Validated microbiological and HPLC methods for the determination of moxifloxacin in pharmaceutical preparations and human plasma

Abstract

The article presents a comparison between microbiological and high performance liquid chromatographic (HPLC) assays for quantification of moxifloxacin in tablets, ophthalmic solutions and human plasma. The microbiological method employed a cylinder-plate agar diffusion assay using a strain of Esherichia coli ATCC 25922 as the test organism and phosphate buffer (pH8) as the diluent. The calibration curves were linear (R²> 0.98) over a concentration range of 0.125 to 16 µgml-1. The within day and between days precisions were < 4.47% and < 6.39% respectively. Recovery values were between 89.4 and 110.2%. The HPLC assay used Hypersil® BDS C18 reversed phase column (250×4.6 mm, 5µm) with a mobile phase comprising 20 mM ammonium dihydrogen orthophosphate (pH3) and acetonitrile (75:25) and flowing at 1.5 ml/min. The detection was at 295nm. The calibration curves were linear (R²> 0.999) over the range of 0.125 to 16 µg ml-1. The within day and between days precisions were < 4.07% and < 5.09% respectively. Recovery values were between 97.7 and 107.6%. Similar potencies were obtained after the analysis of moxifloxacin tablets and ophthalmic solutions by both methods. Also pharmacokinetic parameters were calculated after the analysis of plasma samples of six male healthhy volunteers by both validated methods.

Moxifloxacin; Microbiological assay; HPLC assay; Pharmaceutical analysis; Pharmacokinetics


MEDICAL MICROBIOLOGY

Validated microbiological and HPLC methods for the determination of moxifloxacin in pharmaceutical preparations and human plasma

Ahmed A. Abdelaziz; Tarek E. Elbanna; Noha M. Gamaleldeen* * Corresponding Author. Mailing address: Department of Pharmaceutical Microbiology, Faculty of Pharmacy, University of Tanta, Tanta, Egypt.; Tel.: +2 0106090750 +2 0106090810 Fax: +2 040 3335466.; E-mail: n.gamaleldin@gmail.com

Department of Pharmaceutical Microbiology, Faculty of Pharmacy, University of Tanta, Tanta, Egypt

ABSTRACT

The article presents a comparison between microbiological and high performance liquid chromatographic (HPLC) assays for quantification of moxifloxacin in tablets, ophthalmic solutions and human plasma. The microbiological method employed a cylinder-plate agar diffusion assay using a strain of Esherichia coli ATCC 25922 as the test organism and phosphate buffer (pH8) as the diluent. The calibration curves were linear (R2> 0.98) over a concentration range of 0.125 to 16 µgml-1. The within day and between days precisions were < 4.47% and < 6.39% respectively. Recovery values were between 89.4 and 110.2%. The HPLC assay used Hypersil® BDS C18 reversed phase column (250×4.6 mm, 5µm) with a mobile phase comprising 20 mM ammonium dihydrogen orthophosphate (pH3) and acetonitrile (75:25) and flowing at 1.5 ml/min. The detection was at 295nm. The calibration curves were linear (R2> 0.999) over the range of 0.125 to 16 µg ml-1. The within day and between days precisions were < 4.07% and < 5.09% respectively. Recovery values were between 97.7 and 107.6%. Similar potencies were obtained after the analysis of moxifloxacin tablets and ophthalmic solutions by both methods. Also pharmacokinetic parameters were calculated after the analysis of plasma samples of six male healthhy volunteers by both validated methods.

Key words: Moxifloxacin; Microbiological assay; HPLC assay; Pharmaceutical analysis; Pharmacokinetics.

