Accessibilidad / Informe de Error

Determination of valproic acid in serum by high performance liquid chromatography with diode array detection (HPLC-DAD), after derivatization with phenacyl bromide

Valproic acid (VA) is a drug used to control seizures in several epileptic conditions. In VA pharmacotherapy, therapeutic drug monitoring is recommended to obtain adequate seizure control and avoid toxicity. The aim of this study was to validate a method for the determination of valproic acid in serum, employing high performance liquid chromatography with diode array detection (HPLC-DAD), after derivatization with phenacyl bromide. The calibration curve (y=0.0133x-0.0025) presented good linearity with r²=0.9999. Accuracy (101-115%), intra-assay precision (4.53-8.15%) and inter-assay precision (3.15-6.77%) were acceptable. The quantification limit was 2.0 µg/mL. The method presented similar results to enzyme immunoassay.

valproic acid; phenacyl bromide; HPLC-DAD


Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil