In-Line Incomplete Solid-Liquid Extraction for Spectrophotometric Determination of Soluble Inorganic Phosphorus in Plants and Foods

Abstract

Phosphorus determination in plant materials and foods is important. For accessible inorganic P assessment, solid-liquid extraction (SLE) is required. In this work, a flow-based analytical procedure employing zones penetration approach was developed aiming at the soluble inorganic phosphorus (PI) determination in plants and foods. The analyte was quantified by the spectrophotometric molybdenum blue method. Linear response was observed between 2.0 and 30 mg L-1 PI, and detection limit was estimated at 0.50 mg L-1 PI (99.7% confidence level, n = 20). Determination rate was estimated at 20 h-1 using in-line extraction with 20 mg of sample. Precise timing of the flow system allowed the use of incomplete SLE for the first time prior to PI determination. The online extraction was carried out for 60 s, leading to an extraction efficiency of 65 ± 3% in comparison to batch quantitative extraction. The results obtained by the proposed procedure were in agreement with those obtained with batch extractions (95% confidence level).

Keywords:
flow analysis; incomplete extraction; phosphorus; sample preparation; vegetable material


Introduction

Phosphorus is an important nutrient for plants and animals because it participates in many biochemical processes.11 Schachtman, D.; Reid, R.; Ayling, S. M.; Plant Physiol. 1998, 116, 447.,22 Reddy, N. R. In Food Phytates, 1st ed.; Reddy, N. R.; Sathe, S. K., eds.; Taylor & Francis Group: Boca Raton, USA, 2001, p. 25. This element composes prominent biomolecules such as adenosine triphosphate (ATP), phospholipids, nucleic acids, coenzymes and inositol phosphates. Phosphorus is absorbed by living organisms mainly as orthophosphate (H2PO4- and HPO42- depending on pH), which corresponds to non-condensed inorganic phosphorus in the water soluble fraction (PI). Regarding plant nutrition, the enzymatic conversion of organic phosphorus by phytase increases PI in soils. On the other hand, phosphate leaching by rain and irrigation waters decreases PI content, requiring the use of fertilizers for nutrient maintenance.

One alternative for inorganic and organic fertilizers are green manures (e.g. Crotalaria juncea),33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. which basically are plants that recover soil fertility by increasing nutrient content and ion exchange capacity. Prior to a new sowing, the plants are felled, and their residues remain on the soil as nutrient supply. Therefore, determination of PI in green manures is important for agronomic studies. As PI is also readily bioaccessible for animals, its determination in vegetable derived foodstuffs (e.g. leguminous grains and flours) is also important for human nutritional studies.

Determination of PI in vegetable materials and solid foodstuffs requires solid-liquid extraction (SLE) of the analyte from the solid sample, which is highly susceptible to analyte losses and sample contamination.33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207.,44 Rocha, D. L.; Batista, A. D.; Rocha, F. R. P.; Donati, G. L.; Nóbrega, J. A.; TrAC, Trends Anal. Chem. 2013, 45, 79. Additionally, SLE generally requires tiring and time-consuming steps such as centrifugation and filtration. The use of aqueous medium is recurrent for PI extraction, but similar efficiencies for PI removal have been reported with different extraction times and extractant composition.33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207.,55 Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139.,66 Hogue, E.; Wilcox, G. E.; Cantliffe, D. J.; J. Am. Soc. Hortic. Sci. 1970, 95, 174. A common procedure for PI extraction from plant materials is carried out with HClO4,66 Hogue, E.; Wilcox, G. E.; Cantliffe, D. J.; J. Am. Soc. Hortic. Sci. 1970, 95, 174. being the organic matter removed with active charcoal. PI is then determined with molybdenum blue method. One of the drawbacks is related to acidic waste generation containing perchloric acid. Alternatively, trichloroacetic acid (TCA) has been proposed for PI extraction from plant materials, but systematic errors have been reported77 Noack, S. R.; Smernik, R. J.; McBeath, T. M.; Armstrong, R. D.; McLaughlin, M. J.; Talanta 2014, 126, 122. due to condensed phosphorus extraction and hydrolysis caused by the hydrophobicity and the low pH of TCA solution, respectively.

The replacement of acids by water for PI extraction was also proposed.33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. SLE was carried out for 10 min, thus making this step cleaner and 6-fold faster than previous procedure employing acidic extraction. Afterwards, PI was directly determined in a monosegmented flow system with spectrophotometric detection without the clean-up step with active charcoal. In spite of that, sample preparation was not mechanized, which increases inaccuracy sources inherent to batch extractions.

