Fracture strength of teeth restored with milled ultrathin occlusal veneers made of polymer-infiltrated ceramic

Abstract

The purpose was to evaluate, in vitro, the fracture strength of teeth restored with 0.3- or 0.6-mm-thick occlusal veneers made of Vita Enamic® bonded to enamel or dentin. Fifty third molars were allocated into five groups (n=10): Group C - control (sound teeth); Group 0.3E - 0.3-mm-thick occlusal veneers bonded to enamel; Group 0.6E - 0.6-mm-thick occlusal veneers bonded to enamel; Group 0.3D - 0.3-mm-thick occlusal veneers bonded to dentin; Group 0.6D - 0.6-mm-thick occlusal veneers bonded to dentin. After luting with Single Bond Universal and Filtek Z-100, the samples were subjected to fracture strength test in a universal testing machine. The fractures were classified as irreparable and reparable. Data were analyzed with 2-way and 1-way ANOVA (α=.05) followed by Tukey’s test. The teeth restored with 0.3-mm-thick occlusal veneers (2658 N) achieved significantly higher fracture strength compared to the teeth restored with 0.6-mm-thick occlusal veneers (2070 N) (p=0.017). The teeth restored with occlusal veneers bonded to enamel (2648 N) had significantly higher fracture strength compared to the teeth restored with occlusal veneers bonded to dentin (2049 N) (p=0.015). Group C (3355 N) and Group 0.3E (2965 N) did not differ significantly from each other, being the fracture strength significantly higher than the other groups (p<0.05). The fractures were predominantly reparable. The teeth restored with 0.3-mm-thick occlusal veneers achieved higher fracture strength compared to the teeth restored with 0.6-mm-thick occlusal veneers. The supporting substrate influenced the fracture strength, which was higher for the teeth restored with occlusal veneers bonded to enamel.

Key Words:
Dentin; Dental enamel; Ceramics

Resumo

O objetivo foi avaliar, in vitro, a resistência à fratura de dentes restaurados com facetas oclusais de Vita Enamic® com 0,3 ou o,6 mm de espessura e cimentadas em esmalte ou dentina. Cinquenta terceiros molares foram distribuídos em grupos (n=10): Grupo C-controle (dentes hígidos); Grupo 0,3E - facetas oclusais com 0,3 mm de espessura cimentados em esmalte; Grupo 0,6E - facetas com 0,6 mm de espessura cimentados em esmalte; Grupo 0,3D - facetas com 0,3 mm de espessura cimentados em dentina; Grupo 0,6D - facetas com 0,6 mm de espessura cimentados em dentina. Após cimentação com Single Bond Universal Filtek Z-100, as amostras foram submetidas ao teste de resistência à fratura em máquina de ensaio universal. As falhas foram classificadas em reparáveis e irreparáveis. Os resultados foram analisados pelo teste ANOVA de duas vias e uma via (α=0,05) seguido do teste de Tukey. Os dentes restaurados com facetas oclusais com 0,3 mm de espessura (2658 N) obtiveram resistência à fratura estatisticamente superior aos dentes restaurados com facetas com 0,6 mm (2070 N) (p=0.017). Os dentes restaurados com facetas cimentadas em esmalte (2648 N) tiveram resistência à fratura estatisticamente maior comparada aos dentes restaurados com facetas cimentadas em dentina (2049 N) (p=0.015). Grupo C (3355 N) e Grupo 0,3E (2965 N) não diferiram estatisticamente entre si, tendo resistência à fratura estatisticamente superior aos outros grupos (p<0.05). As fraturas foram predominantemente reparáveis. Os dentes restaurados com facetas oclusais com 0,3 mm de espessura apresentaram maior resistência à fratura comparada aos dentes restaurados com facetas com 0,6 mm. O substrato influenciou a resistência à fratura dos dentes restaurados com facetas oclusais, sendo maior quando cimentadas ao esmalte.

Introduction

The continuous reduction of tooth enamel occurs with advancing age. However, there are events that can cause early loss of this tissue, such as erosion due to gastroesophageal reflux, high consumption of acidic foods and beverages, medications, and nervous bulimia. In addition, bruxism is a common parafunctional habit, and one of the possible sequelae is tooth wear 11. Bartlett D, Phillips K, Smith B. A difference in perspective - The North American and European interpretations of tooth wear. Int J Prosthodont1999;12:401-408.. Regardless of the origin, severe wear of dental tissue causes aesthetic and functional problems 22. Al-Omiri MK, Lamey PJ, Clifford T. Impact of tooth wear on daily living. Int J Prothodont 2006;19:601-605..

Ultrathin occlusal veneers have been used to restore the occlusal vertical dimension (OVD) as well as the aesthetic with minimally invasive preparations that allow greater preservation of the dental structure and maintenance of pulp vitality 33. Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra-thin CAD/CAM composite resin and ceramic occlusal veneers for the treatment of severe dental erosion. J Prosthet Dent 2011;105:217-226.,44. Heck K, Paterno H, Lederer A, Litzenburger F, Hickel R, Kunzelmann KH. Fatigue resistance of ultrathin CAD/CAM ceramic and nanoceramic composite occlusal veneers. Dent Mater 2019;35:1370-1377.. Different materials are available for computer-aided design/computer-aided manufacturing (CAD/CAM) technology, which has been used more frequently in the manufacture of ultrathin restorations in the last 10 years 5>5. Davidowitz G, Kotick PG. The use of CAD/CAM in dentistry. Dent Clin North Am 2011;55:559-570..

Among the different materials, lithium disilicate ceramic for the manufacture of ultrathin occlusal veneers has been shown to be a conservative choice to traditional inlays, onlays and full crowns, demonstrating promising results 33. Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra-thin CAD/CAM composite resin and ceramic occlusal veneers for the treatment of severe dental erosion. J Prosthet Dent 2011;105:217-226.,44. Heck K, Paterno H, Lederer A, Litzenburger F, Hickel R, Kunzelmann KH. Fatigue resistance of ultrathin CAD/CAM ceramic and nanoceramic composite occlusal veneers. Dent Mater 2019;35:1370-1377.,66. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548.. Nanoceramic composite represents another alternative for ultrathin occlusal veneers, showing superior results for fatigue resistance in relation to lithium disilicate ceramic 33. Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra-thin CAD/CAM composite resin and ceramic occlusal veneers for the treatment of severe dental erosion. J Prosthet Dent 2011;105:217-226.,44. Heck K, Paterno H, Lederer A, Litzenburger F, Hickel R, Kunzelmann KH. Fatigue resistance of ultrathin CAD/CAM ceramic and nanoceramic composite occlusal veneers. Dent Mater 2019;35:1370-1377.. Another material, called hybrid ceramic, consists of a polymer-infiltrated ceramic network material (PICN). This material has two different components: a network of sintered ceramic infiltrated by a resin matrix. The aim of this new material is to allow ultrathin restorations (0.2-0.5-mm), faster milling without cracks in the ceramic blocks, and good mechanical behavior after luting 77. Dirxen C, Blunck U, Preisser S. Clinical performance of a new biomimetic double network material. Open Dent J 2013;6:118-122..

Few studies have evaluated the fracture strength of ultrathin occlusal veneers made of PICN bonded to enamel 88. Ioannidis A, Mühlemann S, Özcan M, Hüsler J, Hämmerle CHF. Ultra-thin occlusal veneers bonded to enamel and made of ceramic or hybrid materials exhibit load-bearing capacities not different from conventional restorations. J Mech Behav Biomed Mater 2019;90:433-440.,99. Al-Akhali M, Kern M, Elsayed A, Samran A, Chaar MS. Influence of thermomechanical fatigue on the fracture strength of CAD-CAM-fabricated occlusal veneers. J Prosthet Dent 2019;121:644-650. or bonded to dentin 66. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548.,1010. Maeder M, Pasica P, Enderb A, Özcanc M, Benicc GI, Ioannidisc A. Load-bearing capacities of ultra-thin occlusal veneers bonded to dentin. J Mech Behav Biomed Mater. 2019 Jul;95:165-171.. Regarding fracture, different minimal occlusal thicknesses were evaluated, including 0.5 mm 99. Al-Akhali M, Kern M, Elsayed A, Samran A, Chaar MS. Influence of thermomechanical fatigue on the fracture strength of CAD-CAM-fabricated occlusal veneers. J Prosthet Dent 2019;121:644-650.,1111. Al-Akhali M, Chaar MS, Elsayed A, Samran A, Kern M. Fracture resistance of ceramic and polymer-based occlusal veneer restorations. J Mech Behav Biomed Mater 2017;74:245-250., 0.5 mm or 1.0 mm 88. Ioannidis A, Mühlemann S, Özcan M, Hüsler J, Hämmerle CHF. Ultra-thin occlusal veneers bonded to enamel and made of ceramic or hybrid materials exhibit load-bearing capacities not different from conventional restorations. J Mech Behav Biomed Mater 2019;90:433-440.,1010. Maeder M, Pasica P, Enderb A, Özcanc M, Benicc GI, Ioannidisc A. Load-bearing capacities of ultra-thin occlusal veneers bonded to dentin. J Mech Behav Biomed Mater. 2019 Jul;95:165-171., and 0.6 mm or 1.5 mm 66. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548.. The studies concluded that ultrathin occlusal veneers made of PICN are suitable for use in the posterior region 66. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548.,88. Ioannidis A, Mühlemann S, Özcan M, Hüsler J, Hämmerle CHF. Ultra-thin occlusal veneers bonded to enamel and made of ceramic or hybrid materials exhibit load-bearing capacities not different from conventional restorations. J Mech Behav Biomed Mater 2019;90:433-440.,1010. Maeder M, Pasica P, Enderb A, Özcanc M, Benicc GI, Ioannidisc A. Load-bearing capacities of ultra-thin occlusal veneers bonded to dentin. J Mech Behav Biomed Mater. 2019 Jul;95:165-171.. In addition, the concepts of minimally invasive cosmetic dentistry and minimally invasive prosthetic procedure have been rising with time, demanding restorations with very low thicknesses 1212. Yu H, Zhao Y, Li J, Luo T, Gao J, Liu H, Liu W, Liu F, Zhao K, Liu F, Ma C, Setz JM, Liang S, Fan L, Gao S, Zhu Z, Shen J, Wang J, Zhu Z, Zhou X. Minimal invasive microscopic tooth preparation in esthetic restoration: a specialist consensus. Int J Oral Sci 2019; 11:31.. However, it would be interesting to evaluate thinner occlusal veneers, such as 0.3-mm-thick veneers, since the material has this indication 1313. Mainjot AK, Dupont NM, Oudkerk JC, Dewael TY, Sadoun MJ. From artisanal to CAD-CAM blocks: State of the art of indirect composites. J Dent Res 2016;95:487-495..

There are different degrees of dental erosion. The mineral loss may be only on the enamel, or it may progress and expose the dentin 11. Bartlett D, Phillips K, Smith B. A difference in perspective - The North American and European interpretations of tooth wear. Int J Prosthodont1999;12:401-408.,22. Al-Omiri MK, Lamey PJ, Clifford T. Impact of tooth wear on daily living. Int J Prothodont 2006;19:601-605.. Therefore, restorations are bonded to enamel and/or dentin, and the supporting substrate may affect the fracture strength of the restorations 1414. Facenda JC, Borba M, Benetti P, Della Bona A, Corazza PH. Effect of supporting substrate on the failure behavior of a polymer-infiltrated ceramic network material. J Prosthet Dent 2019;121:929-934.. According to Facenda et al. 1414. Facenda JC, Borba M, Benetti P, Della Bona A, Corazza PH. Effect of supporting substrate on the failure behavior of a polymer-infiltrated ceramic network material. J Prosthet Dent 2019;121:929-934. substrates with greater elastic modulus promoted greater fracture strength values of PICN. Therefore, the literature lacks studies assessing, under the same methodology, the fracture strength of thinner occlusal veneers made of PICN, such as 0.3-mm-thick veneers bonded to enamel or dentin substrates. The purpose of this laboratory study was to evaluate the fracture strength of teeth restored with 0.3- or 0.6-mm-thick occlusal veneers made of PICN bonded to enamel or dentin. The hypotheses of the study were that (i) the thickness of the occlusal veneers and (ii) the substrate influence the fracture strength of the teeth restored with ultrathin occlusal veneers.

Material and Methods

Tooth selection

The Ethics Committee (55675416.7.0000.5336) approved the study protocol. Fifty human third molars, extracted for therapeutic reasons and free of restorations, caries, cracks or fractures, were selected after measurement of the mesiodistal (≅ 10.5 mm) and buccal-lingual (≅ 9.5 mm) dimensions with a digital caliper (500-197-20 Mitutoyo, Kawasaki, Japan). A 0.5-mm variation for each measurement was accepted with the objective to standardize the dimensions of the teeth. The teeth were cleaned, disinfected in 0.5% chloramine T solution for 24 h, stored in distilled water at 4°C, and used within 4 months. The teeth were randomly allocated into five different groups (n=10): Group C - control (sound teeth); Group 0.3E - 0.3-mm-thick occlusal veneers made of Vita Enamic® (VITA Zahnfabrik, Bad Säckingen, Germany) bonded to enamel; Group 0.6E - 0.6-mm-thick occlusal veneers bonded to enamel; Group 0.3D - 0.3-mm-thick occlusal veneers bonded to dentin; and Group 0.6D - 0.6-mm-thick occlusal veneers bonded to dentin. One trained operator performed all experimental steps.

Periodontal ligament simulation

The periodontal ligament was artificially created with vinyl polysiloxane. The root portion was covered by a layer of adhesive (Universal Tray Adhesive, Zhermack, Rovigo, Italy) and left to dry. Subsequently, a thin layer of regular viscosity vinyl polysiloxane (Express Standard - 3M, St. Paul, MN, USA) was applied over the adhesive with a brush.

Tooth preparation

The roots of each tooth were mounted vertically in a plastic cylinder with self-cured acrylic resin up to 2 mm from the enamel-cement junction (ECJ). After polymerization of the acrylic resin, the samples were immersed in distilled water at 4ºC. The preparations were standardized. a) Preparation on enamel: The depth of the preparation was made on enamel using the half active tip of a 2135 diamond bur (KG Sorensen, Cotia, SP, Brazil) at high speed under air-water spray. The finish of the preparation was performed with diamond burs 2135F and 2135FF (KG Sorensen, Cotia, SP, Brazil). The cusps inclination was maintained (Figure 1).

Figure 1:
Preparation of the tooth on enamel.

Preparation on dentin: The exposure of occlusal dentin was performed with a 4138 diamond bur (KG Sorensen, Cotia, SP, Brazil) operating at high speed under air-water spray. The buccal and lingual margins were approximately 2.4 to 2.7 mm above the central groove and 5 mm from the ECJ. The cusps inclination was maintained (Figure 2). The finish of the preparation was performed with diamond burs 4138F and 4138FF (KG Sorensen, Cotia, SP, Brazil).

Figure 2:
Preparation of the tooth on dentin.

Manufacturing the restorations

First, reflective titanium (VITA Zahnfabrik, Bad Säckingen, Germany) was applied on the tooth preparation, which was scanned with an optical 3D intraoral camera (Cerec software, version 4.1, Sirona Dental Systems GmbH, Bensheim, Germany). A three-dimensional virtual image was created, and the CAD/CAM occlusal veneers were shaped according to the occlusal anatomy of a right second lower molar. The thicknesses of 0.3- or 0.6-mm occlusal veneers were defined by the software, and the virtual die spacer was set at 50 µm. Forty occlusal veneers made of a polymer-infiltrated ceramic network (PICN - Vita Enamic®, VITA, Zahnfabrik, Germany) were fabricated in the milling unit inLab MC XL (Dentsply Sirona, Charlotte, NC, USA): twenty 0.3-mm-thick occlusal veneers and twenty 0.6-mm-thick occlusal veneers were fabricated. The occlusal veneers were checked and fitted to the prepared teeth and then polished with a specific Vita Enamic clinical polish kit (VITA, Zahnfabrik, Germany).

Bonding procedure

The inner surface of the occlusal veneers was etched with 5% hydrofluoric acid (Condac Porcelain - FGM, Joinville, SC, Brazil) for 60 s, then rinsed off with forceful water spray for 30 s. The silane Scotchbond Ceramic Primer (3M/ESPE, St. Paul MN, USA) was applied and dried gently for 5 s. On the tooth preparation, phosphoric acid at 37% (Condac - FGM, Joinville, SC, Brazil) was applied on enamel for 30 s and on dentin for 15 s, rinsed for 30 s and dried with cotton buds. The adhesive system Single Bond Universal (3M/ESPE, St. Paul MN, USA) was applied to both fitting surfaces of the occlusal veneer and the tooth, and dried gently for 5 s. The resin composite Filtek Z-100 (3M/ESPE, St. Paul MN, USA), preheated to 69ºC (HotSet - Technolife), was applied to the inner surface of the occlusal veneer. Immediately, the occlusal veneer was positioned on the preparation, followed by the application of a 1-kg load. The excess of preheated resin composite was removed, and each face of the occlusal veneer was light-cured for 20 s with an LED light unit (VALO, Ultradent, South Jordan, UT, USA) with a light intensity of 1,000 mW/cm2. The light intensity was monitored with an LED radiometer (SDI, Bayswater, Vic, Australia). The samples were kept in distilled water at 37°C for 24 h.

Fracture strength testing

After storage, the samples were submitted to a fracture strength test using a DL-2000 universal testing machine (EMIC, São José dos Pinhais, PR, Brazil) with a 10-kN cell load. A 6-mm-diameter metal sphere, connected to the load cell, was positioned on the occlusal surface of the veneers to achieve tripodization of contacts along the cuspal inclines over the central fossa. The compression load was applied at a crosshead speed of 1.0 mm/min until failure. The values of maximum force were recorded in Newtons (N).

Fracture mode analysis

After visual examination, the fractures were classified as follows: a) reparable (fracture at the occlusal veneer only, or fracture that involved one or more cusps); or b) irreparable (fracture divided the tooth into two parts at the level of the pulp chamber floor).

Statistical analysis

SPSS 10.0 software (SPSS Inc., Chicago, IL, USA) was used for the statistical analysis. The Kolmogorov-Smirnov test showed data normality. The fracture strength values of the experimental groups were evaluated by two-way analysis of variance (ANOVA) (occlusal veneer thickness x substrate) and the Tukey post hoc test. One-way ANOVA and the Tukey post hoc test were used to evaluate the fracture strength values of sound teeth with those of the experimental groups (significance level 0.05).

Results

Two-way ANOVA demonstrated that the thickness (p=0.005) and substrate (p=0.004) had significant effects on the fracture strength. The interaction of these factors (p=0.742) was not significant.

Regardless of the substrate, the teeth restored with 0.3-mm-thick occlusal veneers (2658 N) showed significantly higher fracture strength than the teeth restored with 0.6-mm-thick occlusal veneers (2070 N) (Table 1).

Table 1
Fracture strength mean (N) between occlusal veneer thicknesses.

Regardless of the thickness, the teeth restored with occlusal veneers bonded to enamel (2648 N) showed significantly higher fracture strength than the teeth restored with occlusal veneers bonded to dentin (2049 N) (Table 2).

Table 2
Fracture strength mean (N) between substrates.

According to one-way ANOVA, there was no significant difference between Group C (3355 N) and Group 0.3E (2965 N), with fracture strengths that were significantly higher than those of the other groups. Group 0.6E (2331 N), Group 0.3D (2315 N) and Group 0.6D (1808 N) did not differ significantly from each other (Table 3).

Table 3
Fracture strength mean (N) and standard deviation in the different groups.

There was a predominance of reparable fractures in all groups (Table 4). Regarding reparable fractures, there were fractures restricted to restoration only in Group 0.6E (Figure 3). There was predominance of fractures in one or more cusps in Group C, Group 0.3E, Group 0.3D, and Group 0.6D (Figure 4). Irreparable fractures occurred in a few specimens in all groups (Figure 5) (Table 4).

Figure 3
Occlusal veneer fracture (classified as reparable).

Figure 4
Cusp fracture (classified as reparable).

Figure 5
Longitudinal fracture of the tooth into two parts at the pulpal chamber floor (classified as irreparable).

Table 4
Failure analyses of the different groups.

Discussion

In the present study, higher fracture strength was achieved for the teeth restored with 0.3-mm-thick occlusal veneers made of PICN compared to the teeth restored with 0.6-mm-thick occlusal veneers, regardless of whether they were bonded to enamel or dentin. This result showed that even under a constant compressive load, an ultrathin occlusal veneer, such as 0.3 mm thick, achieved high strength. Therefore, the first hypothesis was accepted. Studies have shown higher fracture strength for 1.0-mm-thick compared to 0.5-mm-thick occlusal veneers 1010. Maeder M, Pasica P, Enderb A, Özcanc M, Benicc GI, Ioannidisc A. Load-bearing capacities of ultra-thin occlusal veneers bonded to dentin. J Mech Behav Biomed Mater. 2019 Jul;95:165-171. and higher fracture strength for 1.5-mm-thick compared to 0.6-mm-thick 66. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548. occlusal veneers made of PICN. However, in the present study, thicker PICN occlusal veneer did not contribute to higher fracture strength values when comparing 0.3- and 0.6-mm-thick restorations. This result may be related to the possibility of milling PICN material to thin restorations that are still strong enough to avoid cracks, which are stopped by the interpenetrating polymer within the ceramic network 77. Dirxen C, Blunck U, Preisser S. Clinical performance of a new biomimetic double network material. Open Dent J 2013;6:118-122.. This combination of materials aims to increase the mechanical properties, such as strain at failure and flexural strength 1515. Coldea A, Swain MV, Thiel N. Mechanical properties of polymer-infiltrated-ceramic-network materials. Dent Mater 2013;29:419-426.. Possibly, in very low thicknesses such as those used in the present study, the intrinsic strength of the material at 0.6-mm-thick has not been decisive to obtain higher fracture strength in comparison to 0.3-mm-thick. Another possible explanation for the higher fracture strength of 0.3-mm-thick occlusal veneers is the bonding protocol.

The bonding protocol of the PICN occlusal veneers was standardized. The Single Bond Universal adhesive system was applied on enamel and dentin in the etch-and-rinse technique. The study of Al-Akhali et al. 99. Al-Akhali M, Kern M, Elsayed A, Samran A, Chaar MS. Influence of thermomechanical fatigue on the fracture strength of CAD-CAM-fabricated occlusal veneers. J Prosthet Dent 2019;121:644-650. demonstrated that the self-etch technique on enamel decreased the survival rate of Vita Enamic® ultrathin occlusal veneers. Hydrofluoric acid at 5% was applied on the inner surface of the occlusal veneers for 60 s, followed by the silane. Silane promotes chemical bonding between the silicon contained in the materials and the adhesive material used for bonding 1616. Spohr AM, C Sobrinho LC, Consani S, Sinhoreto MAC, Knowles JC. Influence of surface conditions and silane agent on the bond of resin to IPS Empress 2 ceramic. Int J Prosthodont2003;16:277-282.. Although Single Bond Universal has silane in its composition, studies have shown that the application of silane as a separate step provides greater values of bond strength, independent of the presence of silane within the universal adhesive solution 1717. Moro AFV, Ramos AB, Rocha GM, Perez CDR. Effect of prior silane application on the bond strength of a universal adhesive to a lithium disilicate ceramic. J Prosthet Dent. 2017; 118:666-671..

Following the bonding protocol, a light-curing resin composite was used to bond the occlusal veneers to the tooth preparation. This material was heated, allowing greater flow during the luting procedure. The advantage of using a light-curing resin composite is the extended time to place the restorations in comparison to a dual-cure material. This extended time compensates for the adversity of positioning the restorations because of the lack of an insertion path 33. Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra-thin CAD/CAM composite resin and ceramic occlusal veneers for the treatment of severe dental erosion. J Prosthet Dent 2011;105:217-226.. In addition, resin composite may perform better than resin cement on restoration margins due to more inorganic load filling 1818. Lopes LCP, Terada RSS, Tsuzuki FM, Giannini M, Hirata R. Heating and preheating of dental restorative materials - a systematic review. Clin Oral Investig 2020; 24:4225-35.. Regarding photopolymerization, an important factor related to the results of the present study can be the thickness of the ceramic that influenced the light transmittance and degree of conversion of the resinous materials 1919. Oh S, Shin SM, Kim HJ, Paek J, Kim SJ, Yoon TH, Kim SY. Influence of glass-based dental ceramic type and thickness with identical shade on the light transmittance and the degree of conversion of resin cement. Int J Oral Sci. 2018; 10:1-5.. Then, it cannot be ruled out the possibility that the 0.3-mm-thick occlusal veneers have contributed to a lower attenuation of the transmitted light in comparison with the 0.6-mm-thick occlusal veneers. The possible higher degree of conversion obtained for the teeth restored with 0.3-mm-thick occlusal veneers may have contributed to higher mechanical properties of the Single Bond Universal and Filtek Z-100, promoting higher fracture strength.

A single unit between the tooth and occlusal veneer is obtained with the adhesive luting. Consequently, the forces applied on the restoration bonded to the tooth are dissipated through the tooth structure, periodontal ligament and alveolar bone, preventing concentration of the forces in the inner portion of the restoration, which would lead to fracture 2020. Attia A, Abdelaziz KM, Freitag S, Kern M. Fracture load of composite resin and feldspathic all ceramic CAD/CAM crowns. J Prosthet Dent 2006; 95:117-123.. The fracture types observed in the study corroborate this finding. Most of the reparable fractures occurred in one or more cusps, and there were few irreparable fractures, showing that load dissipated through the tooth. Only in Group 0.6E occurred fractures restricted to the occlusal veneer, without any damage to the tooth substrate. Therefore, the thicker occlusal veneers bonded to enamel transmitted fewer forces to the substrate, concentrating the stresses in the internal part of the restoration.

Vita Enamic® is a brand of PICN. This material associates the properties of ceramic and polymer as a hybrid structure with two interpenetrating networks of ceramic (86 wt%) and polymer (14 wt%) 77. Dirxen C, Blunck U, Preisser S. Clinical performance of a new biomimetic double network material. Open Dent J 2013;6:118-122.. Vita Enamic® presents a flexural strength of 131 MPa and an elastic modulus of 26.3 GPa 2121. de Kok P, Kleverlaan CJ, Jageer N, Kuijs R, Feilzer A. Mechanical performance of implant-supported posterior crowns. J Prosthet Dent 2015; 114:59-66., which is close to that of dentin (18.3 GPa) 2222. Craig RG, Powers JM. Restorative Dental Materials. São Paulo: Santos; 2004. In addition, Vickers hardness of this material lies between enamel and dentin 77. Dirxen C, Blunck U, Preisser S. Clinical performance of a new biomimetic double network material. Open Dent J 2013;6:118-122.. Ceramics with lower elastic modulus, closer to that of dentin, can better resist loads and better avoid crack propagation 1515. Coldea A, Swain MV, Thiel N. Mechanical properties of polymer-infiltrated-ceramic-network materials. Dent Mater 2013;29:419-426.. In this study, the occlusal veneers were bonded to preparations on enamel only or were bonded to preparations on dentin with enamel margins. The enamel elastic modulus (84 GPa) is greater than the dentin elastic modulus (18.3 GPa) 2222. Craig RG, Powers JM. Restorative Dental Materials. São Paulo: Santos; 2004. The study of Piemjai et al. 2323. Piemjai M, Arksornnukit M. Compressive fracture resistance of porcelain laminates bonded to enamel or dentin with four adhesive systems. J Prosthodont 2007;16:457-464. showed that the fracture strength of restorations bonded to enamel was greater than that of those bonded to dentin. The present study corroborates these findings since the teeth restored with occlusal veneers bonded to enamel achieved significantly higher fracture strength than the teeth restored with occlusal veneers bonded to dentin. Thus, the second hypothesis was accepted.

