Evaluation of the automated system Vitek2 for identification and antimicrobial susceptibility testing of Brazilian Gram-positive cocci strains

Abstract

Automated instruments offer many advantages for clinical laboratories. Nevertheless, they can have problems identifying and determining susceptibilities of some pathogens. Vitek® 2 (bioMérieux) is an automated system that was recently introduced to Brazil. We evaluated the performance of this equipment for Brazilian isolates that had been characterized using reference identification and antimicrobial susceptibility testing methods. Ninety-nine strains of Gram-positive cocci from a local reference center collection were analyzed, consisting of 50 coagulasenegative Staphylococcus (CoNS) and 49 Enterococcus and related species. Vitek® 2 correctly identified 79.8% (79/99) of the isolates. Oxacillin resistance was detected in 76% (19/25) of resistant S. epidermidis strains and in 88% (22/25) of other resistant CoNS species strains. Vancomycin resistance was detected in 100% (20/20) of resistant Enterococcus and related species strains. Vitek® 2 performed very well for the identification of S. epidermidis and non-epidermidis staphylococci, and for the detection of vancomycin resistance in Enterococcus and related species. However, the system needs improvement in order to provide reliable results for the characterization of some CoNS species, identification of Enterococcus and related species and for detecting oxacillin resistance in CoNS.

Automated system; Vitek2; Gram-positive cocci; identification; antimicrobial susceptibility testing


Evaluation of the automated system Vitek2 for identification and antimicrobial susceptibility testing of Brazilian Gram-positive cocci strains

Pedro Alves d'AzevedoI; Itacy SiquieraII; Juliana GugelII; Ana Lúcia Souza AntunesI; Carina SecchiI; Jacyr PasternakII; Marinês Dalla Vale MartinoII

IDepartment of Microbiology and Parasitology of the Foundation of Clinical Medical School of Porto Alegre (FFFCMPA), Porto Alegre, RS

IIClinical Laboratory of Hospital Israelita Albert Einstein; São Paulo, SP, Brazil

Address for correspondence Address for correspondence: Dr. Pedro A. d'Azevedo Universidade Federal de Ciências da Saúde de Porto Alegre Rua Sarmento Leite 245/204 Porto Alegre, RS. Zip code: 90050-170, Brazil Phone: 55 51 33038740. Fax: 55 51 32269756 E-mail: pedroaze@ufcspa.edu.br

ABSTRACT

Automated instruments offer many advantages for clinical laboratories. Nevertheless, they can have problems identifying and determining susceptibilities of some pathogens. Vitek® 2 (bioMérieux) is an automated system that was recently introduced to Brazil. We evaluated the performance of this equipment for Brazilian isolates that had been characterized using reference identification and antimicrobial susceptibility testing methods. Ninety-nine strains of Gram-positive cocci from a local reference center collection were analyzed, consisting of 50 coagulasenegative Staphylococcus (CoNS) and 49 Enterococcus and related species. Vitek® 2 correctly identified 79.8% (79/99) of the isolates. Oxacillin resistance was detected in 76% (19/25) of resistant S. epidermidis strains and in 88% (22/25) of other resistant CoNS species strains. Vancomycin resistance was detected in 100% (20/20) of resistant Enterococcus and related species strains. Vitek® 2 performed very well for the identification of S. epidermidis and non-epidermidis staphylococci, and for the detection of vancomycin resistance in Enterococcus and related species. However, the system needs improvement in order to provide reliable results for the characterization of some CoNS species, identification of Enterococcus and related species and for detecting oxacillin resistance in CoNS.

Key-Words: Automated system, Vitek2, Gram-positive cocci, identification, antimicrobial susceptibility testing.

Automated instruments offer many advantages in clinical laboratories, especially in hospitals. Such instruments improve workflow and provide faster results than conventional methods. Nevertheless, they can have problems identifying and determining susceptibilities of some pathogens. Coagulase-negative staphylococci (CoNS) are the most common microorganisms isolated from blood cultures; however, 85% of the isolates are contaminants, usually from skin contamination at the time of collection [1]. Nevertheless, CoNS are recognized as important nosocomial pathogens; in many institutions, they are among the main agents of nosocomial bacteremias [2-4]. Many clinical laboratories do not routinely identify CoNS to the species level when these microorganisms are detected in blood or other body fluids. However, the significance of CoNS as pathogens has increased. Blood culture isolates should therefore be identified to the species level to determine their clinical relevance and monitor their epidemiology [5]. Resistance to methicillin among these microorganisms is a matter of concern, because of the increasingly high levels that have been detected. In a multicenter study in Brazil, methicillin resistance was observed in 87.7% of CoNS isolated from bloodstream infections [6]. Accurate detection of methicillin resistance among CoNS isolates in the clinical laboratory is important to guide therapy and to promote the correct use of glycopeptides [3,7]. The importance of the enterococci as nosocomial pathogens has been widely documented in recent years; vancomycinresistant enterococci (VRE) are particularly relevant due to their increasing prevalence and their capability to acquire antimicrobial resistance. Enterococcus faecalis is the species most frequently associated with human infections, followed by Enterococcus faecium [8].

