Enzyme immobilization using a commercial kit: determination of metabolic parameters in animal blood employing a multicommutation flow system

Automatic flow procedures based on the multicommutation concept, dedicated to the determination of 3-hydroxybutyrate, glucose and cholesterol are proposed. The enzymes were immobilized on glass beads and packed into mini-columns that were coupled to a flow system. Sampling throughputs of 55, 40 and 40 determinations per hour, linear response from 10 to 150, 50 to 600, 25 to 125 mg L-1, detection limits of 1.5, 14 and 4 mg L-1 and relative standard deviations of 1, 2 and 2% for 3-hydroxybutyrate, glucose and cholesterol, respectively, were achieved.

enzymatic reactions; flow injection analysis; commercial kit