Flow-based systems have been successfully employed in a versatile and robust way to minimize errors in SLE.88 Trojanowicz, M.; Kolacinska, K.; Analyst 2016, 141, 2085. These systems overcome drawbacks of batch procedures, minimizing analyte losses, sample contamination, reagent consumption, and waste generation.88 Trojanowicz, M.; Kolacinska, K.; Analyst 2016, 141, 2085. Additionally, analytical frequency and precision are enhanced in procedures employing slow reactions because detection can be carried out without reaching chemical equilibrium, also allowing the exploitation of kinetic aspects. Therefore, analysis can be performed with incomplete decomposition and extraction steps.99 Rocha, D. L.; Kamogawa, M. Y.; Rocha, F. R. P.; Anal. Chim. Acta 2015, 896, 11.

In-line procedures aiming at analyte transference from solid sample to liquid phase have been scarcely exploited.88 Trojanowicz, M.; Kolacinska, K.; Analyst 2016, 141, 2085. Elemental determinations in pharmaceuticals1010 Sweileh, J. A.; Microchem. J. 2000, 65, 87. and in phosphate rocks1111 Sweileh, J. A.; Anal. Chim. Acta 2007, 581, 168. have been carried out after the in-line extraction using a suspension formed in a lab-made device, and carried through the analytical path. In-line filtration step and addition of suspension stabilizers were necessary. The formation of suspension has also been exploited for determination of exchangeable potassium in soils.1212 Almeida, M. I. G. S.; Segundo, M. A.; Lima, J. L. F. C.; Rangel, A. O. S. S.; Microchem. J. 2006, 83, 75. In this case, the sample was transferred to an open vessel for addition of the extractant followed by the formation of the suspension. After magnetic stirring for online SLE, extracts were aspirated to a flow system for potassium determination using paper filter at the tip of the aspirating tubes.

Chambers that also entrap suspension and allow the use of smaller amounts of sample have been recently proposed for in-line SLE of nitrate from vegetables.1313 Galvis-Sánchez, A. C.; Santos, J. R.; Rangel, A. O. S. S.; Anal. Bioanal. Chem. 2016, 408, 7651. A removable part of the apparatus allows sample replacement without disconnecting the whole chamber of the flow system. Additionally, magnetic stir bars were used to improve extraction. Despite appropriate for quantitative SLE extraction, incomplete processes have not been evaluated.

Simpler sample chambers have been proposed for determination of ammonium in building materials,1414 Timofeeva, I. I.; Bulatov, A. V.; Moskvin, A. L.; Kolev, S. D.; Talanta 2015, 142, 140. and for bioavailability studies of antioxidants in foods,1515 Maia, M. A.; Soares, T. R. P.; Mota, A. I. P.; Rosende, M.; Magalhães, L. M.; Miró, M.; Segundo, M. A.; Talanta 2017, 166, 162. and of elements in soils1616 Chomchoei, R.; Miró, M.; Hansen, E. H.; Shiowatana, J.; Anal. Chim. Acta 2005, 536, 183. and wastes.1717 Rosende, M.; Miró, M.; Cerdà, V.; Anal. Chim. Acta 2010, 658, 41. These applications showed some drawbacks such as the need for manual addition of reagents1414 Timofeeva, I. I.; Bulatov, A. V.; Moskvin, A. L.; Kolev, S. D.; Talanta 2015, 142, 140. and collection of extracts in open vessels prior to determinations.1515 Maia, M. A.; Soares, T. R. P.; Mota, A. I. P.; Rosende, M.; Magalhães, L. M.; Miró, M.; Segundo, M. A.; Talanta 2017, 166, 162.

16 Chomchoei, R.; Miró, M.; Hansen, E. H.; Shiowatana, J.; Anal. Chim. Acta 2005, 536, 183.
-1717 Rosende, M.; Miró, M.; Cerdà, V.; Anal. Chim. Acta 2010, 658, 41. Mechanized procedures exploiting dynamic SLE followed by in-line dilution and determination, for example, have not been exploited. In this regard, incomplete processes for SLE has also not been studied. In this work, an analytical procedure is proposed with incomplete in-line extraction of PI from vegetable materials and leguminous foods prior to direct determination with spectrophotometric molybdenum blue method.

