

Conclusion

In the present study, the conventional halogen lamp caused the highest increase in intrapulpal temperature, and the green LED caused the least. There was an increase in temperature with all lights tested and the maximum temperature remained below the critical level (5.5°C). The addition of a bleaching gel led to a higher increase in intrapulpal temperatures.

Tooth bleaching; Light; Temperature

INTRODUCTION

Whitening process has been reinforced by the development of new technologies allowing patients to get faster and better results1616 - Kabbach W, Zezell DM, Pereira TM, Albero FG, Clavijo VR, Andrade MF. A thermal investigation of dental bleaching in vitro. Photomed Laser Surg. 2008;26:489-93.,2222 - Nam SH, Lee HW, Cho SH, Lee JK, Jeon YC, Kim GC. High-efficiency tooth bleaching using non-thermal atmospheric pressure plasma with low concentration of hydrogen peroxide. J Appl Oral Sci. 2013;21:265-70.. This technique can be performed at the office, at home, or a combination of both. In-office bleaching is usually performed using bleaching agents with high concentrations of hydrogen peroxide (15-38%). An external source of hybrid light (LED/Diode Laser) can be applied to activate the bleaching gel agent in less time with good results, with a small increase in the intrapulpal temperature, and less sensitivity2121 - Mondelli RF, Azevedo JF, Francisconi AC, Almeida CM, Ishikiriama SK. Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up. J Appl Oral Sci. 2012;20:435-43..

These external light sources have the ability to heat the hydrogen peroxide gel, thus increasing the rate of oxygen decomposition, which promotes bleaching66 - Caviedes-Bucheli J, Ariza-Garcia G, Restrepo-Méndez S, Ríos-Osorio N, Lombana N, Muñoz HR. The effect of tooth bleaching on substance P expression in human dental pulp. J Endod. 2008;34:1462-5.,99 - De Moor RJ, Verheyen J, Verheyen P, Diachuk A, Meire MA, De Coster PJ, et al. Laser teeth bleaching: evaluation of eventual side effects on enamel and the pulp and the efficiency in vitro and in vivo. ScientificWorldJournal. 2015;2015:835405..

Although the use of hybrid light sources in the bleaching process reduces the time required for performing the process and promotes satisfactory results2121 - Mondelli RF, Azevedo JF, Francisconi AC, Almeida CM, Ishikiriama SK. Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up. J Appl Oral Sci. 2012;20:435-43., some of these light sources can cause an increase in the intrapulpal temperature, which can increase the incidence of post-operative sensitivity (55%-100%)1414 - Goldberg M, Grootveld M, Lynch E. Undesirable and adverse effects of tooth-whitening products: a review. Clin Oral Investig. 2010;14:1-10.,2020 - Marson FC, Sensi LG, Vieira LC, Araújo E. Clinical evaluation of in-office dental bleaching treatments with and without the use of light-activation sources. Oper Dent. 2008;33:15-22..

The equipment on the market includes: incandescent light sources, photopolymerizers, infrared lasers, and high power LEDs, among others. Among the benefits obtained by photo activation, only a small fraction of light is absorbed by the bleaching gel, and its energy is converted into heat1717 - Klaric E, Rakic M, Sever I, Tarle Z. Temperature rise during experimental light-activated bleaching. Lasers Med Sci. 2015;30:567-76.,1919 - Kossatz S, Dalanhol AP, Cunha T, Loguercio A, Reis A. Effect of light activation on tooth sensitivity after in-office bleaching. Oper Dent. 2011;36:251-7.. The light, however, can accelerate the degradation of hydrogen peroxide with less time required for bleaching and increase the temperature of the bleaching agent and the dental pulp2121 - Mondelli RF, Azevedo JF, Francisconi AC, Almeida CM, Ishikiriama SK. Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up. J Appl Oral Sci. 2012;20:435-43..

Many different types of lights are currently used in dentistry for bleaching, including halogen, plasma arc, light emitting diodes (LED), diode lasers, carbon dioxide lasers (CO2), argon lasers, and non-thermal atmospheric pressure plasmas (NAPP). Most of the lights emitted by these lamps are within the visible spectrum, but some lamps also emit the light near an infrared or an ultraviolet spectrum. Some systems are designed solely to be used for tooth whitening, while others have several additional functions (bleaching and polymerization of restorative materials). When these lamps are used, an increase in gel temperature and intrapulpal temperature can be observed.

