Acessibilidade / Reportar erro

Rhodium-Prussian Blue modified carbon paste electrode (Rh-PBMCPE) for amperometric detection of hydrogen peroxide

Prussian Blue was deposited at carbon paste electrode surface from a solution containing 2.0 x 10-3 mol L-1 K3[Fe(CN)6], 3.0 x 10-3 mol L-1 FeCl3 and 1.0 x 10-2 mol L-1 HCl using two controlled potentials. To improve the stability of the modified electrode it was 50 times cycled in a solution containing 1.0 x 10-3 mol L-1 RhCl3, 0.50 mol L-1 KCl and 0.010 mol L-1 HCl in the potential range from - 0.40 V to 0.60 V at 60 mV s-1. The Rh - Prussian Blue carbon paste modified electrode (Rh-PBMCPE) showed good stability during amperometric catalytic determination of H2O2 at 0.040 V, without ascorbic and uric acids interferences. The current changed linearly with H2O2 concentrations in the range of 5.0 x 10-5 - 8.6 x 10-4 mol L-1. The estimated detection limit was 2.8 x 10-5 mol L-1 with sensibility changing from 1.32 to 0.96 A mol-1 L cm-2 along five days (180 determinations).

prussian blue; rhodium; carbon paste electrode; hydrogen peroxide; amperometry


Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br