INTRODUCTION

Since the mid-1980s, the fluoroquinolones (FQs) have become a major group of synthetic antibiotics with activity that ranges from the Enterobacteriaceae and opportunists such as Pseudomonas aeruginosa, to Gram-positive pathogens, including streptococci and staphylococci (4). Moxifloxacin (4aS-cis)-1-Cyclopropyl-6-fluoro-1,4-dihydro-8-methoxy-7-(octahydro-6H-pyrrolo[3,4-b]pyridin-6-yl)-4-oxo-3-quinolinecarboxylic acid, monohydrochloride (Fig.1) (29) is a new fourth generation 8-methoxy FQ developed primarily for the treatment of community acquired pneumonia and upper respiratory tract infections (3). It was also approved for the prevention and treatment of serious ophthalmic infections (e.g. keratitis and endophthalmitis) caused by susceptible bacteria (23). Moxifloxacin is well absorbed from the gastrointestinal tract with absolute bioavailability of approximately 90%. It is 30-50% bound to serum proteins, independent of drug concentration and is widely distributed throughout the body, with tissue concentrations often exceeding plasma concentrations. The pharmacokinetics of moxifloxacin were found to be linear with a wide range of single oral and parental doses from 50 to 800 mg. Maximum plasma concentration (0.29-4.73 mgL-1) is reached after 0.5-4hr (18, 24).


The activity (potency) of antibiotics may be demonstrated under suitable conditions by their inhibitory effect on microorganisms. A reduction in antimicrobial activity will also reveal subtle changes not demonstrable by chemical methods (30). However the assay methodology has changed to chromatographic assays for many antibiotics as they provide quantitative measurements of the purity and the amounts and types of impurities in antibiotics. Nevertheless, a number of commercially important antibiotics continue to require potency determination by microbiological assays (9).

Guerra et al. (11) described a biological assay for analysis of moxifloxacin in tablets. In contrast, several chemical methods for quantitative determination of moxifloxacin in pharmaceutical formulations and biological fluids were described in scientific literature including capillary electrophoresis with laser-induced fluorescence (21), square-wave adsorptive voltammetery (26), spectrophotometery (20), spectrofluorimetery (28), high-performance liquid chromatography (7, 13, 19, 22, 25, 31) amongst others, and ultra HPLC (17). Officially moxifloxacin hydrochloride and moxifloxacin ophthalmic solution were assayed chromatography by an HPLC assay (29).

Method validation is the process that confirms that the analytical procedure employed for a specific test is suitable for its intended use, and it is the corner stone in using the method to evaluate drugs in pharmaceutical industrial committees and regulatory agencies (15). Thus, the scope of this study was to develop and validate an agar diffusion method and HPLC method through the parameters of linearity, precision and accuracy to determine the potency of moxifloxacin in different pharmaceutical formulations as well as monitor the pharmacokinetic parameters of the drug through the analysis of human plasma samples.

EXPERIMENTAL

Materials

Moxifloxacin hydrochloride reference substance (Bayer chemical, Germany) was supplied by Alkan Pharma, Egypt. Avalox® tablet (400 mg moxifloxacin) and Vigamox® sterile ophthalmic solution (5 mg moxifloxacin in each ml) were purchased from the Egyptian markets. Dipotassium hydrogen phosphate and potassium dihydrogen phosphate (BDH, England) were used for buffer preparation. Antibiotic medium No.1 (Difco, England) was purchased for the microbiological assay. Acetonitrile and methanol of HPLC grade (MERCK, Germany), ammonium dihydogen orthophosphate and tinidazole (Reidel de Haein, Germany) were used for the HPLC assay. Water used was double distilled and of high purity.

Methods

Preparation of standard solutions: Moxifloxacin 400 µgml-1 stock solution was prepared by dissolving appropriate amount of moxifloxacin hydrochloride (equivalent to 20 mg moxifloxacin) in phosphate buffer 0.1M (pH8). The working standards of moxifloxacin in concentrations of 2.0, 4.0 and 8.0 µgml-1 were prepared in the same buffer and used for analysis of moxifloxacin in tablet and ophthalmic solution. In addition, eight working standards in the concentration range of 0.125 to 16.0µgml-1 were prepared in human blank plasma. These were employed in the method validation and were used to monitor the pharmacokinetic parameters of the drug.