External heat applied to the teeth can cause pulp damage in varying degrees, depending on the magnitude and duration of the increase in temperature1717 - Klaric E, Rakic M, Sever I, Tarle Z. Temperature rise during experimental light-activated bleaching. Lasers Med Sci. 2015;30:567-76.. In an in vivo study with monkeys, an increase in temperature inside of the pulp chamber of 5.5°C caused irreversible pulpits2929 - Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol. 1965;19:515-30.. Temperatures in the pulp chamber above 46°C can cause irreversible changes, such as stasis and thrombosis in the blood vessels2525 - Pohto M, Scheinin A. Microscopic observations on living dental pulp II. The effect of thermal irritants on the circulation of the pulp in the lower rat incisor. Acta Odontol Scand. 1958;16:315-27.. Some authors claim that the highest temperature tolerated by human pulp22 - Baldissara P, Catapano S, Scotti R. Clinical and histological evaluation of thermal injury thresholds in human teeth: a preliminary study. J Oral Rehabil. 1997;24:791-801., 5.5°C, is considered the critical point of tolerance for intrapulpal temperature.

Mainly due to aggressive marketing of some companies, whitening is becoming increasingly popular and many products that use different energy sources and protocols are available on the market. However, they should be tested in such a way that safety parameters are maintained.

The aim of the present in vitro study was to evaluate the intrapulpal temperature increases generated by a selection of lights used as part of the bleaching process, alone or in combination with a bleaching gel.

MATERIAL AND METHODS

This is an in vitro study, and the experimental design for analysis of temperature variation was divided into two study factors: light sources (in five levels: a green LED light, a high intensity LED light, a LED/Laser hybrid light, tested with and without laser activation, and a conventional halogen lamp) and treatment (in two levels: with or without bleaching gel), the design determined ten experimental groups (G1-G10) as described in Table 1.

Table 1
Mean values of temperature variation, standard deviation, and statistical analysis comparison between groups (Tukey’s test)

Material

The manufacturer’s name and other information of the light sources used are listed in Figure 1. The bleaching agent is an in-office 35% hydrogen peroxide gel (Lase Peroxide, DMC Equipamentos Ltda., São Carlos, SP, Brazil). This product has a red color due to the addition of photoactive colorants (Urucum and Juá, Brazil). Other components include amide, glycol and water.

Figure 1
Light units studied

Tooth preparation

A single, human central upper incisor was used throughout the test. This tooth was donated from an organ bank of the Bauru School of Dentistry (FOB-USP), having obtained approval by the ethics committee of the institution through process no. 002/2008. The root of the tooth was sectioned 2 mm below the cementoenamel junction, perpendicular to the long axis of the tooth, using a slow speed diamond disc (Isomet, Buehler Ltd., Lake Bluff, IL, USA). The apical foramen of the root canal was enlarged using a #2135 diamond bur (KG Sorensen, Cotia, SP, Brazil) and all pulp tissue remnants were removed from the pulp chamber. The empty pulp chamber was filled with a heat absorption compound (Termal Paste, Implastec, Votorantim, SP, Brazil; 55.15.3243.3788) to simulate the pulp tissue as a heat conducting medium. A thin K-type thermocouple probe (MT 401A, Minilpa Ind. Com. Ltda., São Paulo, SP, Brazil) was inserted into the pulp chamber through the apical foramen and placed at its most coronal level. Its position was radiographically checked. During the testing procedure, the root surface of the tooth was partially submerged into a water bath at 37±1°C (Figure 2).

Figure 2
Schematic drawing of experimental set up showing the starting temperature before light activation

Bleaching method

A 1.0 mm layer of 35% hydrogen peroxide-based bleaching gel was applied on the entire vestibular surface of the tooth. The distance between the emitting tip of the light source and the tooth surface was set at 10 mm. Light application protocols, as recommended by the manufacturer for each light source, were used. For the conventional halogen light and the high intensity LED units, 5 applications of 30 s each (with light) and 15 second intervals (without light) were utilized (this sequence is typically used in the clinic in order to avoid tooth overheating). For the hybrid LED/laser light, the protocol was 3 min of light activation and 1 min without light, followed by another 3 min of light activation. The green LED was applied for 1 min, followed by 5 min without light and then another 1 min with light activation.

