Acessibilidade / Reportar erro

Percutaneous closure of ostium secundum atrial septal defect using the memopart™ occlusion device

BACKGROUND: The authors report their initial experience with the Memopart™ device (Shanghai Shape Memory Alloy Co Ltd, Shanghai, China) for the occlusion of secundum atrial septal defect (ASD). METHODS: This was a prospective observational study of a series of patients undergoing percutaneous occlusion of ASD with right ventricle volume overload and favorable anatomic characteristics. The procedure was performed by percutaneous femoral approach. The mean follow-up was 10.3 ± 5 months, with clinical and 24-hour echocardiographic evaluations (or before hospital discharge), 1, 3, 6, and 12 months after implantation. RESULTS: From February/2012 to April/2013, 21 patients, 16 females, mean age 33.1 ± 18.7 years, were submitted to percutaneous occlusion of an ASD. The average diameter of the defect was 19.04 ± 6.25 mm and the device size was 21.42 ± 6.73 mm (8 to 34 mm). Total occlusion of the defect was observed in all cases before hospital discharge. During follow-up, all patients were asymptomatic and without residual shunt. There was no deaths or any other complications in the series. CONCLUSIONS: The percutaneous closure of ASD using a Memopart™ device is an effective and safe procedure within the limits of this investigation. The device is user-friendly and has a high rate of immediate occlusion, even in large defects.

Heart septal defects, atrial; Septal occluder device; Heart defects, congenital; Cardiac catheterization


Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista - SBHCI R. Beira Rio, 45, 7o andar - Cj 71, 04548-050 São Paulo – SP, Tel. (55 11) 3849-5034, Fax (55 11) 4081-8727 - São Paulo - SP - Brazil
E-mail: sbhci@sbhci.org.br