An Alternative Method to Calculate Simplified Projected Aortic Valve Area at Normal Flow Rate

Abstract

Background:

Simplified projected aortic valve area (EOAproj) is a valuable echocardiographic parameter in the evaluation of low flow low gradient aortic stenosis (LFLG AS). Its widespread use in clinical practice is hampered by the laborious process of flow rate (Q) calculation.

Objetive:

This study proposes a less burdensome, alternative method of Q calculation to be incorporated in the original formula of EOAproj and measures the agreement between the new proposed method of EOAproj calculation and the original one.

Methods:

Retrospective observational single-institution study that included all consecutive patients with classic LFLG AS that showed a Q variation with dobutamine infusion ≥ |15|% by both calculation methods.

Results:

Twenty-two consecutive patients with classical LFLG AS who underwent dobutamine stress echocardiography were included. Nine patients showed a Q variation with dobutamine infusion calculated by both classical and alternative methods ≥ |15|% and were selected for further statistical analysis. Using the Bland-Altman method to assess agreement we found a systematic bias of 0,037 cm2 (95% CI 0,004 - 0,066), meaning that on average the new method overestimates the EOAproj in 0,037 cm2 compared to the original method. The 95% limits of agreement are narrow (from -0,04 cm2 to 0,12 cm2), meaning that for 95% of individuals, EOAproj calculated by the new method would be between 0,04 cm2 less to 0,12 cm2 more than the EOAproj calculated by the original equation.

Conclusion:

The bias and 95% limits of agreement of the new method are narrow and not clinically relevant, supporting the potential interchangeability of the two methods of EOAproj calculation. As the new method requires less additional measurements, it would be easier to implement in clinical practice, promoting an increase in the use of EOAproj.

Keywords:
Aortic Valve Stenosis / diagnosis; Aortic Valve Stenosis / diagnostic imaging; Echocardiography, Stress; Heart Valves / physiopathology

Resumo

Fundamento:

A área valvular aórtica projetada simplificada (AEOproj) é um parâmetro ecocardiográfico valioso na avaliação da estenose aórtica de baixo fluxo e baixo gradiente (EA BFBG). Sua utilização na prática clínica é limitada pelo trabalhoso processo de cálculo da taxa de fluxo (Q).

Objetivos:

Este estudo propõe um método alternativo, menos complexo, para o cálculo da Q para ser incorporado na fórmula original da AEOproj, e mede a concordância entre o novo método proposto para o cálculo da AEOproj em comparação ao método original.

Métodos:

Estudo retrospectivo, observacional, unicêntrico que incluiu todos os pacientes com AE BFBG clássica com variação da Q com infusão de dobutamina ≥ |15|% por ambos os métodos.

Resultados:

Foram incluídos 22 pacientes consecutivos com AE BFBG clássico, que se submeteram à ecocardiografia sob estresse com dobutamina. Nove pacientes apresentaram uma variação da Q com infusão de dobutamina calculada tanto pelo método clássico como pelo método alternativo ≥ |15|%, e foram selecionados para análise estatística. Utilizando método Bland-Altman para avaliar a concordância, encontramos um viés sistemático de 0,037 cm2 (IC 95% 0,004 - 0,066), o que significa que, em média, o novo método superestima a AEOproj em 0m037 cm2 em comparação ao método original. Os limites de concordância de 95% são estreitos (de -0,04 cm2 a 0,12 cm2), o que significa que para 95% dos indivídios, a AEOproj calculada pelo novo método estaria entre 0,04 cm2 menos a 0,12 cm2 mais que a AEOproj calculada pela equação original.

Conclusão:

O viés e os limites de 95% de concordância do novo método são estreitos e não são clinicamente relevantes, o que corrobora a intercambialidade dos dois métodos de cálculo da AEOproj. Uma vez que o novo método requer menos medidas, seria mais fácil de ser implementado na prática clínica, promovendo um aumento na utilização da AEOproj.

