Electrochemical hydride generation for selenium determination in a flow injection system with Air-GLP flame atomic absorption spectrometric detection

This paper presents a system for electrochemical hydride generation using flow-injection and atomic absorption spectrometry to determine selenium in biological materials. The electrolytic cell was constructed by assembling two reservoirs, one for the sample and the other for the electrolytic solution separated by a Nafion membrane. Each compartment had a Pt electrode. The sample and electrolyte flow-rates, acidic media, and applied current were adjusted to attain the best analytical performance and ensure the membrane lifetime. The atomisation system used a T quartz tube in an air-LPG flame. The composition of the flame, the observation height, and the argon flow rate used to carry the hydrides were critically investigated. The system allowed to perform thirty determinations per hour with a detection limit of 10 mug L-1 of Se. Relative standard deviations were in general lower than 1.5% for a solution containing 20.0 and 34.0 mug L-1 of Se in a typical sample digest. Accuracy was assessed analysing the certified materials: rice flour (NIST-1568) from National Institute of Standard and Technology and dried fish (MA-A-2), whole animal blood (A-2/1974) from the International Atomic Energy Agency.

electrochemical hydride generation; selenium in biological material; air-LPG flame