Strategies to improve accuracy and sensitivity in phosphorus determinations by inductively coupled plasma quadrupole mass spectrometry

Abstracts

The collision/reaction interface with He introduced through the skimmer at 100 mL min-1 is employed to improve accuracy in 31P+ determinations by ICP-QMS. The 31P16O+ determination in cool plasma conditions result in limits of detection 55-fold lower. This strategy may be an interesting alternative for high ionization potential elements such as P.

collision/reaction interface; dissociation energy; H2 reactivity; 31P16O+; thermodynamics


A interface de colisão e reação com He introduzido pelo skimmer a 100 mL min-1 é empregada para melhorar a exatidão na determinação de 31P+ por ICP-QMS. A determinação de 31P16O+ em plasma frio resulta em limites de detecção 55 vezes menores. Essa é uma alternativa interessante para elementos com altos potenciais de ionização como o P.


SHORT REPORT

Strategies to improve accuracy and sensitivity in phosphorus determinations by inductively coupled plasma quadrupole mass spectrometry

George L. Donati* * e-mail: georgedonati@yahoo.com.br FAPESP has sponsored the publication of this article. ; Renata S. Amais; Joaquim A. Nóbrega

Departamento de Química, Universidade Federal de São Carlos, CP 676, 13560-970 São Carlos-SP, Brazil

ABSTRACT

The collision/reaction interface with He introduced through the skimmer at 100 mL min-1 is employed to improve accuracy in 31P+ determinations by ICP-QMS. The 31P16O+ determination in cool plasma conditions result in limits of detection 55-fold lower. This strategy may be an interesting alternative for high ionization potential elements such as P.

Keywords: collision/reaction interface, dissociation energy, H2 reactivity, 31P16O+, thermodynamics

RESUMO

A interface de colisão e reação com He introduzido pelo skimmer a 100 mL min-1 é empregada para melhorar a exatidão na determinação de 31P+ por ICP-QMS. A determinação de 31P16O+ em plasma frio resulta em limites de detecção 55 vezes menores. Essa é uma alternativa interessante para elementos com altos potenciais de ionização como o P.

Introduction

Inductively coupled plasma mass spectrometry (ICP-MS) is one of the most powerful techniques in modern analytical chemistry. Due to its high sensitivity and multi-element capabilities, ICP-MS has been used in different fields to solve important analytical problems.1-3 The reduction of costs to acquire and maintain quadrupole-based instruments (ICP-QMS) and the increasing demand for fast, sensitive and robust methods have contributed to expand ICP-MS applications in the last decades. One of the most important limitations of ICP-MS is related to its relatively low resolution.4 Spectral overlap from polyatomic ions formed by interactions among matrix, solvent, plasma and atmospheric species may compromise sensitivity and accuracy, frequently hindering ICP-QMS application to complex samples. Different approaches have been proposed to minimize the problem: from mathematical and chemical strategies to more instrumental ones (or a combination of those), all present potentialities and limitations.5-8

An interesting alternative to prevent ICP-QMS spectral interferences is to determine an analyte based on measurements of its oxide. By manipulating the plasma or the reaction cell conditions, it is possible to maximize the oxide formation and improve sensitivity. Since a different, probably less prone to interferences mass-to-charge ratio (m/z) is monitored, the determination of oxide ions may also minimize spectral interferences. This strategy is especially useful for elements with high ionization potentials, such as P, and some works have demonstrated its effectiveness in different applications.9,10

More recently, Kalinitchenko et al.11 proposed promoting physical-chemical interactions directly in the sampler or the skimmer cones as a simpler alternative to pressurized collision/reaction cells.8 In the collision/reaction interface technology (CRI), He or H2 are introduced into the system through modified sampler and skimmer cones to promote interactions capable of destroying interfering species and improving accuracy in ICP-QMS determinations. This strategy has been successfully used for different analytes12 and it has potential for applications in combination with other spectral interference-reducing methods.13

In this work, the CRI technology and the determination of oxide species are evaluated to improve sensitivity and accuracy in P determinations. This is a challenging element in ICP-QMS because of its relatively high ionization potential (i.e., 10.5 eV)14 and its severe spectral interferences. The main polyatomic ions interfering on P determinations at m/z 31 are 14N16OH+ and 15N16O+.15 Considering that P is monoisotopic and that most sample digestions require HNO3,16 the difficulty of determining P by ICP-QMS becomes evident. Different He and H2 CRI flow rates are evaluated for P determinations at m/z 31 (31P+) and 47 (31P16O+). Some preliminary results are presented and physical-chemical aspects affecting the CRI efficiency in each case are discussed.

