Estimate of control voltage tolerances for a photo-electron analyzer of toroidal design

We have run electron optics simulations and determined the tolerance in the control voltages of all elements (retarding input lens, analyzer, accelerating exit lens) of the La Trobe University photoelectron analyzer, recently redesigned to reach a spectral resolution of 5000, and which will be installed at LNLS (Campinas Brasil) and BESSY II (Berlin, Germany).