Experimental

Apparatus

The flow manifold comprised an 8-channel peristaltic pump model IPC-8 (Ismatec, Cole Parmer, Wertheim, Germany) equipped with 1.85 mm i.d. Tygon® tubes (Ismatec), polyethylene tubes (0.76 mm i.d.), a pair of acrylic proportional injectors and a PEEK® 6-way confluence joint. The sample chamber was constructed with a transparent polyvinyl chloride (PVC)-based Tygon® tube (4.0 cm length and 4 mm i.d.) with glass wools in both tips to avoid the escaping of the solid material. This tube was connected to the flow system aiming at dynamic extractions.

Measurements were carried out with a modular multichannel spectrophotometer (Ocean Optics USB4000, Dunedin, FL, USA) composed by a tungsten-halogen light source, a charge-coupled device spectrometer (200-800 nm), a pair of 600 µm i.d. optical fibers, and a sample holder with a 10 mm optical path flow cell (Hellma, Müllheim, Germany; 80 µL internal volume). The software SpectraSuite® (Ocean Optics) was employed for data acquisition using an HP Intel Core i3 computer. A magnetic stirrer (Quimis, Diadema, SP, Brazil) and a centrifuge (Hettich, Kirchlengern, Germany) were also employed.

Reagents, solutions, and samples

All solutions were prepared with deionized water and analytical grade chemicals. The R1 reagent was a 12 mmol L-1 (NH4)6Mo7O24.4H2O (Sigma, St. Louis, MO, USA) solution in 0.30 mol L-1 HNO3 (Sigma). A 60 mmol L-1 ascorbic acid (R2, Sigma) was daily prepared by dissolving the reagent in water. Stock solution of 1000 mg L-1 PI as KH2PO4 (Sigma) was prepared by dissolution of appropriate amounts of the salt in water. Reference solutions of PI were prepared by adequate dilution of the stock in water. Therefore, water was used as blank throughout optimization.

Soy flour and beans were purchased in a local market, and plant materials (Crotalaria juncea, coffee and fig trees) were provided by Center for Nuclear Energy in Agriculture (University of São Paulo, Brazil). These samples were cut in a knife mill (particles < 60 µm) and dried at 80 ºC until they have the same weight. Afterwards, they were kept in a desiccator. Prior to in-line extraction, no further sample treatment was carried out.

Proposed procedure

For the extraction and determination of PI, the flow manifold shown in Figures 1a and 1b was operated. In the position showed in Figure 1a, the sample chamber (SC) containing 20 mg of sample, and coil B1 (which defined the extractant volume) were filled with 2.5 mL of extractant (deionized water) at 2.5 mL min-1. The pump timer was set to 60 s, which was the time required to fill the extraction path, and the pump was automatically turned off after that period. Afterwards, the pump was restarted after injector Inj1 was commuted in the direction represented by the white arrow in Figure 1a. In this position, water was allowed to recirculate for 60 s in the extraction path for PI extraction. The time was also controlled by the pump timer and the device was automatically turned off at the end of the extraction. Then, Inj1 was returned to the previous position and, after turning on the pump, water was used to transport the extract towards sample loop L1 in Inj2 by using a 50-cm coil (B2) between the injectors. By transporting the extract for 70 s (at 2.5 mL min-1), the portion of the extract zone filling the loop was proportional to the necessary dilution (2-fold when collection of the extract was performed batchwise) to avoid PI concentrations out of the linear range for PI determination. The time for this step was also controlled by the pump timer to avoid imprecise sampling. Meanwhile, reagent loops L2 and L3 were filled with R1 and R2 solutions, respectively. Pumping was restarted after injector Inj2 was commuted towards the direction showed by the white arrow in Figure 1b. The sample zone was formed by zones penetration and directed towards B3 and spectrophotometric detection at 730 nm using water as carrier. Sample replacement was carried out by removing the sample chamber from the system, discarding the suspension, and washing the tube with 1% HNO3 and deionized water prior to the introduction of the next sample. Due to sample packing, the rising of air bubbles in the analytical path was rare. For each entrapment of sample, it was possible to register one response. Therefore, for analysis in triplicate, three different portions of the same sample had to be employed (one response for each extraction).

Figure 1
Flow system manifold for (a) extraction and (b) determination of soluble inorganic phosphorus. E: extract; R1: (NH4)6Mo7O24.4H2O + HNO3; R2: ascorbic acid; C: H2O; W: waste; Inj1 and Inj2: proportional injectors; L1-L3: sample and reagent loops; B1-B3: coils (90, 50, and 150 cm, respectively); SC: sample chamber; P: peristaltic pump; D: spectrophotometric detection cell; x: confluence joint. The white arrows indicate the commuting direction of the injectors.