Microbiological condition:

  • Organism and inoculum: The cultures of

    E.coli ATCC 25922 were cultivated on MacConkey' agar and kept in the refrigerator then pealed to nutrient agar and incubated at 37ºC under aerobic condition 24 hr before the assay. A bacterial suspension having a transmittance of 75% ± 2% (equivalent to 0.125 ± 0.01 optical density) at 620 nm was prepared in saline. Portions of 0.5 ml of the inoculated saline were added to 20 ml of antibiotic medium No. 1 at 48 ºC to 50ºC and used as the inoculated layer.

  • Cylinder-plate assay: The agar plate was composed of two separate layers. Twenty milliliter of antibiotic medium No. 1 agar was poured into (100 × 20 mm) sterile Petri dishes as the base agar. After solidification, portions of 5 ml of the inoculated layer were poured onto the base layer. In each agar plate six stainless steel cylinders (8 × 6 × 10 mm) were placed on the surface of inoculated medium, before loading with 200 µl of standard or sample solutions. After incubation at 37ºC for 18 hr, the zone diameters of the growth inhibition were measured in (mm) using a digital caliper. For the analysis of plasma samples Petri dishes of size (150 mm ×20 mm) were used and the base layer comprised of 40 ml aliquots of antibiotic medium No. 1, onto which 15 ml aliquots of the inoculated medium were poured.

Chromatographic condition: An HPLC system equipped with a variable wavelength UV detector and an automatic injector was used in this study (LC - 10 VP, Shimadzu Scientific Instrument). Separation was accomplished at ambient temperature on a 250 mm × 4.6 mm, 5µ Hypersil® BDS C18 column. The mobile phase consisted of 20 mM Ammonium dihydrogen orthophosphate solution (pH adjusted to 3) and acetonitrile (75: 25) and the flow rate was 1.5 ml/min. The effluent was detected at 295 nm and the peak areas were calculated utilizing the data analysis program class - VP (Shimadzu Scientific Instrument).

Method validation: Both microbiological and HPLC methods were validated according to the FDA industrial guidance for bioanalytical method validation (27).

  • Linearity of the calibration curves: Calibration curves of eight concentrations ranging from 0.125 to 16µgml

    -1 of moxifloxacin in pooled human plasma were conducted daily. The linearity was determined by computing linear least-square regression of the: a- Square of the inhibition zone diameters (y) versus the logarithm of moxifloxacin concentration in the microbiological assay, b- Peak area ratio drug/internal standard (y) versus moxifloxacin concentration in the HPLC assay.

  • Accuracy and precision: Both within day and between days accuracy and precision were considered . Within day accuracy and precision of the assay were determined by the analysis of five replicates of the eight concentrations from 0.125 to 16µgml

    -1 in the same day. For the between days accuracy and precision at least six replicates on six different days during the study period were measured. The % recovery of the back calculated concentrations were compared to the nominal concentrations to determine the accuracy of the assay while the relative standard deviation (RSD) was used to determine the precision of the assay.

  • Quantitation limit: The lowest limit of quantification (LOQ) was taken as the concentration

  • giving a 5 times the response obtained from the blank (if any) or the concentration which is detectable and quantifiable with a precision less than 20 % and accuracy of 80 - 120%.

  • Selectivity: Selectivity was assessed by examining the interference from endogenous substances & buffer solutions.

Quality control samples: The performance of the assay during the analysis of the study samples was evaluated by analysis of three independent low, middle and high concentrations (0.3, 3, 9 µgml-1). These were incorporated in each analysis run and were treated as the unknown samples. The recovery of QC samples determined from the calibration curve and the back calculated concentrations was compared with the nominal values.

Determination of moxifloxacin in tablets and ophthalmic solutions: Three tablets were individually pulverized; transferred to 100 ml volumetric flasks with phosphate buffer( pH8), shaken for 10 min before adjusting the volume. These were suitably diluted to obtain final concentrations of 2, 4 and 8 µgml-1. For the ophthalmic solution, three bottles were employed with 1 ml sample being taken from each. These were diluted with the same buffer and three concentrations were prepared as in case of tablets.

Determination of the pharmacokinetic parameters of moxifloxacin: The study protocol was approved by FPTU' s REC (Faculty of Pharmacy Tanta University Research Ethics Committee). The volunteers were informed by the details of the study and a written concert was obtained from each. The study employed six healthy male volunteers (21-26 years old) with a body weight of 60-75kg. Each person was given a single oral dose (400mg) of the drug. Plasma samples were collected in heparinized tubes at predetermined time intervals from 0 to 48 hrs after drug administration. Plasma was separated by centrifugation and stored at -20 °C. For the microbiological analysis, samples were thawed and vortexed for 30 second, then three replicates of 200 µl volumes from each sample were directly introduced into the assay. For HPLC analysis 250 µl of each sample was transferred to a clean test tube; spiked with 25 µl of the internal standard (250 µgml-1 tinidazole stock solutions), treated with 0.5 ml methanol, shaked for 30 seconds, and centrifuged at 1006 ×g for 7 min. A portion of the supernatant (25 µl) was then injected into the HPLC system.

Calculations

To calculate the activity (potency) of moxifloxacin in pharmaceutical preparations microbiologically, symmetrical three dose levels assay and the Hewitt equation were used (14). The calculation was based on the assumption of two straight parallel lines, one being a plot of responses to standard against the logarithm of dose and the other being a plot of responses to unknown against the logarithm of nominal dose. Regression analysis of the slopes of the two lines was used to verify that there was no deviation from parallelism. The quantitative determination of moxifloxacin in commercial samples by HPLC was carried out using the equation of linear regression of the calibration curves. Also analysis of plasma samples by both methods was carried out by the use of regression equation of the calibration curves then the pharmacokinetic parameters Cmax, Tmax, AUC0→∞ and T1/2 were calculated following a non compartment pharmacokinetic approach (12).

RESULTS AND DISCUSSION

Despite of the existence of alternative methods for determination of moxifloxacin (7, 13, 19, 21, 25, 26, 28, 31) only few of them employed the microbiological assay. Even the reported microbiological assays were not fully validated (11). The current study compared the validity of HPLC versus the microbiological assay for determination of moxifloxacin both in pharmaceutical formulations and biological fluids.

Methods validation

The microbiological and HPLC methods were validated according to the FDA guidance and satisfied all requirements for bioanalytical method validation (27). The validation procedures were carried out in human plasma as biological matrix and the following fundamental parameters were studied:

Linearity of the calibration curves: The calibration curves were linear in the range of concentrations of 0.125 to 16µgml-1 and the individual (R2) values were always ≥ 0.98 for the microbiological assay and ≥ 0.999 for the HPLC assay (Fig. 2).


Accuracy and precision: Within day and between days accuracy values were in the range of 89.4-110.2% and 97.7-107.6% as determined by the microbiological assay and HPLC respectively. Whereas, Within day and between days precision values were in the range of 2.02- 6.39% and 1.25- 5.09% as determined by the microbiological assay and HPLC respectively (Tables 1, 2). These results indicate that both methods are accurate and precise (27).

Quantitation limit: The lower limit of quantitation of moxifloxacin by both methods was 0.125 µgml-1 at which the response was identifiable, discrete, and reproducible with a precision of less than 10% and an accuracy of 80 - 120%. This limit provided sufficient coverage of the in vitro and in vivo application of the methods.

Selectivity: Applying blank plasma and buffer solution samples in the analysis procedures did not show any interference with the original drug responses indicating selectivity of the proposed microbiological (Fig. 3) and HPLC (Fig. 4) methods. This reflects the suitability of the developed methods for determination of the drug in human plasma and makes the methods suitable for monitoring drug bioavailability.



Quality control samples: The recorded data revealed good recovery for all the tested concentrations. This is indicated from the obtained values which were in the range of 97.93-113.3% and 99.48-105.7% of the nominal values in cases of the microbiological and HPLC methods respectively (Table 3). These results comply with the acceptance criteria which require that at least 67% of the QC samples must not deviate by more than 15% of the corresponding nominal value with a possibility that 33% of the samples being outside this range (27).