Temperature measurement

The room temperature was controlled at 24+1°C. Each of the light sources were applied with and without the bleaching gel on the tooth surface, for a total of 10 experimental groups. The gel was removed and a new layer was applied every time that a light source was applied to the tooth. At least 15 min were allowed in between the applications of all light sources in order to allow the temperature to return to baseline. Each measurement for all groups was repeated six times, and the temperature was recorded every 30 s during the test. The mean values of the differences between the initial and final temperature readings for each group was calculated.

This protocol was used based on the results of a pilot study performed prior to this study.

Statistical analysis

Two-way analysis of variance (ANOVA) was used to determine if there were statistically significant differences between the groups (p≤0.05) and the Tukey´s test was used for all comparisons among the groups (p≤0.05). The program used for statistical analysis was the Statistica for Windows (version 6.0 StatSoft, Tulsa, Oklahoma, USA).

RESULTS

There were statistically significant differences in the temperature increases among the different light sources and the same light sources, with and without the presence of a bleaching gel. The interaction between the light source and the gel was also significant.

Compared to the other light sources, the conventional halogen lamp applied over the bleaching gel induced the greatest increase in temperature (3.83±0.41°C) (Table 1, Figure 3). The green LED light unit, with and without gel application, did not produce any significant temperature variation. The application of the bleaching gel significantly promoted the temperature increases in all the other units.

Figure 3
Intra-pulpal temperature increase during the in-office bleaching procedures for the different experimental groups

DISCUSSION

Procedures with bleaching agents in high concentrations, coupled with a longer exposure time, can damage the pulp tissue88 - Costa CA, Riehl H, Kina JF, Sacono NT, Hebling J. Human pulp responses to in-office tooth bleaching. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109:e59-64.,2828 - Soares DG, Basso FG, Hebling J, Souza Costa CA. Concentrations of and application protocols for hydrogen peroxide bleaching gels: effects on pulp cell viability and whitening efficacy. J Dent. 2014;42:185-98.. In addition, the temperature variations of the intra-pulpal chamber has been a reason for concern2727 - Sari T, Celik G, Usumez A. Temperature rise in pulp and gel during laser-activated bleaching: in vitro. Lasers Med Sci. 2015;30:577-82..

An increase in real temperature in the intra-pulpal chamber was observed for all groups evaluated, with the exception of the groups who used the green LED light source, possibly due to the low power density1010 - Domínguez A, García JA, Costela A, Gómez C. Influence of the light source and bleaching gel on the efficacy of the tooth whitening process. Photomed Laser Surg. 2011;29:53-9.. The higher temperature of the intra-pulpal chamber was achieved with the use of a conventional halogen lamp coupled with a bleaching gel, which corroborates previous studies1515 - Kabbach W, Zezell DM, Bandeca MC, Pereira TM, Andrade MF. An in vitro thermal analysis during different light-activated hydrogen peroxide bleaching. Laser Phys. 2010;20:1833-7.,3030 - Zhang C, Wang X, Kinoshita J, Zhao B, Toko T, Kimura Y, et al. Effects of KTP laser irradiation, diode laser, and LED on tooth bleaching: a comparative study. Photomed Laser Surg. 2007;25:91-5.. This lamp was widely used at the onset of the in-office bleaching with the use of a light source, but its unstable temperature increases and, consequently, tooth sensitivity has caused it to fall into disuse.

The influence of the diode laser was minimal regarding the increase in temperature when the hybrid light source was also used, probably due to the low power density of the diode laser. This observation is significant because 1W, 2W, and 3W diode lasers are harmful to the pulp, but when used with a low power density, may have therapeutic qualities while reducing inflammatory reactions1313 - Frigo L, Pallota RC, Meneguzzo D, Marcos RL, Penna SC, Lopes-Martins RA. Evaluation of the photo-activated dental bleaching effect on dental pulp in an in vivo rat experimental model. Rev Dental Press Estét. 2009;6:102-14.. Although the present study did not evaluate the sensitivity, the clinical use of this therapeutic laser during bleaching may lead to a reduction in post-operative sensitivity33 - Bortolatto JF, Pretel H, Neto CS, Andrade MF, Moncada G, Junior OB. Effects of LED-laser hybrid light on bleaching effectiveness and tooth sensitivity: a randomized clinical study. Laser Phys Lett. 2013;10:085601..

