Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma – method-comparison evaluations

Analisador portátil para dosar lactato no líquido cefalorraquidiano e plasma – comparação de método

Increased plasma lactate levels can indicate the presence of metabolic disorders in HIV infected individuals.

Objective:

To determine whether a portable analyzer is valid for measuring cerebrospinal fluid (CSF) and plasma lactate levels in HIV infected individuals.

Method:

CSF and plasma were collected from 178 subjects. Samples tested by the Accutrend® portable analyzer were compared to those tested by a reference device (SYNCHRON LX® 20).

Results:

The portable analyzer had in plasma sensitivity of 0.95 and specificity 0.87. For CSF the specificity was 0.95; the sensitivity 0.33; the negative predictive value was 95% and the positive predictive value 33%.

Conclusions:

These findings support the validity of the portable analyzer in measuring lactate concentrations in CSF that fall within the normal range. The relatively poor positive predictive value indicates that a result above the reference range may represent a “false positive test”, and should be confirmed by the reference device before concluding abnormality.

HIV; cerebrospinal fluid; clinical chemistry; lactate