Acessibilidade / Reportar erro

Evaluation and comparison between methods of measurement of systolic blood pressure in healthy conscious cats

ABSTRACT

After two different validation systolic blood pressure measurement methods (PAS), not invasive or indirect (oscillometric and doppler vascular) with invasive or direct (arterial catheterization - Gold Standard) in 12 anesthetized healthy cats of the same age group, not invasive methods were evaluated and compared in 24 healthy conscious cats, divided into four age groups. In each animal, we proceeded to measure the pressure by Doppler and oscillometric methods. The oscillometric method was the first to be performed, and, shortly thereafter, the doppler method. The values considered were obtained from averages from seven consecutive measurements for each method with an interval of thirty seconds between them. The first measurement and disparate measurements (range ˃ 20% PAS) obtained with obvious signs of stress and / or animal movement were discarded. Methods of oscillometric and doppler measurement, although statistically different averages were shown in all groups when compared, each group showed a significant high positive correlation, meaning that when a high value is obtained by a method, the same occurs with the other method, and vice versa. Clinically, the observed difference is not relevant since the mean difference values obtained by the two methods, in each group was lower than 1.1%.

Keywords:
feline; doppler; oscillometric; blood pressure

Universidade Federal de Minas Gerais, Escola de Veterinária Caixa Postal 567, 30123-970 Belo Horizonte MG - Brazil, Tel.: (55 31) 3409-2041, Tel.: (55 31) 3409-2042 - Belo Horizonte - MG - Brazil
E-mail: abmvz.artigo@gmail.com