In order to increase the absorption of light, and consequently the temperature, thus optimizing this effect, some specific bleaching gel colorants are added to the bleaching agents such as red dye (Urucum and Juá, presented in Lase Peroxide, DMC Equipamentos Ltda., São Carlos, SP, Brazil). The groups which used the gel experienced a significantly higher increase in temperature, as shown in previous studies55 - Calatayud JO, Calatayud CO, Zaccagnini AO, Box MJ. Clinical efficacy of a bleaching system based on hydrogen peroxide with or without light activation. Eur J Esthet Dent. 2010;5:216-24.,1111 - Eldeniz AU, Usumez A, Usumez S, Ozturk N. Pulpal temperature rise during light-activated bleaching. J Biomed Mater Res B Appl Biomater. 2005;72:254-9.. In contrast, the application of a bleaching gel or vaseline has been shown to act as a physical barrier, causing a decrease in the intrapulpal chamber temperature1717 - Klaric E, Rakic M, Sever I, Tarle Z. Temperature rise during experimental light-activated bleaching. Lasers Med Sci. 2015;30:567-76..

An increase of 5.5°C in the intrapulpal temperature has been shown to cause irreversible pulpitis in 15% of the monkey teeth tested2929 - Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol. 1965;19:515-30.. However, extrapolation of values obtained from humans to monkeys is highly questionable77 - Choi SH, Roulet JF, Heintze SD, Park SH. Influence of cavity preparation, light-curing units, and composite filling on intrapulpal temperature increase in an in vitro tooth model. Oper Dent. 2014;39:E195-205.. In the present study, an increase of 5.5°C in the intra-pulpal temperature is regarded as the threshold, which should not increase to prevent irreversible damage to the pulp.

Some studies have measured the pulp temperature increases induced by the bleaching gel when the tooth is exposed to different energy sources. The CO2 lasers, diode lasers3030 - Zhang C, Wang X, Kinoshita J, Zhao B, Toko T, Kimura Y, et al. Effects of KTP laser irradiation, diode laser, and LED on tooth bleaching: a comparative study. Photomed Laser Surg. 2007;25:91-5., conventional halogen lamps, and LED1111 - Eldeniz AU, Usumez A, Usumez S, Ozturk N. Pulpal temperature rise during light-activated bleaching. J Biomed Mater Res B Appl Biomater. 2005;72:254-9. with high densities reported temperature values above the critical temperature. However, another study reported a lower temperature increase with halogen lamps, Plasma arcs, LED, and LED/Lasers1111 - Eldeniz AU, Usumez A, Usumez S, Ozturk N. Pulpal temperature rise during light-activated bleaching. J Biomed Mater Res B Appl Biomater. 2005;72:254-9.. It is very difficult to compare the different studies and evaluate their results because of the number of variables present such as time, energy, initial temperature of the pulp chamber, and distance from the photoactivator tip99 - De Moor RJ, Verheyen J, Verheyen P, Diachuk A, Meire MA, De Coster PJ, et al. Laser teeth bleaching: evaluation of eventual side effects on enamel and the pulp and the efficiency in vitro and in vivo. ScientificWorldJournal. 2015;2015:835405..

Although an attempt in vitro was made to simulate the temperature of the pulp chamber, these results should be interpreted with a certain amount of caution. The pulpal blood flow acts as a heat diffuser, with the pulp circulation capable of dissipating some heat prior to damaging the pulp cells1818 - Kodonas K, Gogos C, Tziafa C. Effect of simulated pulpal microcirculation on intrachamber temperature changes following application of various curing units on tooth surface. J Dent. 2009;37:485-90.,2424 - Park SH, Roulet JF, Heintze SD. Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow rates. Oper Dent. 2010;35:353-61.. The current in vitro model was unable to replicate the pulp flow, therefore, it is possible that the increase in temperature was lower. The heating of the bleaching agent not only leads to an increase in temperature in the intra-pulpal chamber, but also a greater penetration of peroxide in the pulp chamber44 - Bowles WH, Ugwuneri Z. Pulp chamber penetration by hydrogen peroxide following vital bleaching procedures. J Endod. 1987;13:375-7..

The diffusion of hydrogen peroxide in the pulp leads to oxidative stress, which could adversely affect the cell metabolism of the pulp66 - Caviedes-Bucheli J, Ariza-Garcia G, Restrepo-Méndez S, Ríos-Osorio N, Lombana N, Muñoz HR. The effect of tooth bleaching on substance P expression in human dental pulp. J Endod. 2008;34:1462-5.. Histological evaluations of extracted human teeth have shown that heat can cause mild inflammatory responses. In some cases, however, it could also be caused by the use of 35% H2O21.