Determination of moxifloxacin in tablets and ophthalmic solutions

Microbiological method: An experimental 3×3 design, using three dose levels (2, 4 and 8µgml-1) for each standard and sample were analyzed as 6 replicates of each dose level (2, 8). The log dose-response line recorded for the tested samples was parallel to that obtained for the authentic standard concentrations. As a pre request for application of the parallel-line model (2, 8), these findings were verified by means of statistical regression analysis of the slopes of the parallel-lines. Results showed that the two lines were linear with no deviation from parallelism at 95% confidence levels.

Accordingly, the Hewitt equation (14) was employed in calculation of the potency of the drug in the tested formulations. The results are presented in table 6. The results indicated the suitability of the microbiological assay for determination of drug concentration in dosage forms with RSD value of 1.38% between Avalox® samples and 0.40 % between Vigamox® samples (Table 4).

HPLC: The recovered drug values were calculated from the calibration curves which were constructed as before. The results of the determination of moxifloxacin in Avalox® tablets & Vigamox® ophthalmic solution are summarized in Table 5. The method was shown to be precise with RSD values of 1.83% between Avalox® samples and 0.66% between Vigamox® samples.

The efficiency of both methods in determination of drug content of tablets and eye drops was compared. This was achieved using the Student t test. Results revealed similar efficacy as indicated from the non-significant difference between the revealed values (P > 0.05).

Determination of the pharmacokinetic parameters of moxifloxacin

The pharmacokinetic parameters of the drug were determined after oral administration of the drug. This study employed six healthy male volunteers. The drug concentration in the plasma samples was determined using both methods. The recorded plasma concentration versus time plots are shown in Fig. 5.


The calculated pharmacokinetic parameters are presented in Table 6. The maximum plasma concentration (Cmax) of the drug was 3.14 µgml-1 after HPLC determination and was 3.06 µgml-1 after microbiological assay. These were shown to be statistically similar, indicating the suitability of both methods for determination of the pharmacokinetics of moxifloxacin. The other pharmacokinetic parameters revealed the same trend (Table 6). Pharmacokinetic parameters obtained using proposed methods are in agreement with those of the previously reported in the literature reports (6, 18, 32), all summarized in Table 6.

To further prove the similarity of both methods the correlation between the recorded concentrations in both cases was assessed (Fig. 6). This revealed good correlation between the recorded concentrations in each method with a correlation coefficient of 0.909. In earlier reports compared microbiological assays versus HPLC for the determination of FQs in plasma, Griggs and Wise achieved a coefficient of 0.956 (10), while other authors reported coefficients between 0.845 and 0.555 (1, 5, 16).


CONCLUSION

Although biological assays have a high variability incidence, the current study proved that the developed microbiological method for quantitative determination of moxifloxacin was accurate, precise, and sensitive. Thus, it is a valuable method for monitoring the quality of this drug in its different pharmaceutical preparations as well as studying its bioavailability and bioequivalence. Being similar to HPLC in efficacy the microbiological methods can provide an adequate determination of biological potency which offer additional advantage. Finally, the combination of assay methodologies designed to achieve similar results was a novel approach in an analytical laboratory and should be investigated to a greater extent.

Submitted: April 04, 2011

Returned to authors for corrections: November 17, 2011

Approved: June 07, 2012.

