Repeatability of the tomographic study of the cornea with a rotational Scheimpflug in normal eyes, eyes with keratoconus and eyes with crosslinking for keratoconus

ABSTRACT

Objective

To evaluate the accuracy of corneal tomography with Scheimpflug images (Pentacam® AXL, OCULUS, Wetzlar, Germany) in scanning patterns with 25 and 50 images, by analyzing their repeatability and reproducibility in normal eyes, eyes with keratoconus and with keratoconus after crosslinking.

Methods

Prospective, cross-sectional, mixed study including patients who had normal corneas, or who had keratoconus, who had not undergone corneal surgery, with the exception of crosslinking. Patients who had undergone surgery for intracorneal ring implantation or corneal transplantation were excluded. Three measurements were performed with 25 images and three measurements were performed with 50 images on each eye included in the study.

Results

The study evaluated a total of 54 eyes of 41 patients. Of these, 10 (18.52%) eyes had normal corneas, without previous eye surgery, 22 (40.74%) were eyes with keratoconus, without previous eye surgery, and 22 (40.74%) were eyes with keratoconus, with crosslinking for keratoconus. In the study of repeatability, we noticed a greater variation of values in all indices in patients with KC compared to those with normal eyes, in most of the parameters analyzed.

Conclusion

There is no significant difference between the scanning methods with 25 and 50 images. However, 25 scans has advantages that make it the main method suggested. The presence of keratoconus has a negative impact on the repeatability of both types of examination, and there is great agreement between the methods in the three groups studied. The high repeatability coefficient in some parameters suggests that these data should be used with care in the assessment of keratoconus progression.

Repeatability; Keratoconus; Multimodal propaedeutics; Corneal crosslinking; Corneal tomography with rotational Scheimpflug