Palavras-chave:
Estenose da Valva Aórtica / diagnóstico; Estenose da Valva Aórtica / diagnóstico por imagem; Ecocardiografia sob Estresse; Valvas Cardíacas / fisiopatologia

Introduction

Classical low-flow, low-gradient (LFLG) aortic stenosis (AS) is characterized by the combination of a calcified aortic valve with an effective orifice area (EOA) compatible with severe stenosis, a low transvalvular velocity or pressure gradient suggestive of moderate stenosis and a low left ventricular ejection fraction (LVEF).11 Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Guyton RA, et al; ACC/AHA Task Force Members. 2014 AHA/ACC Guideline for the Management of Patients With Valvular Heart Disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation. 2014;129(23):e521-643. doi: 10.1161/CIR.0000000000000031. Erratum in: Circulation. 2014;130(13):e120.
https://doi.org/10.1161/CIR.000000000000...
Dobutamine stress echocardiography (DSE) may aid in the distinction between patients with true severe AS and those with pseudo-severe AS by promoting a potential increase in flow. Hence, traditional hemodynamic indices of stenosis severity could be evaluated at normal flow rates and easily interpreted.22 Awtry E, Davidoff R. Low-flow/low-gradient aortic stenosis. Circulation. 2011;124(23):e739-41. doi: 10.1161/CIRCULATIONAHA.111.075853.
https://doi.org/10.1161/CIRCULATIONAHA.1...
The main limitation of this exam is the unpredictability of flow augmentation, leading to ambiguous changes of mean pressure gradient and EOA.33 Garnier F, Eicher JC, Jazayeri S, Bertaux G, Bouchot O, Aho LS, et al. Usefulness and limitations of contractile reserve evaluation in patients with low-flow, low-gradient aortic stenosis eligible for cardiac resynchronization therapy. Eur J Heart Fail. 2014;16(6):648-54. doi: 10.1002/ejhf.78.
https://doi.org/10.1002/ejhf.78...
Projected aortic valve area at normal transvalvular flow rate (250 mL/min) - EOAproj - is an echocardiographic parameter that was developed in order to overcome this limitation. It consists of the effective orifice aortic area that would have occurred at a standardized flow rate of 250 mL/min, enabling the comparison of AS severity between patients with different flow rate profiles with dobutamine infusion.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
The determination of this new parameter requires the calculation of at least the basal and peak flow rate in each patient. The original formula of EOAproj published by Blais et al. proposed the calculation of flow rate as the quotient between stroke volume and the ejection time (ET), which requires 3 different measurements: 1) left ventricular outflow tract (LVOT) diameter; 2) LVOT velocity-time integral and 3) ET measured at the aortic velocity spectrum.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
Flow rate can also be determined by the product of LVOT area and LVOT mean velocity, which requires only 2 measurements: 1) LVOT diameter and 2) LVOT mean velocity.55 Quiñones MA, Otto CM, Stoddard M, Waggoner A, Zoghbi WA; Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. Recommendations for quantification of Doppler echocardiography: A report from the Doppler quantification task force of the nomenclature and standards committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167-84. PMID: 11836492. This alternative method to calculate flow rate is less cumbersome and less susceptible to inter-observer and intra-observer variability as it requires less measurements.

The aim of the present study is to measure the agreement between two methods of calculation of simplified EOAproj using two different approaches of flow rate determination in patients with classical LFLG AS.

Methods

Retrospective observational single-institution study that included all consecutive patients with LFLG AS with depressed LVEF (definition in accordance with the 2014 AHA/ACC Guidelines for the Management of Valvular Heart Disease11 Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Guyton RA, et al; ACC/AHA Task Force Members. 2014 AHA/ACC Guideline for the Management of Patients With Valvular Heart Disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation. 2014;129(23):e521-643. doi: 10.1161/CIR.0000000000000031. Erratum in: Circulation. 2014;130(13):e120.
https://doi.org/10.1161/CIR.000000000000...
) referred for DSE evaluation between September/2011 and November/2015.

Patients admitted to the study had to fulfill all the following criteria: 1) age ≥ 18 years old; 2) EOA ≤ 1.0 cm2 or EOA indexed to body surface area ≤ 0.6 cm2/m2 and maximal transaortic velocity (Vmax) < 4 m/s or mean transaortic gradient (Gmean) < 40 mmHg and 3) LVEF < 50%. Patients with more than mild aortic regurgitation or more than mild mitral regurgitation or stenosis were excluded.