Extraction efficiency (EE) was evaluated by comparing PI content obtained with dynamic and batch sample preparations, considering the latter yielded quantitative extraction. Accuracy was evaluated by comparing the results obtained with the developed and batch procedures.

For PI determination, chemical and physical parameters were optimized using the flow injection system shown in Figure 1b with the univariate method, aiming at maximum sensitivity and determination rate, better precision and minimum reagent consumption and waste generation. The initial concentrations and volumes were as reported before in the literature for PI determination:55 Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139. 25 µL of 0.32 mmol L-1 P (10 mg L-1), 20 µL of 7.5 mmol L-1 (NH4)6Mo7O24.4H2O in 0.15 mol L-1 HNO3 and 40 µL of 60 mmol L-1 ascorbic acid.

Batch procedure

For accuracy assessment, PI extraction and determination in vegetable materials and leguminous was carried out according to batch extraction procedures.33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207.,55 Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139. In a 50-mL polypropylene flask, 200 mg of plant sample, and 15 mL of water were mixed and kept under agitation for 10 min. Afterwards, samples were centrifuged and filtered for phase isolation. The volume was completed with water up to 20 mL prior to the spectrophotometric determination of PI using molybdenum blue method. For soy flour and beans, 3.0 g of sample was transferred to a 50-mL polypropylene flask and suspended with 30 mL of deionized water. After 1 h of orbital agitation, the suspension was centrifuged, filtered, and the extract volume was made up to 50 mL. The determinations were carried out batchwise using the same reaction conditions as the monosegmented flow procedure,33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. by mixing the extract, a 2.5 mmol L-1 heptamolybdate in 0.25 mol L-1 H2SO4 solution and 0.15 mol L-1 ascorbic acid solution (1.0 mL of each). The mixture rested for 10 min, and the volume was rounded up to 5.0 mL with water. Spectrophotometric responses were obtained at 730 nm.

Results and Discussion

The reaction stoichiometry of molybdenum blue (MB) is not fixed because different molybdate species are formed depending on chemical conditions, especially pH.1818 Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60. Reaction kinetics is also easily altered, depending on the reducing reagent, as well as the number of generated MB species. Nevertheless, a simplified and general reaction can be represented as in equations 1 and 2 1818 Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60. with ascorbic acid as reducing agent.

(1) PO 4 3 + 12 MoO 4 2 + 27 H + H 3 PO 4 MoO 3 12 + 12 H 2 O

(2) H 3 PMo VI 12 O 40 + 2 C 6 H 8 O 6 H 4 PMo VI 8 Mo V 4 O 40 + 2 C 6 H 6 O 6 + 3 H +

Maximum absorption and absorptivity change with the composition of reagents because of different reaction products, such as the ones containing Sb.1919 Murphy, J.; Riley, J. P.; Anal. Chim. Acta 1962, 27, 31. In this work, Sb was neglected, as PI concentration in plant extracts does not require high sensitivity, also yielding a greener approach.

In order to yield pH 0-11818 Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60. for MB formation, sulfuric acid is widely employed despite the evidences that blank signals are higher in the presence of sulfate ions due to formation of species that absorb at 750-770 nm in the absence of phosphorus.1818 Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60. Therefore, analytical signals were obtained in sulfuric and nitric acids media at the same pH. The difference between responses was < 2%, indicating that both acids can be employed. A slight increase of the blank was observed with sulfuric acid solutions, but a paired t-test showed that they did not differ significantly from blank signals obtained with nitric acid. Reagent R1 was then prepared with 0.15 mol L-1 HNO3. Schlieren effect due to acidity of the sample zone was minimized with an intermittent flow of water, which also improved determination rate.

Flow-based determination of PI using zone penetration approach was exploited for the first time. As the volumes of solutions were defined by sampling loops, continuous reagent pumping was avoided, minimizing waste generation and reagents consumption. Merging zones approach was also considered, but the response decreased 42% due to the sample and reagent dilution at the confluence joint. In this condition, the order of the injected solutions did not alter the response significantly.

System optimization for PI determination

The first optimized parameter was sample volume, which was varied from 25 to 300 µL (5.0-60 cm sample loop). The signals increased with sample volume until reaching a plateau, as expected. In spite of minimizing dispersion, the increase of the sample volume hinders the overlap of the zones, causing minimal signal variations above 200 µL due to poorer mixing. Thus, this volume was selected for further studies. Maximum blank absorbance was 0.010 ± 0.001.