In general, these two species are responsible for about 90 to 95% of enterococcal infections in humans. The remaining 5 to 10% are caused by other members of the genus.

Automated systems, such as Vitek® legacy (bioMérieux) and MicroScan® (DadeBehring), are also frequently used for rapid identification and antimicrobial susceptibility testing of Gram-positive cocci in clinical laboratories around the world [9-13]. Vitek®2 (bioMérieux) is an automated system recently introduced to Brazil. We evaluated the performance of this system with Brazilian isolates that had been characterized using reference methods of identification (ID) and antimicrobial susceptibility testing (AST).

Material and Methods

Isolates

Ninety-nine isolates of Gram-positive cocci from a local reference center collection were analyzed: 50 Coagulasenegative Staphylococcus (CoNS) and 49 Enterococcus spp. and related species. They were obtained from a culture collection maintained in the Microbiology department, Universidade Federal de Ciências da Saúde de Porto Alegre. The isolates were kept at -20ºC in skim milk (Difco) plus 20% glycerol.

Quality control of the tests was done using Enterococcus faecalis ATCC 29212, Enterococcus faecium SS 1274, Staphylococcus epidermidis ATCC 12228, Staphylococcus saprophyticcus CCM 883, Staphylococcus hominis ATCC 27844 and Staphylococcus haemolyticcus CCM 2737.

The isolates were identified by conventional biochemical tests [8,14].

Most isolates were also identified by analysis of electrophoretic whole-cell protein profiles according to the procedures described by Merquior et al. [15] for enterococci and by molecular (sodA sequencing) methods for CoNS [16]. Conventional antimicrobial susceptibility testing - Susceptibilities to oxacillin and vancomycin were detected by the disk-diffusion (DD), Etest® (AB Biodisk, Solna, Suécia) and agar dilution (CLSI) [17] procedures. The mecA and vanA alleles were detected by PCR; this was considered the "gold standard" method [18,19].

Identification and Antimicrobial Susceptibility Testing Using Automated Systems

Species identification and antimicrobial susceptibility were determined using the automated systems Vitek2® (bioMérieux - software 4,03), panel GP card (identification) and GP01 (susceptibility testing). The manufacturer's instructions were followed for the preparation of the inoculua and incubation of the isolates.

Results

Based on the combination of results obtained from the conventional biochemical tests, the 99 isolates of CoNS, enterococci and related genera were identified as follows: CoNS - 25 S. epidermidis, 3 S. haemolyticus, 3 S. sciuri, 2 S. warneri, 2 S. cohnii subsp. cohnii, 2 S. capitis subsp. capitis, 2 S. hominis subsp. hominis, 2 S. hominis subsp. novobiosepticus, 2 S. cohnii subsp. urealyticus, 1 S. caprae, 1 S. auricularis, 1 S. equorum, 1 S. lugdunensis, 1 S. saprophyticus, 1 S. simulans and 1 S. xilosus. Enterococci and related genera - 17 E. faecium, 9 E. faecalis, 7 E. gallinarum, 3 E. avium, 2 E. casseliflavus, 2 E. hirae, 1 E. durans, 1 E. raffinosus, 1 Lactococcus garvieae, 2 Leuconostoc pseudomesenteroides, 2 Streptococcus bovis and 2 Pediococcus sp.

Vitek2 correctly identified 80 of the 99 isolates (accuracy of 80.8%), 43/50 (86%) of CoNS and 37/49 (75.5%) of Enterococcus sp. and related species. Among CoNS, the identifications were concordant for S. epidermidis, S. caprae, S. capitis subsp. capitis, S. cohnii subsp. urealyticus, S. lugdunensis, S. haemolyticus, S. hominis subsp. hominis, S. sciuri, and S. warneri. The system failed to identify the other species and subspecies tested. Among Enterococcus sp. and related species the identifications were as follows: Enterococcus sp. 71.4% (30/42), Pediococcus sp. 0% (0/2), Leuconostoc pseudomesenteroides 100% (2/2), Lactococcus garviae 100% (1/1), and Streptococcus bovis 0% (0/2) (Table 1). Methicillin resistance was associated with the mecA gene in 19 of the 25 samples (76.0%) in S. epidermidis isolates and in 22/25 (88%) in other CoNS species. Vancomycin resistance was detected in 100% of the Enterococcus species and related genera (20/20) (Tables 1 and 2).