After completing DSE, patients were classified into groups in terms of severity of the stenosis in agreement with the 2014 AHA/ACC Guidelines for the Management of Valvular Heart Disease:

  • Patients with true severe LFLG AS: EOA ≤ 1.0 cm2 with Vmax ≥ 4 m/s at any flow rate

  • • Patients who did not fulfill the criteria for true severe LFLG AS having: a) EOA ≤ 1.0 cm2 with Vmax < 4m/s (persistent area - gradient mismatch), b) EOA > 1.0 cm2 with Vmax ≥ 4 m/s or c) EOA > 1.0 cm2 with Vmax < 4 m/s (pseudo-severe AS)

Echocardiographic assessment

Echocardiographic examination was performed using commercially available equipment (Vivid - 7; General Electric Vingmed, Milwaukee, WI) with a 3.5-MHz transducer.

After the acquisition of the baseline study, a low dose dobutamine infusion protocol was begun at 5 ug/Kg body weight per minute, titrated upward in stages of 5 ug/Kg per minute every 5 minutes up to a maximal dose of 20 ug/Kg per minute. Systemic blood pressure and the 12-lead electrocardiogram were monitored throughout the test. Continuous wave Doppler of the aortic valve velocity spectrum and pulsed-wave Doppler of the LVOT velocity spectrum were recorded at baseline and in the last 2 minutes of each stage of the protocol. LVOT diameter was measured in the basal parasternal long axis view and was assumed to have remained constant during the test protocol.

Raw data was stored digitally and analysis was performed off-line by a single independent operator, using the EchoPac Clinical Workstation Software (General Electric, Vingmed, Milwaukee, WI). For each Doppler measurement, three cycles were averaged, avoiding post-extrasystolic beats. Transaortic gradients were calculated using the simplified Bernoulli equation (P = 4ν2, where ΔP is in mmHg and v is the aortic velocity in m/s). EOA of the aortic valve was calculated from the continuity equation - EOA = CSALVOT × (LVOTVTI ÷ AoVTI) -, where EOA is in cm2, LVOTVTI is the subaortic velocity -time integral and AoVTI is the aortic velocity-time integral both in cm. CSALVOT is the cross sectional area (in cm2) of the LVOT calculated from the LVOT diameter measured in the parasternal long axis view (d in cm) assuming a circular geometry - CSALVOT = π x (d/2)(2. Left ventricular end diastolic and end systolic volumes (LVEDV and LVESV, respectively) and LVEF were assessed by standard 4 chamber and 2 chamber views using the biplane Simpson method. Stroke volume (SV) was calculated from the following equation: = LVOTVTI × CSALVOT, where SV is in mL/beat, LVOTVTI is in cm and CSALVOT is in cm2. Flow rate (Q) was calculated using 2 different methods:

  • a classical method using the formula Qclassic=1000×LVOTVTI×CSALVOTTE, where Qclassic is in mL/sec, LVOTVTI is in cm, CSALVOT is in cm2and ET is the ejection time in ms measured in the continuous wave Doppler of the aortic valve velocity spectrum.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
    https://doi.org/10.1161/CIRCULATIONAHA.1...

  • an alternative method using the formula Qalternative = CSALVOT × VmeanLVOT × 100, where Qalternative is in mL/sec, CSALVOT is in cm2 and VmeanLVOT is the mean velocity of blood in the LVOT during the ejection period in m/sec and is measured in the pulsed-wave Doppler of the LVOT velocity spectrum.55 Quiñones MA, Otto CM, Stoddard M, Waggoner A, Zoghbi WA; Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. Recommendations for quantification of Doppler echocardiography: A report from the Doppler quantification task force of the nomenclature and standards committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167-84. PMID: 11836492.

