Analysis of the Mechanical Behavior of Porcine Graft Fixation in a Polyurethane Block Using a 3D-printed PLA Interference Screw* * Trabalho desenvolvido na UFERSA - Universidade Federal Rural do Semi-Árido, Mossoró, RN, Brasil.

Abstract

Objective

The interest in using 3D printing in the healthcare field has grown over the years, given its advantages and potential in the rapid manufacturing of personalized devices and implants with complex geometries. Thus, the aim of the present study was to compare the mechanical fixation behavior of a 3D-printed interference screw, produced by fused deposition modeling of polylactic acid (PLA) filament, with that of a titanium interference screw.

Methods

Eight deep flexor porcine tendons, approximately 8 mm wide and 9 cm long, were used as graft and fixed to a 40 pounds-per-cubic-foot (PCF) polyurethane block at each of its extremities. One group was fixed only with titanium interference screws (group 1) and the other only with 3D-printed PLA screws (BR 20 2021 018283-6 U2) (group 2). The tests were conducted using an EMIC DL 10000 electromechanical universal testing machine in axial traction mode.

Results

Group 1 (titanium) obtained peak force of 200 ± 7 N, with mean graft deformation of 8 ± 2 mm, and group 2 (PLA) obtained peak force of 300 ± 30 N, and mean graft deformation of 7 ± 3 mm. Both the titanium and PLA screws provided good graft fixation in the polyurethane block, with no slippage or apparent deformation. In all the samples, the test culminated in graft rupture, with around 20 mm of deformation in relation to the initial length.

Conclusion

The 3D-printed PLA screw provided good fixation, similar to that of its titanium counterpart, producing satisfactory and promising results.

Keywords
bone screws; lactic acid; printing, three-dimensional