Volumes of reagents R1 and R2 were varied from 15 to 125 µL. Lower volumes were not studied because the injector geometry did not allow the use of loops smaller than 3.0 cm. For R1, higher sensitivity was observed with 25 µL. Lower responses were observed for volumes higher than 50 µL because a longer R1 zone increased the distance between sample and R2. The minimum possible volume of R2 was chosen because higher volumes did not affect the analytical signal, as R2 is at the front of the zones, and in great excess. Thus, 25 and 15 µL of R1 and R2 solutions were respectively chosen.

Residence time was evaluated with the variation of reaction coil B3 from 250 to 1000 µL (50-200 cm). Significant increase of the analytical signal with B3 was observed up to 750 µL (150 cm) because from this point reaction development did not compensate signal loss due to dispersion. Still regarding residence time, flow rate was also studied from 0.3 to 3.5 mL min-1. As expected, an increase of the analytical signal was observed for lower flow rates. By employing 1.0 and 2.0 mL min-1, the response decreased ca. 11% with the latter, but determination rate increased 2-fold. Therefore, for PI determination, a 750 µL reaction coil, and 2.0 mL min-1 flow rate were chosen for further experiments.

Intermittent flow was also exploited aiming at decreasing the cleaning time. By inserting water into the analytical path after maximum signal was obtained, the determination rate was increased in twofold due to the higher flow rate after confluence joint x. A 0.10 mol L-1 NaOH solution was also evaluated for cleaning, but determination rate was not significantly affected.

Phosphomolybdic acid formation and reducing reactions strongly depend on acidity and molybdate concentration, making important the optimization of reagents conditions. Nitric acid, ammonium heptamolybdate, and ascorbic acid concentrations were evaluated from 100 to 700, 8.0 to 20 and 40 to 100 mmol L-1, respectively. Similar response trends were observed in these studies. Analytical signal increased up to a certain concentration, and reached a plateau for higher values. The selected parameters were 300 mmol L-1 HNO3, 12 mmol L-1 ammonium heptamolybdate, and 60 mmol L-1 ascorbic acid. Due to the undefined reaction stoichiometry, it is not a simple task to predict which species of MB were formed. Considering the chemical conditions of the sample zone, the precursor of MB was octamolybdate.1818 Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60.,2020 Cruywagen, J. J.; Draaijer, A. G.; Heyns, J. B. B.; Rohwer, E. A.; Inorg. Chim. Acta 2002, 331, 322.

Analytical features

After optimization, the procedure was evaluated for PI determination in plants, soy flour, and beans. A summary of the chosen parameters and the studied ranges is shown in Table 1. Linear response was observed between 2.0 and 30 mg L-1 PI (64-960 µmol L-1), described by equation A = 0.049 CP + 0.067, r = 0.9998, where A is absorbance and CP is PI concentration in mg L-1. Coefficient of variation (n = 20), detection limit (n = 20, 99.7% confidence level), and determination rate were estimated at 1.2%, 0.50 mg L-1 and 40 h-1, respectively. 3.0 mL of waste was generated per determination, consuming 55 µg of (NH4)6Mo7O24.4H2O, and 120 µg of ascorbic acid.

Table 1
Optimized parameters of the flow system (Figure 1b) for determination of soluble inorganic phosphorus

Extraction efficiency evaluation

For dynamic SLE of PI from vegetable materials, Crotalaria juncea was employed as a model, and water was used as extractant as previously recommended.33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. Reference PI content was obtained by a batch extraction,33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. and it was estimated at 1.66 mg g-1. This value was considered as quantitative extraction of PI and was used for evaluating extraction efficiency (EE). Initially, injectors Inj1 and Inj2 in Figure 1 were not connected. After flow-based extraction with 2.3 mL of water, the extracts were collected in 15-mL graduated polyethylene flasks, and diluted 2-fold prior to PI determination with the manifold depicted in Figure 1b. The initial conditions for optimization were 40 mg of sample in an 8.0-cm Tygon® tube (4.0 mm i.d.), extractant flow rate and volume at 2.5 mL min-1 and 3.5 mL, respectively. The recirculation time for extraction was initially carried out for 60 s, which was counted from the moment that injector Inj1 was commuted in the direction of the white arrow in Figure 1a until it was returned to the initial position.

The flow rate of water during recirculation for extraction was evaluated (Figure 2a). Flow rates at 1.5 and 2.5 mL min-1 led to a difference of EE < 2.0%. On the other hand, higher flow rates decreased EE because the time of contact between sample and extractant was increased as the extraction is carried out in a loop. This led to a pronounced water absorption by the solid, retaining part of the extract. This means that the water containing previously extracted analyte was absorbed by the solid, hindering EE. By performing three consecutive extractions with water for 60 s each, and collecting the extract in a vial prior to PI determination, an EE of ca. 97% was achieved. Washing steps and sequential extractions of the same sample would be difficult to implement in the proposed flow system due to the need for entrapping several aliquots of extracts sequentially because of the volumes of coil B2 and sample loop L1 (Figure 1). As flow injection systems show highly reproducible timing, a fixed EE was achieved using flow rate at 2.5 mL min-1 with high reproducibility and determination rate.

Figure 2
Variation of online extraction efficiency with (a) extractant flow rate; (b) sample mass; (c) extractant volume; and (d) recirculation time.

Sample mass was evaluated by altering the volumes of sample chamber and extraction coil B1 proportionally. In this way, extractant volume remained constant during the experiments. In Figure 2b, EE was inversely proportional to sample mass, as expected. With fixed extractant volume and extraction time, higher amounts of sample decreased the contact time of the extractant with the solid. By using 40 mg, in one second, ca. 0.96 µL of water was in contact with 1 mg of sample. This volume decreases 2-fold for 80 mg, implying that poorer efficiency was due to smaller amounts of PI extracted from the sample. Additionally, extract absorption increased with the amount of solid. Therefore, sample mass was reduced to 20 mg (in a 4.0 cm tube). Lower masses were not evaluated to avoid errors related to sample weighing.

Extractant volume and extraction time were evaluated from 1.8 to 2.5 mL (Figure 2c), and from 60 to 300 s (Figure 2d), respectively. The former was studied by altering the volume of extraction coil B1 (Figure 1a). For the latter, the time for recirculation was varied. In both studies, the increased contact time between sample and water (lower volumes of water and higher extraction times) led to poorer EE. This was also observed when flow rate was evaluated, as previously discussed. Water absorption by the sample was noted because of the slight decrease of extract volume (< 100 µL) after collection in the vials. Therefore, 2.5 mL of extractant and 60 s recirculation time were selected. Shorter extraction times were not evaluated due to the time needed for a complete turn of the extract in the extraction path.

After optimization, EE of Crotalaria juncea was estimated at 65 ± 3% (n = 5). Aiming at analytical application to other plant materials, the comparison between batch and dynamic extractions were carried out for leaves and stem of coffee and fig trees (n = 5), leading to extraction efficiencies of 65 ± 3 and 63 ± 4%, respectively. These samples originated from Coffea sp. and Ficus carica. For soy flour and beans, EE reached 69 ± 4%. This value was not statistically different from EE of plants (95% confidence level). In-line extraction led to coefficients of variation < 5.0%, which is lower than the observed for batch extractions (< 15%), demonstrating better reproducibility of the proposed procedure.

The direct transportation of the extract from Inj1 (Figure 1a) towards sample loop L1 of Inj2 (Figure 1b) was evaluated to avoid the collection of the extract in vessels prior to determination of PI. As a twofold dilution was required for analysis, the concentration gradient of the extract was exploited to yield this condition for PI determination. Therefore, a 50-cm tube (0.76 mm i.d.) was employed to connect the extract outlet of Inj1 to sample loop inlet of Inj2. The time was controlled by using the peristaltic pump timer. Signals obtained for a sample of Crotalaria juncea after in-line extraction followed by 2-fold dilution in vials, and direct determination after 70 s transportation were 0.325 ± 0.005 and 0.337 ± 0.006, respectively. The signals did not show significant difference at 95% confidence level (n = 3), indicating that a twofold dilution was achieved with both conditions. Depending on the required dilution, the time for transportation can be varied with coil length, and flow rate. Considering the time for in-line extraction and determination of PI, sample throughput was estimated at 20 h-1. As the sample chamber is cheap and easily constructed (comprising a simple polymeric tube), several samples can be prepared, and the dynamic extractions can be carried out sequentially by changing the sample. After extraction, the sample chambers can be disposed or reused after washing with diluted nitric acid, and water.