Patients with flow rate variation with dobutamine infusion ≥ |15|% in both classical and alternative methods were selected and simplified aortic valve area at 250 mL/s flow rate (EOAproj) was calculated according to the formula published by Blais et al44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
: EOAproj=EOAbasal+ΔEOAΔQ×250Qbasal, where EOAproj is in cm2, Q is the mean transvalvular flow rate, EOAbasal and Qbasal are the EOA and Q at rest and ΔEOA and ΔQ are the absolute variation in EOA and Q with dobutamine infusion.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
As we used two different methods to calculate flow rate we obtained two sets of values of simplified EOAproj in each eligible patient: 1) a classical simplified EOAproj using the classical method of flow rate calculation and 2) an alternative simplified EOAproj using the alternative method of flow rate calculation.

Statistical analysis

Categorical variables are described by frequencies and percentages. Continuous variables are presented as mean ± standard deviation.

A scatter plot and a linear regression model were constructed to assess the strength of linear relation between the classic and the alternative methods of calculation of EOAproj and to quantify the proportion of variance that the two methods have in common. Finally, in order to evaluate the agreement between the two methods (i.e., how much the new method is likely to differ from the old), we built a Bland-Altman plot - a plot of the paired differences between the two methods against their mean. Normal distribution of the paired differences was verified by the use of Shapiro-Wilk normality test. The bias was computed as the mean of the differences of the two methods. A one sample t test was conducted against the null hypothesis of no bias to evaluate the statistical significance of the calculated bias. Ninety-five percent-limits of agreement were computed as the mean bias plus or minus 1.96 time its standard deviation.66 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. PMID: 2868172. Two-tailed p values < 0,05 were considered statistically significant.

IBM SPSS Statistics version 23 (IBM, Vienna, Austria) and GraphPad Prism version 7.0 (GraphPad Software, La Jolla California, USA) were used for statistical analysis.

Results

Baseline characteristics

Between September/2011 and November/2015, 22 patients [15 (68%) men, mean age 72 ± 9 years] with classical LFLG AS underwent a low dose dobutamine stress echocardiography in order to evaluate the true severity of the AS. No major adverse events were reported. Table 1 shows the baseline clinical and echocardiographic features of these patients as well as the hemodynamic evolution with dobutamine infusion. 8 (36%) patients reached the AHA/ACC criteria for true severe aortic stenosis, 11(50%) patients maintained the valve area - gradient discordance present at baseline and 3 (14%) patients showed a progression of hemodynamic indices suggestive of pseudo severe aortic stenosis. No patient ended up the stress exam with inversion of the area - gradient mismatch (ie, aortic valve area > 1,0 cm2 and Vmax ≥ 4 m/s).

Table 1
Clinical and echocardiographic characteristics of the low-flow low-gradient aortic stenosis patients at baseline and at 20 ug/Kg/min Dobutamine infusion

Flow rate at baseline and at peak dobutamine infusion was calculated using both the classic Qclassic=1000×LVOTVTI×CSALVOTET and the alternative equations (Qalternative = CSALVOT × VmeanLVOT × 100) in all patients. Only 9 (41%) patients achieved a flow rate variation with dobutamine infusion assessed by both methods ≥ |15|%, enabling the simultaneous determination of the simplified projected aortic valve area at normal flow rate by the classic and the alternative formulas. Table 2 shows the baseline and peak dobutamine echocardiographic characteristics of this subset group of patients.

Table 2
Clinical and Echocardiographic Characteristics of the Low Flow Low Gradient Aortic Stenosis Patients with Flow Variation calculated by both methods ≥ |15| % with Dobutamine Infusion

A scatter plot showing the classic simplified projected aortic valve area values against the respective alternative simplified projected aortic valve area values was built (Figure 1). As suggested by the scatter plot, a strong linear association between the two methods of calculation was found - r (7) = 0,99, p < 0,001.

Figure 1
Scatter plot showing the classic simplified projected aortic valve area values against the alternative simplified projected aortic valve area values with a superimposed regression line (solid line) with 95% confidence bands (dashed lines).

Simple regression was conducted to find the best line that predicts the simplified projected aortic valve area calculated by the alternative method from the simplified projected aortic valve area calculated by the classic method. The results were statistically significant, F (1,7) = 245,5, p < 0,0001. The identified equation to understand this relationship was: alternative EOAproj = 1.00 (95% CI 0.85 - 1.15) x Classic EOAproj + 0,036 (95% CI -0.111 - 0.182). The adjusted R2 was 0.97, meaning that 97% of the variance of the alternative EOAproj can be explained by classic EOAproj.