Experimental

Instrumentation

An inductively coupled plasma mass spectrometer equipped with a quadrupole mass analyzer and the CRI technology (ICP-QMS, 820-MS, Varian, Mulgrave, Australia) was used in all experiments. This instrument presents modified skimmer and sampler cones which allow the introduction of H2 or He directly into the plasma/mass spectrometer interface. The CRI technology is complemented by a 90° angle design in which plasma source and mass spectrometer are positioned off axis to each other to reduce instrumental noise. Details on the instrument design and the CRI configuration can be found in the literature.11,17

The sample introduction system is composed of an automatic sampler (SPS3, Varian), a double pass, Scott-type spray chamber, and a concentric nebulizer. The mass‑to‑charge ratios (m/z) monitored were 31 and 47 for 31P+ and 31P16O+, respectively. Table 1 presents the ICP-QMS operating conditions used in this work.

A closed-vessel microwave oven (Ethos 1600, Milestone, Sorisole, Italy) equipped with 45 mL PFA vessels was used for sample preparation.

Reagents, reference solutions and samples

Ultrapure HNO3 was produced using a sub-boiling distillation apparatus (Milestone). This reagent was used to prepare all aqueous solutions and to digest the sample aliquots. Hydrogen peroxide 30% (m/m) was also used in the microwave-assisted acid digestion of the sample aliquots. Distilled-deionized water (18.2 MΩ cm, Milli-Q, Millipore, Bedford, MA, USA) was used as solvent in all experiments. Standard reference solutions of P were prepared by appropriated dilution of a 1000 mg L-1 stock solution (Tec-Lab, Hexis, São Paulo, SP, Brazil). Freshly collected tap water was spiked with 1.0 mg L-1 P and used as test solution to optimize the CRI conditions. A standard reference material (SRM 1548a, "Typical Diet") from the National Institute of Standards and Technology (NIST, Gaithersburg, MD, USA) was used to check the method accuracy.

All polypropylene material was kept in 10% (v/v) HNO3 overnight and rinsed with distilled-deionized water before use to prevent contaminations. Liquid Ar and the CRI gases (He and H2) used in this work were 99.999% pure (White Martins, Sertãozinho, SP, Brazil).

Sample preparation

Freshly collected tap water aliquots of 5 mL were spiked with P and diluted to 10 mL with distilled-deionized water to a final concentration of 1.0 mg L-1.

Typical Diet (NIST 1548a) aliquots of approximately 250 mg were accurately weighted directly in microwave PFA vessels and submitted to a pre-digestion period of 60 min with 2.5 mL of concentrated HNO3 (14 mol L-1) without any external heating. Then, 2.5 mL of distilled-deionized water were added to the mixture and an additional pre-digestion period of 30 min at room temperature was observed. Finally, a volume of 3.0 mL of H2O2 30% (m/m) was added and the mixture was digested in a closed-vessel cavity microwave oven (Milestone) using the heating program presented in Table 2. The resulting solutions were then allowed to cool down, transferred to 15 mL polypropylene flasks, and diluted to a final volume of 10 mL with distilled-deionized water. Additional dilution was performed just before analysis to adjust the P concentration to the analytical calibration curve linear dynamic range.

Optimization of the CRI conditions

Different CRI flow rates were evaluated for He or H2 introduced through the skimmer or the sampler cones. For 31P+ determinations, He flow rates of 40, 60, 80 and 100 mL min-1 introduced through the skimmer, and of 500, 1000 and 1500 mL min-1 introduced through the sampler were evaluated. Hydrogen gas introduced through the skimmer at 20, 40 and 60 mL min-1, and through the sampler at 200, 1000 and 1500 mL min-1 was also evaluated. For 31P16O+, determinations without the CRI and with H2 flowing at 40 mL min-1 through the skimmer were compared.

Results and Discussion

Determinations at m/z 31

The CRI efficiency at reducing spectral interferences in 31P+ determinations was evaluated by comparing recoveries obtained for a tap water sample spiked with 1.0 mg L-1 P in different He or H2 flow rates. As observed in previous works,11-13 the sampler cone interface was not efficient and recoveries from 22 to 80% were obtained. For determinations using the skimmer cone interface, recoveries varied from 60 to 94%. Table 3 presents the results for He and H2 in different flow rates.

Based on these results, conditions presenting recoveries higher than 85%, i.e., He at 80 and 100 mL min-1, and H2 at 40 mL min-1, were used to determine P in a standard reference material (Typical Diet, NIST 1548a). The goal in this case was to check the method efficiency for a more complex matrix. No statistically significant difference at a 95% confidence level was observed between the certified value (3486 ± 245 mg kg-1) and the ones determined using He at 80 or 100 mL min-1 in the skimmer cone (3960 ± 270 and 3350 ± 250 mg kg-1, respectively). On the other hand, a lower accuracy was observed for determinations using H2 at 40 mL min-1 in the skimmer, i.e., 3010 ± 60 mg kg-1.