Discussion

Automated methods are faster and improve laboratory workflow. Most of the discrepant results in identification involved the less-frequently-isolated species. We need to be cautious when CoNS other than S. epidermidis are identified with this system. Nonhoff et al. (2005) [20] found similar limitations when he tested Vitek®2 on methicillin-resistant staphylococci from Belgium. We must also be cautious with the less frequent Enterococcus sp. and related species. Vitek®2 performed well and identified these two species; it correctly identified 25/26 (96.2%) of E. faecalis and E. faecium, which are normally responsible for approximately 90% to 95% of enterococcal infections in humans.

For less-frequently-isolated species, Vitek®2 may need improvement. A major mistake was made with E. gallinarum, an important vancomycin-resistant pathogen, which should be correctly identified in a clinical hospital laboratory.

The system had good performance in the determination of methicillin resistance, especially for S. non-epidermidis 22/25 (88%). However, the interpretative criteria of the CLSI (2005) may overestimate resistance of the other species. The discrepancies in the susceptibility tests for CoNS included major errors. The Minimum Inhibitory Concentration (MICs) determination of the strains that showed false resistance were near the established breakpoints to classify them as susceptible. According to the CLSI (2008) [17], strains isolated from serious infections that have MICs varying from 0.5 to 2.0 µL/mL should be tested for the presence of the mecA gene or for the protein expressed by this gene, considering that they may present confusing phenotypes. Less frequent species of CoNS have been associated with serious infections in hospital institutions, and these have become increasingly common.

The advantages of automated systems for species identification and antimicrobial susceptibility testing are speed and better workflow. Improvements in their accuracy would help make them practical.

In the analysis of vancomycin resistance in Enterococcus sp. and related species the automated system detected 20/20 isolates (100%). Considering that vancomycin-resistant enterococci require implementation of infection control measures, some species identification is needed.

The reporting time for CoNS identification by the VITEK 2 system ranged from 4.25h to 8h, and the mean time to result was 5.79h. Enterococci required from 3.25 h to 8 h to identify; the mean time was 5.78 h.

In conclusion, the automated system Vitek2 needs further improvement in order to provide reliable results for the characterization of the other CoNS and enterococci-related species and for detection of oxacillin resistance.

Acknowledgements

The authors thank bioMérieux, the Microbiology group of the Hospital Israelita Albert Einstein, and the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) for financial technical support.

Received on 10 May 2008; revised 12 December 2008.

This study was presented in part at the 107th General Meeting of the American Society for Microbiology, Toronto, Canada, 2007. Financial support: CNPq, FFFCMPA.