A Bland-Altman analysis was performed to assess agreement between the two methods of EOAproj calculation. In Figure 2 the Y axis shows the differences between the two paired EOAproj measurements (alternative method - classic method) and the X axis represents the average of these measurements Alternativemethod+Classicmethod2. Normal distribution of the differences between paired measurements was verified by use of the Shapiro-Wilk test for normal distribution (test statistics = 0,854, df = 9, p = 0,082). There is no trend in increases in the variability of the differences in relation to their mean. The calculated bias (the average of the paired differences) is 0.037 cm2 (95% CI 0.004 - 0.066), meaning that on average EOAproj calculated by the alternative method measures 0.037 cm2 more than EOAproj calculated by the classic method. This bias is statistically significant (t = 2.619, df = 8, p = 0.031). The calculated 95% limits of agreement between the two methods are -0,04 and 0,12, which means that for 95% of the individuals, the EOAproj calculated by the alternative method would be between 0,04 cm2 less and 0,12 cm2 more than the EOAproj calculated by the classic method.

Figure 2
Bland-Altman plot, in which the difference of the two paired EOAproj measurements is plotted against their mean. The solid line parallel to the x axis represents the bias and the dashed lines parallel to the x axis represent the limits of agreement.

Discussion

The EOAproj is defined as the EOA of the aortic valve that would have occurred at a hypothetical standardized flow rate of 250 mL/s. This new echocardiographic index was developed in order to overcome the variable and unpredictable effect of dobutamine in flow rate.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
In fact, patients with classic LFLG AS undergoing DSE have a wide variable response in terms of flow rate progression, which may be due to multiple factors including the variable presence of myocardial contractile reserve, the unpredictable chronotropic response to dobutamine and the potential development of left ventricle dyssynchrony with dobutamine infusion33 Garnier F, Eicher JC, Jazayeri S, Bertaux G, Bouchot O, Aho LS, et al. Usefulness and limitations of contractile reserve evaluation in patients with low-flow, low-gradient aortic stenosis eligible for cardiac resynchronization therapy. Eur J Heart Fail. 2014;16(6):648-54. doi: 10.1002/ejhf.78.
https://doi.org/10.1002/ejhf.78...
Such variability in flow rate response may impose an insurmountable obstacle in the interpretation of ambiguous changes in mean pressure gradient and EOA. By normalizing the EOA at a hypothetical flow rate of 250 mL/s, the EOAproj enables direct comparison of AS severity in patients with classic LFLG AS that present different flow rate profiles with dobutamine infusion. In addition to make the interpretation of DSE results easier, this new parameter has also been shown to be related to actual AS severity (calcification at surgery) and to have an important value in mortality prediction.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
,77 Clavel MA, Fuchs C, Burwash IG, Mundigler G, Dumesnil JG, Baumgartner H, et al. Predictors of outcomes in low-flow, low-gradient aortic stenosis: results of the multicenter TOPAS Study. Circulation. 2008;118(14 Suppl):S234-42. doi: 10.1161/CIRCULATIONAHA.107.757427.
https://doi.org/10.1161/CIRCULATIONAHA.1...

In order to calculate the EOAproj, EOA is plotted against the mean transvalvular flow rate at different stages of DSE. The slope of this curve - called compliance - is then used to predict EOA at 250 mL/min.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
A simplified version of the original formula substitutes the curve slope for an easier to calculate quotient PeakEOARestEOAPeakQRestQ. Thus, the simplified version of the EOAproj formula can be expressed as EOAproj=EOAbasal+PeakEOARestEOAPeakQRestQ×250Qrest.88 Clavel MA, Burwash IG, Mundigler G, Dumesnil JG, Baumgartner H, Bergler-Klein J, et al. Validation of conventional and simplified methods to calculate projected valve area at normal flow rate in patients with low flow, low gradient aortic stenosis: the multicenter TOPAS (True or Pseudo Severe Aortic Stenosis) study. J Am Soc Echocardiogr. 2010;23(4):380-6. doi: 10.1016/j.echo.2010.02.002.
https://doi.org/10.1016/j.echo.2010.02.0...