Kinetic and thermodynamic aspects must be considered while applying the CRI to minimize ICP-QMS spectral interferences. Helium is essentially used as a collision gas that may destroy interfering ions by collision induced dissociation (CID). However, it may only be efficient if the collision energy is higher than the interfering ion dissociation energy (Do).18 As previously demonstrated by Kovacevic et al.,15 the main interfering ions affecting 31P+ determinations in HNO3 medium are 14N16OH+ (65% abundance) and 15N16O+ (32% abundance). By calculating the collision energy in the center of mass between He and these ions (Ecm He/31 = 0.46 eV ),19 and comparing it with the dissociation energies (Do = 0.395 and 10.85 eV for 14N16OH+ and 15N16O+, respectively),14,19 one can assume that the CRI on collision mode could only be effective at minimizing spectral overlap from 14N16OH+. Considering the ICP-MS conditions and 15N low natural abundance (i.e., 0.37%), it may be supposed that 14N16OH+ is the species responsible for most significant spectral interferences, which could explain the adequate accuracies obtained for determinations using He in the skimmer cone.

It is important to note that the accuracy obtained in this work may also be related to more efficient sample matrix decompositions. It has been demonstrated that high residual carbon contents (RCC) can compromise accuracy in 31P+ determinations, probably due to a higher concentration of carbon-based interfering ions such as 12C18OH+.15 In a previous work using the same standard reference material and the CRI on collision mode, poor accuracies were observed probably because no pre-digestion step was employed and digested sample solutions with higher RCCs were probably obtained.7 Considering 12C18OH+ as the main carbon-based interfering ion in high RCC solutions,15 and comparing Ecm He/31 and Do values (0.46 and 6.14 eV, respectively),14,19 the reason for such poor accuracies7 becomes evident.

On the other hand, no kinetic data are available for reactions between H2 and either of these interfering ions.20 In addition, reactions represented by equations 1-314,21 indicate that they are thermodynamically non-favorable since they are all endothermic. Therefore, as it would be expected and it is confirmed by the Typical Diet results, the CRI on reaction mode is not effective at improving accuracy in 31P+ determinations.

Determinations at m/z 47

As previously demonstrated in different works,9,10 the determination of P as PO+ at m/z 47 may be an interesting alternative to improve sensitivity and minimize interferences in ICP-QMS analyses. The PO+ ion is readily formed in the plasma as demonstrated in equation 4 (k = 8.70 × 10-10 cm3 s-1).22 Since this is an exothermic reaction, its formation may be maximized by reducing the plasma temperature. This effect can be obtained by reducing the radio-frequency (rf) applied power and increasing the nebulizer gas flow rate (Table 1). It is interesting to note that a higher nebulizer gas flow rate can have a two-fold positive effect on maximizing the PO+ formation since it may also increase the number of non-ionized water molecules available.

The efficiency of a method combining the CRI strategy and PO+ determination in cooler plasma conditions was evaluated by analyzing recoveries in a tap water sample spiked with 1.0 mg L-1 P. Two aspects need to be considered while using the CRI for oxide ion determinations: the analyte should present both high dissociation energy and limited reactivity with H2 so it is not destroyed in the interface. Table 4 presents the dissociation energies for PO+ and its main interfering ions (i.e., 14N16O2H+, 15N16O2+, 12C35Cl+, 13C18O16O+, and 12C18O16OH+),15 as calculated using thermodynamic data from the literature.14 By comparing these values with the collision energy in the center of mass between He and those species (Ecm He/47 = 0.31 eV),19 it can be assumed that neither the analyte will be dissociated nor the CRI on collision mode will be effective at minimizing the main spectral interferences at m/z 47. The same reasoning may be applied for determinations using the CRI on reaction mode and by analyzing the energies involved in reactions represented by equations 5-10.14,21,23 Based on these data and considering the sampler interface low efficiency,11-13 only two CRI conditions were evaluated for 31P16O+ determinations: H2 at 0 or 40 mL min-1 introduced through the skimmer. The results for spiked tap water and Typical Diet are presented in Table 5. As expected, no accuracy improvements were observed by employing the CRI in each case.

Limits of detection and quantification

The instrumental limits of detection (LOD) and quantification (LOQ) for different CRI conditions were calculated and compared with determinations without the CRI. The results are presented in Table 6. As it can be seen, significant improvements were obtained for determinations at m/z 47, probably due to lower background signals in a less energetic plasma. The LODs were calculated as 3 times the background equivalent concentration (BEC) multiplied by the relative standard deviation of twenty consecutive blank determinations (RSD, n = 20). The BEC was calculated by dividing the analyte concentration in one of the calibration curve points by its respective signal-to-background ratio (SBR). SBR is the net analytical signal divided by the blank signal.24 The LOQ as calculated as 10 times BEC multiplied by RSD.