  • 1. Auten, G.M.; Preheim, L.C.; Sookpranee, M.; Bittner, M.J.; Sookpranee, T.; Vibhagool, A. (1991). High-pressure liquid chromatography and microbiological assay of serum ofloxacin levels in adults receiving intravenous and oral therapy for skin infections. Antimicrob Agents Chemother 35, 2558-2561.
  • 2. B.P. British Pharmacopeia, Volume IV 5th ed., (2008). (general notices), Her Majesty's Stationary OfficeApendix XIV, London, UK, p. A323-329.
  • 3. Balfour, J.A.; Wiseman, L.R. (1999). Moxifloxacin. Drugs 57, 363-373; discussion 374.
  • 4. Ball, P. (2000). The quinolones: history and overview. In andriole, V.T. (Ed.), The quinolones, 3rd ed. Academic Press, San Diego, p. 2-24.
  • 5. Böttcher, S.; von Baum, H.; Hoppe-Tichy, T.; Benz, C.; Sonntag, H.G. (2001). An HPLC assay and a microbiological assay to determine levofloxacin in soft tissue, bone, bile and serum. J Pharm Biomed Anal 25, 197-203.
  • 6. Clinical Pharmacology, (2009). www.rxlist.com/cgi/generic/ moxifloxacin-cp.htm
  • 7. Djurdjevic, P.; Ciric, A.; Djurdjevic, A.; Stankov, M.J. (2009). Optimization of separation and determination of moxifloxacin and its related substances by RP-HPLC. J Pharm Biomed Anal 50, 117-126.
  • 8. European Pharmacopoeia, 4th ed. (2002). Conseil de l'Europe, Strasbourg, 2.7.2., p.160-166.
  • 9. Greco, G.M. (1997). Microbiological assay systems for the analysis of antibiotics in pharmaceutical formulations. (PhD. Dissertation. the State University of New Jersey. USA).
  • 10. Griggs, D.J.; Wise, R. (1989). A simple isocratic high-pressure liquid chromatographic assay of quinolones in serum. J Antimicrob Chemother 24, 437-445.
  • 11. Guerra, F.L.; Paim, C.S.; Steppe, M.; Schapoval, E.E. (2005). Biological assay and liquid chromatographic method for analysis of moxifloxacin in tablets. J AOAC Int 88, 1086-1092.
  • 12. Hedaya, M.A. (2007). Basic pharmacokinetics. CRC Press, Boca Raton.
  • 13. Hemanth Kumar, A.K.; Ramachandran, G. (2009). Simple and rapid liquid chromatography method for determination of moxifloxacin in plasma. J Chromatogr B Analyt Technol Biomed Life Sci 877, 1205-1208.
  • 14. Hewitt, W. (2004). Microbiological assay for pharmaceutical analysis : a rational approach. Interpharm/CRC, Boca Raton.
  • 15. Huber, L. (2003). Validation of analytical methods and processes. In Nash, R.A., Wachter, A.H. (Eds.). Pharmaceutical process validation 3 ed, New York, USA, p. 507-524.
  • 16. Immanuel, C.; Hemanth Kumar, A.K. (2001). Simple and rapid high-performance liquid chromatography method for the determination of ofloxacin concentrations in plasma and urine. J Chromatogr B Biomed Sci Appl 760, 91-95.
  • 17. Jain, G.K.; Jain, N.; Pathan, S.A.; Akhter, S.; Talegaonkar, S.; Chander, P.; Khar, R.K.; Ahmad, F.J.(2010). Ultra high-pressure liquid chromatographic assay of moxifloxacin in rabbit aqueous humor after topical instillation of moxifloxacin nanoparticles. J Pharm Biomed Anal 52, 110-113.
  • 18. Kim, M.K.; Nightingale, C.H. (2000). Pharmacokinetics and pharmacodynamics of the fluoroquinolones. In andriole, V.T. (Ed.). The quinolones 3rd ed. Academic Press, San Diego, p. 170-191.
  • 19. Liang, H.; Kays, M.B.; Sowinski, K.M. (2002). Separation of levofloxacin, ciprofloxacin, gatifloxacin, moxifloxacin, trovafloxacin and cinoxacin by high-performance liquid chromatography: application to levofloxacin determination in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci 772, 53-63.