Both the original and simplified version of the EOAproj formulae recommend the calculation of flow rate as the quotient between stroke volume and ET which requires 3 different measurements: 1) LVOT diameter (LVOTD); 2) LVOT velocity-time integral (LVOTVTI) and 3) ET measured at the aortic velocity spectrum. Both LVOTD and LVOTVTI are measures routinely done in DSE protocols performed for classic LFLG AS evaluation as they are needed to calculate EOA of the aortic valve by the continuity equation. However, the need for ET measured at the aortic velocity spectrum adds the requirement for an extra measurement in the usual protocol of DSE. Furthermore, this flow rate formula involves measurements acquired in different places and, inevitably, in different time points, encompassing an intrinsic bias.

Flow rate can also be determined by the product of left ventricular outflow tract area and left ventricular outflow tract mean velocity, which requires only 2 measurements: 1) LVOTD and 2) mean velocity of blood at LVOT during the ejection period (LVOTVmean). LVOTVmean is given automatically in most echocardiography software when assessing LVOTVTI (a fundamental step in EOA calculation by the continuity equation). This alternative formula is less cumbersome to calculate as it does not need an additional measurement in the aortic velocity spectrum. Also, as it only requires 2 different measurements, it is less prone to increased inter-observer and intra-observer variability.

This study aimed to assess how much the EOAproj calculated using an alternative method to estimate flow rate differs from the EOAproj calculated by the standard formula. The Bland-Altman method was used to assess agreement between the two methods. As previously published, Pearson correlation and linear regression analysis can be misleading in terms of assessing agreement between two measurement methods, as data which seem to be in poor agreement (for instance, a change in scale of measurement) can be highly correlated.66 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. PMID: 2868172.,99 Giavarina D. Understanding Bland Altman analysis. Biochem Med (Zagreb). 2015;25(2):141-51. doi: 10.11613/BM.2015.015
https://doi.org/10.11613/BM.2015.015...
Bland-Altman method assesses how well the methods agree on average (by estimating the mean of the differences for individuals - the systematic bias) and how well the measurements agree for individuals (by examining the variability of the differences and the calculation of the limits of agreement which quantify the range of values that can be expected to cover agreement for most of the subjects).1010 Bunce C. Correlation, agreement, and Bland-Altman analysis: statistical analysis of method comparison studies. Am J Ophthalmol. 2009;148(1):4-6. doi: 10.1016/j.ajo.2008.09.032.
https://doi.org/10.1016/j.ajo.2008.09.03...

Using the Bland-Altman method, we found a systematic bias of 0.037 cm2 (95% CI 0.004 - 0.066), meaning that on average the alternative method overestimates the EOAproj in 0,037 cm2 compared to the classic method. Despite being statistically significant, this bias is not clinically significant as it is less than 0.1 cm2. Also, the 95% limits of agreement are quite narrow (from -0,04 cm2 to 0,12 cm2), meaning that for 95% of individuals, EOAproj calculated by the alternative method would be between 0,04 cm2 less to 0,12 cm2 more than the EOAproj calculated by the classic equation. Such narrow range is the largest likely differences between the two methods, and do not compromise the clinical agreement between the two methods. Therefore, it is reasonable to acknowledge the potential interchangeability of the two methods of EOAproj calculation in clinical practice.

Conclusion

This study presented a new method to calculate the simplified EOA of the aortic valve at normal flow rate using a less cumbersome equation to estimate flow rate and tested the agreement of this new method with the previous reported by Blais et al.44 Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
https://doi.org/10.1161/CIRCULATIONAHA.1...
The bias and 95% limits of agreement of the new method are narrow and not clinically relevant, supporting the potential interchangeable use of both methods in clinical practice. As the new method requires less additional measurements, it would be easier to implement it in clinical practice, promoting an increase in the use of EOAproj - a valuable echocardiographic parameter in the evaluation of LFLG AS.