  • 1. Weinstein M.P., Towns M.L., Quartey S.M., et al. The clinical significance of positive blood culture in the 1990s: a prospective comprehensive evaluation of the microbiology, epidemiology and outcome of bacteremia and fungemia in adults. Clin Infect Dis 1997;24:584-602
  • 2. Cockerill F.R., III J.G., Hughes E.A., Vetter E.A., et al. Analysis of 281797 consecutive blood cultures performed over an eight-year period: trends in croorganisms isolated and the value of anaerobic culture of blood. Clin Infect Dis 1997;24:403-18.
  • 3. Hussain Z., Stoakes L., Lannigan R., et al. Evaluation of Screening and Commercial Methods for Detection of Methicillin Resistance in Coagulase-Negative Staphylococci. J Clin Microbiol 1998;36:273-4.
  • 4. Petinaki E., Kontos F., Miriagou V., et al. The bacterial resistance group 2001. Survey of methicillin-resistant coagulase-negative staphylococci in the hospitals of central Greece. Int J Antimicrob Agents 2001;18:563-6.
  • 5. Kloos W.E., Bannerman T.L. Update on clinical significance of coagulase-negative staphylococci. Clin Microbiol Rev 1994;7:117-40.
  • 6. Sader H.S., Sampaio J.L.M., Zocolli C., Jones R.N. Results of the 1997 SENTRY antimicrobial surveillance program in three Brazilian medical centers. Braz J Infect Dis 1999;3:63-79.
  • 7. Yamazumi T.I.F., Diekema D.J., Pfaller M.A., Jones R.N. Comparison of the Vitek Gram-Positive Susceptibility 106 Card, the MRSA-Screen Latex Agglutination Test, and mecA Analysis for Detecting Oxacillin Resistance in a Geographically Diverse Collection of Clinical Isolates of Coagulase-Negative Staphylococci. J Clin Microbiol 2001;39:3633-6.
  • 8. Facklam R.R., Sahm D., Teixeira L.M. Enterococcus, p.297-305. In PR Murray, EJ Baron, JH Jorgensen, MA Pfaller, RH Yolken (eds), Manual of Clinical Microbiology, American Society of Microbiology, Washington, D.C., 1999
  • 9. Jett B., Free L., Sahm D.F. Factors influencing the Vitek Gram-positive susceptibility system's detection vanB encoded vancomycin-resistance among enterococci. J Clin Microbiol 1996;34:701-6.
  • 10. d'Azevedo P.A., Dias C.A.G., Gonçalves A.L.S., et al. Evaluation of an automated system for the identication and antimicrobial susceptibility testing of enterococci. Diag Microbiol Infect Dis 2001;40:157-61.
  • 11. Martinez F., Chandler L.J., Reisner B.S., Woods G.L. Evaluation of the Vitek card and VTK-RO7.01 software for detection of oxacillin resistance in clinically relevant coagulase-negative Staphylococci J Clin Microb 2001;39:3733-5.
  • 12. d'Azevedo P.A., Cantarelli V., Inamine E., et al. Avaliação de um sistema automatizado na identificação de espécies de Enterococcus J Bras Patol Med Lab 2004;40:237-9.
  • 13. Caierão J., Superti S., Dias C.A.G., d'Azevedo P.A. Automated systems in the identification and determination of methicillin resistance among coagulase negative staphylococci. Mem Inst Oswaldo Cruz 2006;101:277-9.
  • 14. Bannerman T.L. Staphylococcus, Micrococcus, and other catalasepositive cocci that grow aerobically, p. 384-404. In: PR Murray, EJ Baron, JH Jorgensen, MA Pfaller, RH Yolken (eds), Manual of Clinical Microbiology, American Society of Microbiology, Washington, D.C., 2003
  • 15. Merquior V.L.C., Peralta J.M., Facklam R.R., Teixeira L.M. Analysis of electrophoretic whole-cell protein profiles as a tool for characterization of Enterococcus species. Curr Microbiol 1994;28:149-53.
  • 16. Poyart C., Quesne G., Boumaila C., Trieu-Cuot P. Rapid and Accute Species-Level identification of Coagulase-Negative Staphylococci by using the sodA Gene as a Target. J Clin Microbiol 2001;39:4296-301.
  • 17. CLSI - Clinical and Laboratory Standards Institute. Performance standards for antimicrobial susceptibility testing - Fifteenth informational supplement. document M100-S18. Clinical and Laboratory Standards Institute, Wayne, PA, US, 2008
  • 18. Clarck N.C., Teixeira L.M., Facklam R.R., Tenover F. Detection and differentiation of vanC-1, vanC-2, and vanC-3 glycopeptide resistance genes in enterococci. J Clin Microbiol 1998;36:2294-7.
  • 19. Vannuffel P., Gigi J., Ezzedine H., et al. Specific detection of methicilin resistant Staphylococcus species by multiplex PCR. J Clin Microbiol 1998;33:2864-7.
  • 20. Nonhoff C., Rottiers S., Struelens M.J. Evaluation on the Vitek 2 system for identification and antimicrobial susceptibility testing of Staphylococcus spp. Clin Microbiol Infect 2005;11:150-3.
  • Address for correspondence:
    Dr. Pedro A. d'Azevedo
    Universidade Federal de Ciências da Saúde de Porto Alegre
    Rua Sarmento Leite 245/204
    Porto Alegre, RS. Zip code: 90050-170, Brazil
    Phone: 55 51 33038740. Fax: 55 51 32269756
    E-mail:
  • Publication Dates

    • Publication in this collection
      01 Feb 2010
    • Date of issue
      Apr 2009

    History

    • Reviewed
      12 Dec 2008
    • Received
      10 May 2008
    Brazilian Society of Infectious Diseases Rua Augusto Viana, SN, 6º., 40110-060 Salvador - Bahia - Brazil, Telefax: (55 71) 3283-8172, Fax: (55 71) 3247-2756 - Salvador - BA - Brazil
    E-mail: bjid@bjid.org.br