  • 20. Motwani, S.K.; Chopra, S.; Ahmad, F.J.; Khar, R.K. (2007). Validated spectrophotometric methods for the estimation of moxifloxacin in bulk and pharmaceutical formulations. Spectrochim Acta A Mol Biomol Spectros 68, 250-256.
  • 21. Mӧller, J.G.; Stass, H.; Heinig, R.; Blaschke, G. (1998). Capillary electrophoresis with laser-induced fluorescence: a routine method to determine moxifloxacin in human body fluids in very small sample volumes. J Chromatogr B Biomed Sci Appl 716, 325-334.
  • 22. Nguyen, H.A.; Grellet, J.; Ba, B.B.; Quentin, C.; Saux, M.C. (2004). Simultaneous determination of levofloxacin, gatifloxacin and moxifloxacin in serum by liquid chromatography with column switching. J Chromatogr B Analyt Technol Biomed Life Sci 810, 77-83.
  • 23. Schlech, B.A.; Alfonso, E. (2005). Overview of the potency of moxifloxacin ophthalmic solution 0.5% (VIGAMOX). Surv Ophthalmol 50 Suppl 1, S7-15.
  • 24. Stass, H.; Dalhoff, A.; Kubitza, D.; Schuhly, U. (1998). Pharmacokinetics, safety, and tolerability of ascending single doses of moxifloxacin, a new 8-methoxy quinolone, administered to healthy subjects. Antimicrob Agents Chemother 42, 2060-2065.
  • 25. Tatar Ulu, S. (2007). High-performance liquid chromatography assay for moxifloxacin: pharmacokinetics in human plasma. J Pharm Biomed Anal 43, 320-324.
  • 26. Trindade, M.A.G.; da Silva, G.M.; Ferreira, V.S. (2005). Determination of moxifloxacin in tablets and human urine by square-wave adsorptive voltammetry. Microchemical Journal 81, 209-216.
  • 27. U.S. Food and Drug Administration, Guidance for Industry, Bioanalytical Method Validation, (2001). U.S. Department of Health and Human Services, Center for Drug Evaluation and Research (CDER), Centre of Veterinary Medicine (CVM), Rockville, USA Internet:// www.fda.gov/cvm
  • 28. Ulu, S.T. (2009). Spectrofluorimetric determination of fluoroquinolones in pharmaceutical preparations. Spectrochim Acta A Mol Biomol Spectrosc 72, 138-143.
  • 29. US Pharmacopoeia & National Formulary (2009). Second supplement, Moxifloxacin, Official Monographs, USP32-NF27, US Pharmacopeial Convention Inc., Rockville, p. 4264-4267.
  • 30
    US Pharmacopoeia & National Formulary (2009). ‹81› Antibiotics-Microbial assays, USP32-NF27, US Pharmacopeial Convention Inc.,Rockville, p. 86-93.
  • 31. Wagenlehner, F.M.; Kees, F.; Weidner, W.; Wagenlehner, C.; Naber, K.G. (2008). Concentrations of moxifloxacin in plasma and urine, and penetration into prostatic fluid and ejaculate, following single oral administration of 400 mg to healthy volunteers. Int J Antimicrob Agents 31, 21-26.
  • 32. Zhanel, G.G.; Noreddin, A.M. (2001). Pharmacokinetics and pharmacodynamics of the new fluoroquinolones: focus on respiratory infections. Curr Opin Pharmacol 1, 459-463.
  • *
    Corresponding Author. Mailing address: Department of Pharmaceutical Microbiology, Faculty of Pharmacy, University of Tanta, Tanta, Egypt.; Tel.: +2 0106090750 +2 0106090810 Fax: +2 040 3335466.; E-mail:
  • Publication Dates

    • Publication in this collection
      19 Feb 2013
    • Date of issue
      Dec 2012

    History

    • Received
      04 Apr 2011
    • Accepted
      07 June 2012
    • Reviewed
      17 Nov 2011
    Sociedade Brasileira de Microbiologia USP - ICB III - Dep. de Microbiologia, Sociedade Brasileira de Microbiologia, Av. Prof. Lineu Prestes, 2415, Cidade Universitária, 05508-900 São Paulo, SP - Brasil, Ramal USP 7979, Tel. / Fax: (55 11) 3813-9647 ou 3037-7095 - São Paulo - SP - Brazil
    E-mail: bjm@sbmicrobiologia.org.br