Group C (sound teeth) achieved higher fracture strength and did not differ significantly from Group 0.3E. A possible factor related to this finding is the wear of the dental structure that was less in enamel preparation in comparison with dentin preparation, preserving greater dental substrate. However, regardless of the substrate (enamel or dentin) or the restoration thickness (0.3 or 0.6 mm), the fracture strength values were higher than the bite forces registered in patients. In forced bite tests, the value of 922 N was reached, and 1,120 N in patients with bruxisms 2424. Sakaguchi RL, Douglas WH, Delong R, Pintado MR. The wear of a posterior composite in an artificial mouth: a clinical correlation. Dent Mater 1986;2:235-240.. These are important data, showing that even with a 0.3-mm-thick veneer in the occlusal region; the restored teeth had a fracture strength similar to that of sound teeth.

Structural defects in ceramics and their absence of plastic deformation contribute to the strength of these materials 2222. Craig RG, Powers JM. Restorative Dental Materials. São Paulo: Santos; 2004. PICN restorations, such as Vita Enamic®, are obtained by CAD/CAM technology. The ceramic blocks of PICN have a more homogeneous structure, with fewer defects 1313. Mainjot AK, Dupont NM, Oudkerk JC, Dewael TY, Sadoun MJ. From artisanal to CAD-CAM blocks: State of the art of indirect composites. J Dent Res 2016;95:487-495.. This is an important characteristic and must have a significant contribution to the high fracture strength values of the teeth restored with ultrathin occlusal veneers. The stresses are more effectively distributed in a more homogeneous material and promote strength against breakdown phenomena 1313. Mainjot AK, Dupont NM, Oudkerk JC, Dewael TY, Sadoun MJ. From artisanal to CAD-CAM blocks: State of the art of indirect composites. J Dent Res 2016;95:487-495..

Most of the failures were classified as reparable. In this case, the tooth structure would not be lost, and a new restorative procedure could be performed. However, the interpretation of the results must include the limitations of a laboratory experiment that do not reproduce the complexity of the oral environment or the masticatory biomechanics. During the fracture strength test, a constant axial load was applied to the sample until fracture. It can be considered a limitation of the study since fractures by a constant axial load do not occur clinically 2525. Shirakura A, Lee H, Geminiani A, Ercoli C, Feng C. The influence of veneering porcelain thickness of all-ceramic and metal ceramic crowns on failure resistance after cyclic loading. J Prosthet Dent 2009;101:119-127.. Despite this limitation, the results of this mechanical test allow a comparison of the fracture strengths among different restorative techniques and with sound teeth. Another limitation could be the small sample size of the study, even though it was possible to show significant differences in fracture strength between the substrates as well as between the different thicknesses of the occlusal veneers.

The results obtained in this study seem to provide good perspectives in relation to the clinical use of ultrathin occlusal veneers made of PICN in cases in which the remaining dental structure must be preserved or in cases in which there is limited interocclusal space due to the wear of the teeth. In addition, the dentists are advised to maintain as much enamel surface area as possible during tooth preparation, since the results of fracture strength are higher when the occlusal veneers were bonded to this substrate.

Considering the limitations of this study, it was possible to conclude that the teeth restored with 0.3-mm-thick occlusal veneers achieved higher fracture strength than the teeth restored with 0.6-mm-thick occlusal veneers. The substrate influenced the fracture strength of the teeth restored with ultrathin occlusal veneers, which was higher when bonded to enamel.

Acknowledgments

This study was funding in part by the Coordination for the Improvement of Higher Education Personnel (CAPES, Brazil) - Finance Code 001.

