Acessibilidade / Reportar erro

Preanalytical and validation studies for the determination of malondialdehyde in human plasma through high performance liquid chromatography after derivatization with 2,4-dinitrophenylhydrazine

Malondialdehyde (MDA) is an important biomarker for the evaluation of oxidative stress status. The aim of this study was to develop a method for plasma MDA quantification by high performance liquid chromatography with diode-array detection (HPLC-DAD), following derivatization with 2,4-dinitrophenylhydrazine (DNPH), evaluating the main preanalytical variables. The calibration curve in plasma (0 to 40 µM) presented high linearity (r² = 0.998). Main validation parameters were: recovery: 78%; LOD: 0.11 µM and LOQ: 0.38 µM. The MDA values obtained in healthy volunteers (n=38) were 3.31 ± 0.38 µM (mean ± SD). Stability studies of MDA standard solution and derivatizing reagent and plasma MDA indicated that the standard solution can be stored at -20 and 4 ºC, remaining stable for at least 30 days.

Oxidative stress; Malondialdehyde; High performance liquid chromatography; 2,4-dinitrophenylhydrazine


Divisão de Biblioteca e Documentação do Conjunto das Químicas da Universidade de São Paulo Av. Prof. Lineu Prestes, 950 , Caixa Postal 66083 , 05315-970 - São Paulo - SP - Brasil , Tel.: (011) 3091.3804 , Fax: (011) 3097.8627 - São Paulo - SP - Brazil