SUMMARY
Objective:
The presente study aims to draw, build and test a simple and low cost surgical keratometer.
Methods:
The principle described by Plácido da Costa was used to construct the prototype. A circular ring of light was built and attached to a surgical microscope. The zoom system present in the microscope provided a bigger or smaller image of the ring to the observer. lts size would then be compared to a circular reticle located inside the ocular lens of the microscope.Numeric values were achieved by a graduated scale. The flattest and steepest meridians were also located.
Results:
A precision between 0.29 and 0.62D h as been achieved in the evaluation of astigmatism. The meridian orientation showed a precision of about 10 degrees of inclination in astigmatisms greater than 1.60 D, and 6 degrees in astigmatisms greater than 1.96 D.
Conclusion:
This dévice showed a slightly inferior precision in astigmatism measurement, when compared to conventional surgical keratometers (0.10 to 0.25 D). Equal to high precision was observed when evaluating meridian location (conventional precision = 10 degrees). Itshould also be mentioned that its low cost and easy use may render it a very useful device.
Keywords:
Cornea surgery; Instrumentation
RESUMO
Objetivos:
Desenhar, construir e testar um ceratômetro cirúrgico simples e de baixo custo.
Métodos:
Para a montagem do aparelho foram utilizados os princípios descritos por Plácido da Costa. Um anel circular iluminado foi construído e acoplado a um microscópio cirúrgico. O sistema de "zoom " do microscópio era acionado pelo observador, que deveria aumentar ou diminuir o tamanho da imagem do anel, produzida pela superficie de teste, até a coincidência com uma retícula circular, presente na lente ocular do microscopio. Uma escala graduada foi montada para a obtenção de valores numéricos. A localização dos meridianos também foi realizada.
Resultados:
O aparelho apresenta uma precisão de 0,29 a 0,62 D, na medida do astigmatismo. A precisão na medida da localização do astigmatismo é de aproximadamente 1O graus de inclinação, em astigmatismos maiores do que 1,60 D e de aproximadamente 6 graus em astigmatismos maiores do que 1,96 D.
Conclusão:
O aparelho tem precisão pouco inferior aos ceratômetros convencionais (0,10 a 0,25 D) na determinação do valor do astigmatismo. A precisão na orientação do meridiano mais plano/curvo é igual ou superior aos aparelhos convencionais (precisão de 10 graus). Além disso, seu baixo custo e alta simplicidade de uso, devem torná-lo muito útil em nosso meio.
Palavras-chave:
Córnea - Cirurgia; Instrumentação
Full text available only in PDF format.
REFERENCES
-
1Pearlman JT. Hippocrates and Ophthalmology. Am J Ophthalmol 1969;68:1069-76.
-
2Bores LD (ed). Refractive Eye Surgery. Boston: Blackwell Scientific Publications, 1993.
-
3Lebensohn JE. A chronology of ophthalmic progress. Am J Ophthalmol 1965;59:883-95.
-
4Bicas HEA: Sobre um novo princípio de ceratometria (e sua aplicação através de um instrumento servindo também a outras finalidades, adaptável aos biomicroscópios. Tese (Doutorado), Ribeirão Preto, USP, 1967.
-
5Gamma L. Estudos sobre a oftalmometria. Rev Med (RJ) 1873;1:39.
-
6Prado D. De Wollaston a Javal. Arq Bras Oftalmol 1941; 4:342-8.
-
7Barraquer JJ. The history of the microscope in ocular surgery. J Microsurg 1980; I:288-99.
-
8Troutman R, Kelly S, Kaye D, Clahane A. The use and preliminary results of the Troutman surgical keratometer in cataract and corneal surgery. Trans Am Acad Ophthalmol Otolaryngol 1977;83:232-8.
-
9Schor P, Miller D. Diferentes tipos de topógrafos corneanos. In: Bechara S, Nosé W, eds. Cirurgia Refrativa. São Paulo:. ... ?
-
10Fowler CW, Dave TN. Review of past and present techniques of measuring corneal topography. Ophthalmol Physiol Opt 1994;14:49-58.
-
11Morlet N, Lindsay P, Cooke P. A comparison of two semi-quantitative surgical keratometer: the modified Hyde ruler and the Barrett keratoscope with "astigmatic dial". Ophthalmic Surg 1994;25:144-9.
-
12Levene JR. An evaluation of the of the hand keratoscope as a diagnostic instrument for corneal astigmatism. Thesis (MSc), Indiana University, 1961.
-
13Frantz J. Reidy, McDonald M. A comparison of surgical keratometers. Refract Corneal Surg 1989;5:409-13.
Publication Dates
-
Publication in this collection
June 1998