References

  • 1
    Bartlett D, Phillips K, Smith B. A difference in perspective - The North American and European interpretations of tooth wear. Int J Prosthodont1999;12:401-408.
  • 2
    Al-Omiri MK, Lamey PJ, Clifford T. Impact of tooth wear on daily living. Int J Prothodont 2006;19:601-605.
  • 3
    Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra-thin CAD/CAM composite resin and ceramic occlusal veneers for the treatment of severe dental erosion. J Prosthet Dent 2011;105:217-226.
  • 4
    Heck K, Paterno H, Lederer A, Litzenburger F, Hickel R, Kunzelmann KH. Fatigue resistance of ultrathin CAD/CAM ceramic and nanoceramic composite occlusal veneers. Dent Mater 2019;35:1370-1377.
  • >5
    Davidowitz G, Kotick PG. The use of CAD/CAM in dentistry. Dent Clin North Am 2011;55:559-570.
  • 6
    Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett L Júnior H, Spohr AM. Effect of different computer-aided design/computer-aided manufacturing (CAD/CAM) materials and thicknesses on the fracture resistance of occlusal veneers. Oper Dent 2018;43:539-548.
  • 7
    Dirxen C, Blunck U, Preisser S. Clinical performance of a new biomimetic double network material. Open Dent J 2013;6:118-122.
  • 8
    Ioannidis A, Mühlemann S, Özcan M, Hüsler J, Hämmerle CHF. Ultra-thin occlusal veneers bonded to enamel and made of ceramic or hybrid materials exhibit load-bearing capacities not different from conventional restorations. J Mech Behav Biomed Mater 2019;90:433-440.
  • 9
    Al-Akhali M, Kern M, Elsayed A, Samran A, Chaar MS. Influence of thermomechanical fatigue on the fracture strength of CAD-CAM-fabricated occlusal veneers. J Prosthet Dent 2019;121:644-650.
  • 10
    Maeder M, Pasica P, Enderb A, Özcanc M, Benicc GI, Ioannidisc A. Load-bearing capacities of ultra-thin occlusal veneers bonded to dentin. J Mech Behav Biomed Mater. 2019 Jul;95:165-171.
  • 11
    Al-Akhali M, Chaar MS, Elsayed A, Samran A, Kern M. Fracture resistance of ceramic and polymer-based occlusal veneer restorations. J Mech Behav Biomed Mater 2017;74:245-250.
  • 12
    Yu H, Zhao Y, Li J, Luo T, Gao J, Liu H, Liu W, Liu F, Zhao K, Liu F, Ma C, Setz JM, Liang S, Fan L, Gao S, Zhu Z, Shen J, Wang J, Zhu Z, Zhou X. Minimal invasive microscopic tooth preparation in esthetic restoration: a specialist consensus. Int J Oral Sci 2019; 11:31.
  • 13
    Mainjot AK, Dupont NM, Oudkerk JC, Dewael TY, Sadoun MJ. From artisanal to CAD-CAM blocks: State of the art of indirect composites. J Dent Res 2016;95:487-495.
  • 14
    Facenda JC, Borba M, Benetti P, Della Bona A, Corazza PH. Effect of supporting substrate on the failure behavior of a polymer-infiltrated ceramic network material. J Prosthet Dent 2019;121:929-934.
  • 15
    Coldea A, Swain MV, Thiel N. Mechanical properties of polymer-infiltrated-ceramic-network materials. Dent Mater 2013;29:419-426.
  • 16
    Spohr AM, C Sobrinho LC, Consani S, Sinhoreto MAC, Knowles JC. Influence of surface conditions and silane agent on the bond of resin to IPS Empress 2 ceramic. Int J Prosthodont2003;16:277-282.
  • 17
    Moro AFV, Ramos AB, Rocha GM, Perez CDR. Effect of prior silane application on the bond strength of a universal adhesive to a lithium disilicate ceramic. J Prosthet Dent. 2017; 118:666-671.
  • 18
    Lopes LCP, Terada RSS, Tsuzuki FM, Giannini M, Hirata R. Heating and preheating of dental restorative materials - a systematic review. Clin Oral Investig 2020; 24:4225-35.
  • 19
    Oh S, Shin SM, Kim HJ, Paek J, Kim SJ, Yoon TH, Kim SY. Influence of glass-based dental ceramic type and thickness with identical shade on the light transmittance and the degree of conversion of resin cement. Int J Oral Sci. 2018; 10:1-5.
  • 20
    Attia A, Abdelaziz KM, Freitag S, Kern M. Fracture load of composite resin and feldspathic all ceramic CAD/CAM crowns. J Prosthet Dent 2006; 95:117-123.
  • 21
    de Kok P, Kleverlaan CJ, Jageer N, Kuijs R, Feilzer A. Mechanical performance of implant-supported posterior crowns. J Prosthet Dent 2015; 114:59-66.
  • 22
    Craig RG, Powers JM. Restorative Dental Materials. São Paulo: Santos; 2004
  • 23
    Piemjai M, Arksornnukit M. Compressive fracture resistance of porcelain laminates bonded to enamel or dentin with four adhesive systems. J Prosthodont 2007;16:457-464.
  • 24
    Sakaguchi RL, Douglas WH, Delong R, Pintado MR. The wear of a posterior composite in an artificial mouth: a clinical correlation. Dent Mater 1986;2:235-240.
  • 25
    Shirakura A, Lee H, Geminiani A, Ercoli C, Feng C. The influence of veneering porcelain thickness of all-ceramic and metal ceramic crowns on failure resistance after cyclic loading. J Prosthet Dent 2009;101:119-127.