Application and critical evaluation

Results of the analysis of plant and food extracts by the proposed procedure were compared with those obtained by batch procedures33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207.,55 Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139. described in “Batch procedure” sub-section. Table 2 shows the comparison of results for samples analyzed with the manifold in Figure 1b using batch (Crotalaria 3, soy flours, and beans) and dynamic (Crotalaria 1-3, coffee and fig trees, and soy flours) extractions. In this way, the accuracy of PI determination was assessed after batch and in-line extractions. The results agreed with the attained by the batch procedures (95% confidence level), considering the EE of 65%.

Table 2
Main values and uncertainties for soluble inorganic phosphorus determination in plant materials, soy flours, and beans by the proposed and batch procedures

The proposed procedure is faster and cleaner than other strategies for extraction and determination of PI. In comparison to a reported procedure,33 Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207. this work showed 10-fold higher detection limit due to the shorter residence time. This feature is not critical for PI determination in vegetable materials. On the other hand, the consumption of molybdate and ascorbic acid were 15 and 40-fold lower, respectively. Extraction time was reduced 10-fold using significative lower amounts of sample and extractant. Reagent consumption was also decreased compared to a proposed procedure for determination of P species in foodstuffs using multicommutation.55 Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139. In spite of the advantages of solenoid micro pumps for solution handling and automation, reagent consumption was 7-fold higher than the proposed procedure with similar detection limit (0.5 mg L-1 PI). In comparison to a reference procedure for PI extraction in plant materials,66 Hogue, E.; Wilcox, G. E.; Cantliffe, D. J.; J. Am. Soc. Hortic. Sci. 1970, 95, 174. this work did not employ neither mineral acids for extraction nor active charcoal for clean-up steps. In all previous works, phosphorus extractions were carried out batchwise, affecting analytical frequency and reproducibility.

Recent works1010 Sweileh, J. A.; Microchem. J. 2000, 65, 87.,1111 Sweileh, J. A.; Anal. Chim. Acta 2007, 581, 168.,1313 Galvis-Sánchez, A. C.; Santos, J. R.; Rangel, A. O. S. S.; Anal. Bioanal. Chem. 2016, 408, 7651. have proposed SLE in flow-based systems with elaborated chambers that might perform agitation and filtering. In this work, stirring and washing steps were not necessary during the dynamic process because incomplete extractions were exploited reproducibly. Additionally, slurries were not formed in the extraction path because the solid was trapped by glass wools at both tips of the chamber. Online SLE approaches have also been proposed using simple units for extraction aiming at bioavailability studies,1515 Maia, M. A.; Soares, T. R. P.; Mota, A. I. P.; Rosende, M.; Magalhães, L. M.; Miró, M.; Segundo, M. A.; Talanta 2017, 166, 162.,1717 Rosende, M.; Miró, M.; Cerdà, V.; Anal. Chim. Acta 2010, 658, 41. but direct determination after dynamic extractions have not been exploited.

The comparison of some important features of proposed procedures is presented in Table 3. In this work, sample mass was similar to or smaller than other procedures with dynamic SLE (from 0.010 to 2.0 g). As determination of PI is carried out after incomplete extraction, a shorter time for extracting the analyte (60 s) was achieved compared to most procedures. This comparison should not be extended to dynamic extractions for bioavailability studies because they require longer periods due to multiple extractions. Another feature is that washing step of the sample is not required as normally is necessary in quantitative processes.1313 Galvis-Sánchez, A. C.; Santos, J. R.; Rangel, A. O. S. S.; Anal. Bioanal. Chem. 2016, 408, 7651.

Table 3
Analytical features of some procedures with in-line and online solid-liquid extraction

Direct analyte determination after extraction is also an inherent feature of the proposed procedure. Most strategies using dynamic SLE require sample collection in vials,1515 Maia, M. A.; Soares, T. R. P.; Mota, A. I. P.; Rosende, M.; Magalhães, L. M.; Miró, M.; Segundo, M. A.; Talanta 2017, 166, 162.,1616 Chomchoei, R.; Miró, M.; Hansen, E. H.; Shiowatana, J.; Anal. Chim. Acta 2005, 536, 183. and separation units1010 Sweileh, J. A.; Microchem. J. 2000, 65, 87.

11 Sweileh, J. A.; Anal. Chim. Acta 2007, 581, 168.
-1212 Almeida, M. I. G. S.; Segundo, M. A.; Lima, J. L. F. C.; Rangel, A. O. S. S.; Microchem. J. 2006, 83, 75. (e.g. filtration) prior to determinations. Additionally, the whole system is mechanized, avoiding manual solution handling as previously proposed.11 Schachtman, D.; Reid, R.; Ayling, S. M.; Plant Physiol. 1998, 116, 447.