Conclusions

The collision/reaction interface can be effective at improving accuracy in ICP-QMS determinations. However, kinetic and thermodynamic aspects need to be considered while developing methods based on this technology. The CRI on collision mode may be effective for determinations involving interfering species with both low mass and low dissociation energies. Considering accuracy and limits of detection in 31P+ determinations, the introduction of He at 100 mL min-1 through the skimmer is the most recommended condition.

Monitoring PO+ in cool plasma conditions is an interesting alternative to improve ICP-QMS sensitivity. Comparing this strategy with the optimized CRI method for 31P+, a 55-fold improvement in the LOD was obtained. On the other hand, a larger number of interfering ions with relatively high ionization energies and low reactivity with H2 can compromise both the CRI efficiency and the method accuracy for determinations at the m/z 47.

Acknowledgments

The authors would like to thank the Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP) for the grants and fellowships provided (2006/59083-9, 2010/50238-5, and 2010/17387-7). The support from the Instituto Nacional de Ciências e Tecnologias Analíticas Avançadas and Conselho Nacional de Desenvolvimento Científico e Tecnológico (INCTAA and CNPq) is also greatly appreciated.

Submitted: November 22, 2011

Published online: February 2, 2012

  • 1. Avula, B.; Wang, Y.; Duzgoren-Aydin, N. S.; Khan, I. K.; Food Chem. 2011, 127, 54.
  • 2. Sussulini, A.; Banzato, C. E. M.; Arruda, M. A. Z.; Int. J. Mass Spectrom. 2011, 307, 182.
  • 3. Oliveira, E. P.; Yang, L.; Sturgeon, R. E.; Santelli, R. E.; Bezerra, M. A.; Willie, S. N.; Capilla, R.; J. Anal. At. Spectrom. 2011, 26, 578.
  • 4. Becker, J. S.; Inorganic Mass Spectrometry: Principles and Applications; Wiley: Chichester, 2008.
  • 5. Thomas, R.; Spectroscopy 2002, 17, 24.
  • 6. Tanner, S. D.; J. Anal. At. Spectrom. 1995, 10, 905.
  • 7. Donati, G. L.; Amais, R. S.; Nóbrega, J. A.; J. Anal. At. Spectrom. 2011, 26, 1827.
  • 8. D'Ilio, S.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6.
  • 9. Yang, C.; Jiang, S.; Spectrochim. Acta, Part B 2004, 59, 1389.
  • 10. Ciavardelli, D.; Sacchetta, P.; Federici, G.; Di Ilio, C.; Urbani, A.; Talanta 2010, 80, 1513.
  • 11. Kalinitchenko, I.; Wang, X.; Sturman, B.; Spectroscopy 2008, 23, 38.
  • 12. Pereira, C. D.; Garcia, E. E.; Silva, F. V.; Nogueira, A. R. A.; Nóbrega, J. A.; J. Anal. At. Spectrom. 2010, 25, 1763.
  • 13. Fialho, L. L.; Pereira, C. D.; Nóbrega, J. A.; Spectrochim. Acta, Part B 2011, 66, 389.
  • 14
    National Institute of Standards and Technology, Chemistry WebBook, NIST Standard Reference Database. http://webbook.nist.gov/chemistry/form-ser.html accessed in October 2011.
  • 15. Kovacevic, M.; Goessler, W.; Mikac, N.; Veber, M.; Anal. Bioanal. Chem. 2005, 383, 145.
  • 16. Nóbrega, J. A.; Donati, G. L. In Encyclopedia of Analytical Chemistry; Meyers, R. A.; Bings, N. H., eds.; Wiley: Chichester, 2011.
  • 17. Elliot, S.; Knowles, M.; Kalinitchenko, I.; Spectroscopy 2004, 19, 30.
  • 18. Bandura, D. R.; Baranov, V. I.; Tanner, S. D.; Fresenius J. Anal. Chem. 2001, 370, 454.
  • 19. Houk, R. S.; Praphairaksit, N.; Spectrochim. Acta, Part B 2001, 56, 1069.
  • 20. Anicich, V. G.; Astrophys. J. Suppl. S. 1993, 84, 215.
  • 21. Oka, T.; Proc. Natl. Acad. Sci. U.S.A. 2006, 103, 12235.
  • 22. Thorne, L. R.; Anicich, V. G.; Huntress, Jr., W. T.; Chem. Phys. Lett.1983, 98, 162.
  • 23. Green, S.; Schor, H.; Siegbahn, P.; Thaddeus, P.; Chem. Phys. 1976, 17,479.
  • 24. Thomsen, V.; Roberts, G.; Burguess, K.; Spectroscopy 2000, 15, 33.
  • *
    e-mail:
    FAPESP has sponsored the publication of this article.