Clinically, patients treated with bleaching agents activated with heat often complain of dental hypersensitivity up to 48 h after the bleaching procedure1212 - Féliz-Matos L, Hernández LM, Abreu N. Dental bleaching techniques; hydrogen-carbamide peroxides and light sources for activation, an update. Mini review article. Open Dent J. 2014;8:264-8..

Light sources have been widely used to accelerate the whitening process, with advantages such as an immediate lower tooth sensitivity2323 - Nutter BJ, Sharif MO, Smith AB, Brunton PA. A clinical study comparing the efficacy of light activated in-surgery whitening versus in-surgery whitening without light activation. J Dent. 2013;41:e3-7.,2626 - Rosa R, Mondelli R. Comparação clínica entre clareamento com e sem fotoativação. DMC Journal. 2009;5:14-7.,2828 - Soares DG, Basso FG, Hebling J, Souza Costa CA. Concentrations of and application protocols for hydrogen peroxide bleaching gels: effects on pulp cell viability and whitening efficacy. J Dent. 2014;42:185-98., besides decreasing the clinical time2121 - Mondelli RF, Azevedo JF, Francisconi AC, Almeida CM, Ishikiriama SK. Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up. J Appl Oral Sci. 2012;20:435-43.. However, more studies are needed to establish the precise correlations between the new light sources and bleaching gels available in the market and its implication to the pulp organ.

CONCLUSION

Within the limitations of the present in vitro study, the results showed that the conventional halogen lamp caused the greatest increase in intra-pulpal temperature and the green LED the lowest.

There was an increase in intra-pulpal temperature with all light sources tested, and the maximum temperature remained below the critical level (5.5°C).The addition of bleaching gel led to a higher increase in intra-pulpal temperatures.