Conclusions

A flow-based procedure with in-line SLE for PI was developed and successfully applied to the accurate analysis of vegetable materials and solid foodstuffs. Reagent consumption was low, especially due to the zone penetration approach for PI determination, and the employment of water as extractant. The analytical features demonstrated the potential of the system for PI determination. In addition, sample and extractant consumptions, extraction time, and waste generation were minimized in comparison to other proposed procedures.

Incomplete extraction was applied for analytical ends due to the reproducible timing of the flow system, allowing faster determination of PI. A simple chamber for sample entrapment was used, encompassing a polymeric tube with two small pieces of glass wool. In this way, the procedure does not require neither expensive nor laborious devices for extraction.

The procedure is then a reliable alternative for phosphorus determination, bringing advantages of incomplete in-line sample preparation. Agronomic studies can be benefited by in-line SLE, as well as laboratories for routine analysis of solid foodstuffs.

Acknowledgments

The authors acknowledge the fellowships and financial support from Brazilian agencies Coordination for the Improvement of Higher Education Personnel (CAPES) and São Paulo Research Foundation (FAPESP), process No. 2015/12172-6.

References

  • 1
    Schachtman, D.; Reid, R.; Ayling, S. M.; Plant Physiol. 1998, 116, 447.
  • 2
    Reddy, N. R. In Food Phytates, 1st ed.; Reddy, N. R.; Sathe, S. K., eds.; Taylor & Francis Group: Boca Raton, USA, 2001, p. 25.
  • 3
    Maruchi, A. K.; Rocha, F. R. P.; Microchem. J. 2006, 82, 207.
  • 4
    Rocha, D. L.; Batista, A. D.; Rocha, F. R. P.; Donati, G. L.; Nóbrega, J. A.; TrAC, Trends Anal. Chem. 2013, 45, 79.
  • 5
    Rocha, D. L.; Rocha, F. R. P.; Microchem. J. 2013, 109, 139.
  • 6
    Hogue, E.; Wilcox, G. E.; Cantliffe, D. J.; J. Am. Soc. Hortic. Sci. 1970, 95, 174.
  • 7
    Noack, S. R.; Smernik, R. J.; McBeath, T. M.; Armstrong, R. D.; McLaughlin, M. J.; Talanta 2014, 126, 122.
  • 8
    Trojanowicz, M.; Kolacinska, K.; Analyst 2016, 141, 2085.
  • 9
    Rocha, D. L.; Kamogawa, M. Y.; Rocha, F. R. P.; Anal. Chim. Acta 2015, 896, 11.
  • 10
    Sweileh, J. A.; Microchem. J. 2000, 65, 87.
  • 11
    Sweileh, J. A.; Anal. Chim. Acta 2007, 581, 168.
  • 12
    Almeida, M. I. G. S.; Segundo, M. A.; Lima, J. L. F. C.; Rangel, A. O. S. S.; Microchem. J. 2006, 83, 75.
  • 13
    Galvis-Sánchez, A. C.; Santos, J. R.; Rangel, A. O. S. S.; Anal. Bioanal. Chem. 2016, 408, 7651.
  • 14
    Timofeeva, I. I.; Bulatov, A. V.; Moskvin, A. L.; Kolev, S. D.; Talanta 2015, 142, 140.
  • 15
    Maia, M. A.; Soares, T. R. P.; Mota, A. I. P.; Rosende, M.; Magalhães, L. M.; Miró, M.; Segundo, M. A.; Talanta 2017, 166, 162.
  • 16
    Chomchoei, R.; Miró, M.; Hansen, E. H.; Shiowatana, J.; Anal. Chim. Acta 2005, 536, 183.
  • 17
    Rosende, M.; Miró, M.; Cerdà, V.; Anal. Chim. Acta 2010, 658, 41.
  • 18
    Nagul, E. A.; McKelvie, I. D.; Worsfold, P.; Kolev, S. D.; Anal. Chim. Acta 2015, 890, 60.
  • 19
    Murphy, J.; Riley, J. P.; Anal. Chim. Acta 1962, 27, 31.
  • 20
    Cruywagen, J. J.; Draaijer, A. G.; Heyns, J. B. B.; Rohwer, E. A.; Inorg. Chim. Acta 2002, 331, 322.