Limitations

This is a small retrospective single-institution study that is inherently underpowered to assess small differences in echocardiographic variables between groups. A higher number of patients is needed to investigate potential discrepancies in the performance of both EOAproj calculation methods in different subsets of LFLG AS patients. Therefore, the results presented here must be interpreted with caution.

  • Sources of Funding
    There were no external funding sources for this study.
  • Study Association
    This study is not associated with any thesis or dissertation work.
  • Ethics approval and consent to participate
    This study was approved by the Ethics Committee of the Faculdade de Medicina da Universidade de Coimbra under the protocol number CE-016/2017. All the procedures in this study were in accordance with the 1975 Helsinki Declaration, updated in 2013. Informed consent was obtained from all participants included in the study.

References

  • 1
    Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Guyton RA, et al; ACC/AHA Task Force Members. 2014 AHA/ACC Guideline for the Management of Patients With Valvular Heart Disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation. 2014;129(23):e521-643. doi: 10.1161/CIR.0000000000000031. Erratum in: Circulation. 2014;130(13):e120.
    » https://doi.org/10.1161/CIR.0000000000000031
  • 2
    Awtry E, Davidoff R. Low-flow/low-gradient aortic stenosis. Circulation. 2011;124(23):e739-41. doi: 10.1161/CIRCULATIONAHA.111.075853.
    » https://doi.org/10.1161/CIRCULATIONAHA.111.075853
  • 3
    Garnier F, Eicher JC, Jazayeri S, Bertaux G, Bouchot O, Aho LS, et al. Usefulness and limitations of contractile reserve evaluation in patients with low-flow, low-gradient aortic stenosis eligible for cardiac resynchronization therapy. Eur J Heart Fail. 2014;16(6):648-54. doi: 10.1002/ejhf.78.
    » https://doi.org/10.1002/ejhf.78
  • 4
    Blais C, Burwash IG, Mundigler G, Dumesnil JG, Loho N, Rader F, et al. Projected valve area at normal flow rate improves the assessment of stenosis severity in patients with low-flow, low-gradient aortic stenosis: the multicenter TOPAS (Truly or Pseudo-Severe Aortic Stenosis) study. Circulation. 2006;113(5):711-21. doi: 10.1161/CIRCULATIONAHA.105.557678.
    » https://doi.org/10.1161/CIRCULATIONAHA.105.557678
  • 5
    Quiñones MA, Otto CM, Stoddard M, Waggoner A, Zoghbi WA; Doppler Quantification Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. Recommendations for quantification of Doppler echocardiography: A report from the Doppler quantification task force of the nomenclature and standards committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167-84. PMID: 11836492.
  • 6
    Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. PMID: 2868172.
  • 7
    Clavel MA, Fuchs C, Burwash IG, Mundigler G, Dumesnil JG, Baumgartner H, et al. Predictors of outcomes in low-flow, low-gradient aortic stenosis: results of the multicenter TOPAS Study. Circulation. 2008;118(14 Suppl):S234-42. doi: 10.1161/CIRCULATIONAHA.107.757427.
    » https://doi.org/10.1161/CIRCULATIONAHA.107.757427
  • 8
    Clavel MA, Burwash IG, Mundigler G, Dumesnil JG, Baumgartner H, Bergler-Klein J, et al. Validation of conventional and simplified methods to calculate projected valve area at normal flow rate in patients with low flow, low gradient aortic stenosis: the multicenter TOPAS (True or Pseudo Severe Aortic Stenosis) study. J Am Soc Echocardiogr. 2010;23(4):380-6. doi: 10.1016/j.echo.2010.02.002.
    » https://doi.org/10.1016/j.echo.2010.02.002
  • 9
    Giavarina D. Understanding Bland Altman analysis. Biochem Med (Zagreb). 2015;25(2):141-51. doi: 10.11613/BM.2015.015
    » https://doi.org/10.11613/BM.2015.015
  • 10
    Bunce C. Correlation, agreement, and Bland-Altman analysis: statistical analysis of method comparison studies. Am J Ophthalmol. 2009;148(1):4-6. doi: 10.1016/j.ajo.2008.09.032.