REFERENCES

  • 1
    - Anderson DG, Chiego DJ Jr, Glickman GN, McCauley LK. A clinical assessment of the effects of 10% carbamide peroxide gel on human pulp tissue. J Endod. 1999;25:247-50.
  • 2
    - Baldissara P, Catapano S, Scotti R. Clinical and histological evaluation of thermal injury thresholds in human teeth: a preliminary study. J Oral Rehabil. 1997;24:791-801.
  • 3
    - Bortolatto JF, Pretel H, Neto CS, Andrade MF, Moncada G, Junior OB. Effects of LED-laser hybrid light on bleaching effectiveness and tooth sensitivity: a randomized clinical study. Laser Phys Lett. 2013;10:085601.
  • 4
    - Bowles WH, Ugwuneri Z. Pulp chamber penetration by hydrogen peroxide following vital bleaching procedures. J Endod. 1987;13:375-7.
  • 5
    - Calatayud JO, Calatayud CO, Zaccagnini AO, Box MJ. Clinical efficacy of a bleaching system based on hydrogen peroxide with or without light activation. Eur J Esthet Dent. 2010;5:216-24.
  • 6
    - Caviedes-Bucheli J, Ariza-Garcia G, Restrepo-Méndez S, Ríos-Osorio N, Lombana N, Muñoz HR. The effect of tooth bleaching on substance P expression in human dental pulp. J Endod. 2008;34:1462-5.
  • 7
    - Choi SH, Roulet JF, Heintze SD, Park SH. Influence of cavity preparation, light-curing units, and composite filling on intrapulpal temperature increase in an in vitro tooth model. Oper Dent. 2014;39:E195-205.
  • 8
    - Costa CA, Riehl H, Kina JF, Sacono NT, Hebling J. Human pulp responses to in-office tooth bleaching. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109:e59-64.
  • 9
    - De Moor RJ, Verheyen J, Verheyen P, Diachuk A, Meire MA, De Coster PJ, et al. Laser teeth bleaching: evaluation of eventual side effects on enamel and the pulp and the efficiency in vitro and in vivo ScientificWorldJournal. 2015;2015:835405.
  • 10
    - Domínguez A, García JA, Costela A, Gómez C. Influence of the light source and bleaching gel on the efficacy of the tooth whitening process. Photomed Laser Surg. 2011;29:53-9.
  • 11
    - Eldeniz AU, Usumez A, Usumez S, Ozturk N. Pulpal temperature rise during light-activated bleaching. J Biomed Mater Res B Appl Biomater. 2005;72:254-9.
  • 12
    - Féliz-Matos L, Hernández LM, Abreu N. Dental bleaching techniques; hydrogen-carbamide peroxides and light sources for activation, an update. Mini review article. Open Dent J. 2014;8:264-8.
  • 13
    - Frigo L, Pallota RC, Meneguzzo D, Marcos RL, Penna SC, Lopes-Martins RA. Evaluation of the photo-activated dental bleaching effect on dental pulp in an in vivo rat experimental model. Rev Dental Press Estét. 2009;6:102-14.
  • 14
    - Goldberg M, Grootveld M, Lynch E. Undesirable and adverse effects of tooth-whitening products: a review. Clin Oral Investig. 2010;14:1-10.
  • 15
    - Kabbach W, Zezell DM, Bandeca MC, Pereira TM, Andrade MF. An in vitro thermal analysis during different light-activated hydrogen peroxide bleaching. Laser Phys. 2010;20:1833-7.
  • 16
    - Kabbach W, Zezell DM, Pereira TM, Albero FG, Clavijo VR, Andrade MF. A thermal investigation of dental bleaching in vitro Photomed Laser Surg. 2008;26:489-93.
  • 17
    - Klaric E, Rakic M, Sever I, Tarle Z. Temperature rise during experimental light-activated bleaching. Lasers Med Sci. 2015;30:567-76.
  • 18
    - Kodonas K, Gogos C, Tziafa C. Effect of simulated pulpal microcirculation on intrachamber temperature changes following application of various curing units on tooth surface. J Dent. 2009;37:485-90.
  • 19
    - Kossatz S, Dalanhol AP, Cunha T, Loguercio A, Reis A. Effect of light activation on tooth sensitivity after in-office bleaching. Oper Dent. 2011;36:251-7.
  • 20
    - Marson FC, Sensi LG, Vieira LC, Araújo E. Clinical evaluation of in-office dental bleaching treatments with and without the use of light-activation sources. Oper Dent. 2008;33:15-22.
  • 21
    - Mondelli RF, Azevedo JF, Francisconi AC, Almeida CM, Ishikiriama SK. Comparative clinical study of the effectiveness of different dental bleaching methods - two year follow-up. J Appl Oral Sci. 2012;20:435-43.
  • 22
    - Nam SH, Lee HW, Cho SH, Lee JK, Jeon YC, Kim GC. High-efficiency tooth bleaching using non-thermal atmospheric pressure plasma with low concentration of hydrogen peroxide. J Appl Oral Sci. 2013;21:265-70.
  • 23
    - Nutter BJ, Sharif MO, Smith AB, Brunton PA. A clinical study comparing the efficacy of light activated in-surgery whitening versus in-surgery whitening without light activation. J Dent. 2013;41:e3-7.
  • 24
    - Park SH, Roulet JF, Heintze SD. Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow rates. Oper Dent. 2010;35:353-61.
  • 25
    - Pohto M, Scheinin A. Microscopic observations on living dental pulp II. The effect of thermal irritants on the circulation of the pulp in the lower rat incisor. Acta Odontol Scand. 1958;16:315-27.
  • 26
    - Rosa R, Mondelli R. Comparação clínica entre clareamento com e sem fotoativação. DMC Journal. 2009;5:14-7.
  • 27
    - Sari T, Celik G, Usumez A. Temperature rise in pulp and gel during laser-activated bleaching: in vitro Lasers Med Sci. 2015;30:577-82.
  • 28
    - Soares DG, Basso FG, Hebling J, Souza Costa CA. Concentrations of and application protocols for hydrogen peroxide bleaching gels: effects on pulp cell viability and whitening efficacy. J Dent. 2014;42:185-98.
  • 29
    - Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol. 1965;19:515-30.
  • 30
    - Zhang C, Wang X, Kinoshita J, Zhao B, Toko T, Kimura Y, et al. Effects of KTP laser irradiation, diode laser, and LED on tooth bleaching: a comparative study. Photomed Laser Surg. 2007;25:91-5.

Publication Dates

  • Publication in this collection
    Mar-Apr 2016

History

  • Received
    2 Apr 2015
  • Reviewed
    17 Sept 2015
  • Accepted
    30 Sept 2015
Faculdade De Odontologia De Bauru - USP Serviço de Biblioteca e Documentação FOB-USP, Al. Dr. Octávio Pinheiro Brisolla 9-75, 17012-901 Bauru SP Brasil, Tel.: +55 14 32358373 - Bauru - SP - Brazil
E-mail: jaos@usp.br